An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of phenobarbital in human serum and plasma

被引:1
|
作者
Schierscher, Tobias [1 ]
Salzmann, Linda [1 ]
Singh, Neeraj [2 ]
Bachmann, Martina [2 ]
Kobel, Anja [1 ]
Wild, Janik [1 ]
Bauland, Friederike [3 ]
Geistanger, Andrea [2 ]
Risch, Lorenz [1 ]
Geletneky, Christian [2 ]
Seger, Christoph [1 ]
Taibon, Judith [2 ,4 ]
机构
[1] Dr Risch Ostschweiz AG, Buchs, Switzerland
[2] Roche Diagnost GmbH, Penzberg, Germany
[3] Chrestos Concept GmbH & Co KG, Essen, Germany
[4] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
关键词
phenobarbital; ID-LC-MS/MS; reference measurement procedure; SI units; traceability; ANTIEPILEPTIC DRUGS; SPECIFICATIONS; PERFORMANCE; METABOLITES; IONIZATION; EPILEPSY; EFFICACY; POSITION; BLOOD; MS;
D O I
10.1515/cclm-2023-1104
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Phenobarbital serves as an antiepileptic drug (AED) and finds application in the treatment of epilepsy either as monotherapy or adjunctive therapy. This drug exhibits various pharmacodynamic properties that account for its beneficial effects as well as potential side effects. Accurate measurement of its concentration is critical for optimizing AED therapy through appropriate dose adjustments. Therefore, our objective was to develop and validate a new reference measurement procedure (RMP) for the accurate quantification of phenobarbital levels in human serum and plasma. Methods: A sample preparation protocol based on protein precipitation followed by a high dilution step was established in combination with a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method using a C8 column to separate target analytes from known and unknown interferences. Assay validation and determination of measurement uncertainty were performed based on current guidelines. Selectivity and Specificity were assessed using spiked serum and plasma samples; to investigate possible matrix effects (MEs) a post-column infusion experiment and a comparison of standard line slopes was performed. Precision and accuracy were determined within a multiday precision experiment. Results: The RMP was shown to be highly selective and specific, with no evidence of matrix interferences. It can be used to quantify phenobarbital in the range of 1.92 to 72.0 mu g/mL. Intermediate precision was less than 3.2 %, and repeatability coefficient of variation (CV) ranged from 1.3 to 2.0 % across all concentration levels. The relative mean bias ranged from -3.0 to -0.7 % for native serum levels, and from -2.8 to 0.8 % for Li-heparin plasma levels. The measurement uncertainties (k=1) for single measurements and target value assignment were 1.9 to 3.3 % and 0.9 to 1.6 %, respectively. Conclusions: A novel LC-MS/MS-based candidate RMP for the quantification of phenobarbital in human serum and plasma is presented which can be used for the standardization of routine assays and the evaluation of clinically relevant samples.
引用
收藏
页码:1314 / 1326
页数:13
相关论文
共 50 条
  • [41] Candidate reference measurement procedure for determination of urea in serum by liquid chromatography-tandem mass spectrometry
    Han, Liqiao
    Huang, Xiaoting
    Zhang, Lu
    Zhang, Qiaoxuan
    Wang, Jianbing
    Lin, Haibiao
    Yan, Jun
    Zhuang, Junhua
    Huang, Xianzhang
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2019, 162 : 124 - 129
  • [42] Evaluation and clinical application of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for serum theophylline
    Du, Yuanyuan
    Yan, Qiaofang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Zhang, Pengwei
    Yan, Jun
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (22) : 4897 - 4906
  • [43] Quantification of psilocin in human whole blood using liquid chromatography-tandem mass spectrometry (LC-MS/MS)
    Gomonit, Munchelou M.
    Skillman, Britni
    Swortwood, Madeleine J.
    JOURNAL OF FORENSIC SCIENCES, 2024, 69 (02) : 678 - 687
  • [44] Development of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry
    Lee, Hwashim
    Park, Sangryoul
    Kim, Byungjoo
    Lee, Gaeho
    ANALYTICAL SCIENCES, 2007, 23 (03) : 299 - 303
  • [45] Development of a Candidate Reference Method for Determination of Tyrosine in Serum by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Hwashim Lee
    Sangryoul Park
    Byungjoo Kim
    Gaeho Lee
    Analytical Sciences, 2007, 23 : 299 - 303
  • [46] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [47] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [48] Simultaneous Determination of Creatinine and Creatine in Human Serum by Double-Spike Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) and Gas Chromatography-Mass Spectrometry (GC-MS)
    Fernandez-Fernandez, Mario
    Rodriguez-Gonzalez, Pablo
    Anon Alvarez, M. Elena
    Rodriguez, Felix
    Alvarez Menendez, Francisco V.
    Garcia Alonso, J. Ignacio
    ANALYTICAL CHEMISTRY, 2015, 87 (07) : 3755 - 3763
  • [49] Measurement of Plasma Free Metanephrines Using a Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) Method
    MacMahon, M. M.
    Mulready, K. J.
    Fitzgibbon, M. C.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2013, 182 : S406 - S406
  • [50] Development of isotope dilution-liquid chromatography/tandem mass spectrometry as a candidate reference method for the determination of folic acid
    Jung, Minyoung
    Kim, Byungjoo
    Boo, Doo Wan
    So, Hun-Young
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2007, 28 (05): : 745 - 750