HEMOLYSIS DETECTION FOR ETHANOL MEASUREMENT IN WHOLE BLOOD SAMPLES BEFORE CENTRIFUGATION: HEMCHECK DEVICE EVALUATION

被引:0
|
作者
Zengi, Oguzhan [1 ]
Boz, Meltem [1 ]
Yesil, Beyazit Semih [1 ]
Gumus, Alper [1 ]
机构
[1] Hlth Sci Univ, Istanbul Basaksehir Cam & Sakura City Hosp, Dept Med Biochem, Olimpiyat Yolu Basaksehir, Istanbul, Turkiye
关键词
hemolysis; pre-analytical phase; ethanol; point-of-care testing; quality improvement; SPECIMENS; ASSAY;
D O I
10.5937/jomb0-41574
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Background: As previously reported, the measurement of ethanol can also be affected by interference from hemolysis. This is a matter of concern since ethanol is widely regarded as the most commonly abused substance globally. When sample re-collection is ordered to eliminate hemolysis effects for ethanol testing, this can have unfavourable consequences for these patients. Rapid detection of hemolysed specimens would alleviate some issues associated with forensic samples. This study aimed to assess the qualitative analytical performance of a novel point-of-care testing device per the guidelines specified in CLSI-EP-12A document. HemCheckT" is a novel POCT device that qualitatively detects free-hemoglobin levels on the specimen shortly after drawing the sample.Methods: The system consists of two components. One is a cartridge with a needle that is used to transfer a small volume of whole blood from a vacuum tube to vertical and lateral flow filtration. The second component is the reader. The consumable cartridges are designed to be inserted into the reader without requiring the syringe or blood collection tube removal. A red indicator led illuminates, indicating that the sample has been hemolysed. To assess the imprecision of the method, we determined the C5-C95 interval and C50, using the Roche Cobas clinical chemistry analyser as the comparator. For this study, we utilised residual samples.Results: Our C50 concentrations were 45 mg/dL, the C5- C95 30 mg/dL and 60 mg/dL, respectively. Two methods reveal an overall agreement of 89.2%. The diagnostic sensitivity and specificity of the study showed a positive agree-ment of 95.7% and a negative agreement of 80.0%. The calculated Cohens Kappa value was 77.3%.Conclusion: According to current literature, a bias in ethanol levels can be noticed even with small free hemo-globin concentrations. When it came to forensic samples, this bias was crucial. Our study confirmed that the HemCheck device has acceptable analytical performance, as outlined in the CLSI-EP-12A document. Although the possible beneficial impacts of the innovative POCT hemol-ysis detection device on the preanalytical phase are pro-posed, we have not studied them in this study.
引用
收藏
页码:600 / 606
页数:7
相关论文
共 50 条
  • [1] Optofluidic Sensor for Inline Hemolysis Detection on Whole Blood
    Zhou, Chen
    Hedayati, Mehdi Keshavarz
    Zhu, Xiaolong
    Nielsen, Frank
    Levy, Uriel
    Kristensen, Anders
    ACS SENSORS, 2018, 3 (04): : 784 - 791
  • [2] Detection of progesterone in whole blood samples
    Ehrentreich-Förster, E
    Scheller, FW
    Bier, FF
    BIOSENSORS & BIOELECTRONICS, 2003, 18 (04): : 375 - 380
  • [3] Evaluation of an electrochemical biosensor for uric acid measurement in human whole blood samples
    Liao, Li-Ting
    Liao, Chi-Chih
    Liu, Chiu-Ching
    Yang, Ting-Ya
    Wang, Giueng-Chueng
    CLINICA CHIMICA ACTA, 2014, 436 : 72 - 77
  • [4] Arginine measurement in dry and whole blood samples
    Manan, Z.
    Makhdoom, Az
    Jain, S.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2025, 369 : S299 - S300
  • [5] The Evaluation of Uncertainty for the Measurement of Blood Ethanol
    Ince, F. Demet
    Arslan, Banu
    Kaplan, Y. Cem
    Ellidag, Hamit
    JOURNAL OF CLINICAL AND ANALYTICAL MEDICINE, 2016, 7 (05) : 648 - 651
  • [6] Pre-analytical stability before centrifugation of 7 biochemical analytes in whole blood
    Perrier-Cornet, Andreas
    Moineau, Marie-Pierre
    Narbonne, Valerie
    Plee-Gautier, Emmanuelle
    Le Saos, Fabienne
    Carre, Jean-Luc
    ANNALES DE BIOLOGIE CLINIQUE, 2015, 73 (04) : 454 - 460
  • [7] Measurement of the Phagocytic Activity of Human Peripheral Blood Using a Highly Sensitive Fluorometric Detection Device Without Hemolysis
    Zhang, Ran
    Inagawa, Hiroyuki
    Takahashi, Masaru
    Kawanishi, Ami
    Kazumura, Kimiko
    Tsuchiya, Hiroshi
    Morishita, Naokazu
    Kobayashi, Yutaro
    Masaki, Tsutomu
    Kobara, Hideki
    Soma, Gen-Ichiro
    ANTICANCER RESEARCH, 2017, 37 (07) : 3897 - 3903
  • [8] Hemolysis Rates in Whole Blood Samples for Blood Gas/Electrolyte Analysis by Point-of-Care Testing
    Wilson, Amanda
    Sweeney, Maranna
    Lynch, P. L. Mark
    O'Kane, Maurice J.
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2018, 3 (01): : 144 - 145
  • [9] Multicentre evaluation of an immunological rapid test for the detection of troponin T in whole blood samples
    Collinson, PO
    Gerhardt, W
    Katus, HA
    MullerBardorff, M
    Braun, S
    Schricke, U
    Vogt, W
    Nagel, D
    Zander, M
    Leinberger, R
    Mangold, D
    Zerback, R
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1996, 34 (07): : 591 - 598
  • [10] Fast and sensitive smartphone colorimetric detection of whole blood samples on a paper-based analytical device
    Yan, Xiang-Hong
    Ji, Bin
    Fang, Fang
    Guo, Xiao-Lin
    Zhao, Shuang
    Wu, Zhi-Yong
    Talanta, 2024, 270