An audit of patient radiation doses in interventional radiology at a South African hospital

被引:1
|
作者
Slave, Oneile [1 ]
Mahomed, Nasreen [1 ]
机构
[1] Univ Witwatersrand, Fac Hlth Sci, Dept Radiol, Johannesburg, South Africa
来源
SA JOURNAL OF RADIOLOGY | 2023年 / 27卷 / 01期
关键词
Interventional radiology; diagnostic reference level; kerma air product; reference point air kerma; fluoroscopy time; DIAGNOSTIC REFERENCE LEVELS; MULTICENTER; EXPOSURE; IR;
D O I
10.4102/sajr.v27i1.2559
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Interventional radiology (IR) is becoming more relevant in patient care and is associated with increased patient radiation exposure and radiation-induced adverse effects. Diagnostic reference levels (DRLs) are crucial for radiation control. There is a paucity of published DRLs for IR in South Africa and sub-Saharan Africa. Objectives: This study aimed to determine local DRLs for fluoroscopically-guided IR procedures and compare the achieved DRLs with published local and international DRLs. Method: Retrospective, descriptive, single-centre study. Kerma air product (KAP), reference point air kerma (Ka,r) and fluoroscopy time (FT) were collected for patients (12 years and older) who underwent IR procedures at a university hospital from 01 January 2019 to 31 December 2019. The 75th percentile of the distribution of each dose parameter (KAP, Ka,r and FT) per procedure was calculated and taken as the local diagnostic reference levels (LDRL). The established LDRLs were compared to published DRLs. Results: A total of 564 cases were evaluated. The 13 most frequent procedures (with 15 or more cases) represented 86.1% (487/564). Percutaneous transhepatic biliary drainage was the most common procedure (n = 146, 25.9%). Diagnostic cerebral angiogram DRLs exceeded the published DRL data ranges for all parameters (DRL 209.3), and interventional cerebral angiogram exceeded published ranges (DRL 275). Uterine artery embolisation (UAE) exceeded these ranges for KAP and Ka,r. (KAP-954.9 Gy/cm2, Ka,r-2640.8 mGy). Conclusion: The LDRLs for diagnostic cerebral angiogram, interventional cerebral angiogram and UAE exceeded published international DRL ranges. These procedures require radiation optimisation as recommended by the International Commission on Radiological Protection (ICRP). Contribution: In addition to informing radiation protection practices at the level of the institution, the established LDRLs contribute towards Regional and National DRLs.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Staff Radiation Doses in Interventional Cardiology: Correlation With Patient Exposure
    Vano, Eliseo
    Ubeda, Carlos
    Leyton, Fernando
    Miranda, Patricia
    Gonzalez, Luciano
    PEDIATRIC CARDIOLOGY, 2009, 30 (04) : 409 - 413
  • [32] Staff Radiation Doses in Interventional Cardiology: Correlation With Patient Exposure
    Eliseo Vano
    Carlos Ubeda
    Fernando Leyton
    Patricia Miranda
    Luciano Gonzalez
    Pediatric Cardiology, 2009, 30 : 409 - 413
  • [33] Patient radiation doses in paediatric interventional cardiology and optimization actions
    Ubeda, Carlos
    Vano, Eliseo
    Riquelme, Nemorino
    Aguirre, Daniel
    Vasquez, Hector
    Chavez, Cristian
    Dalmazzo, Dandaro
    RADIATION PHYSICS AND CHEMISTRY, 2020, 168
  • [34] Patient radiation doses in paediatric interventional cardiology procedures: a review
    Harbron, R. W.
    Dreuil, S.
    Bernier, M-O
    Pearce, M. S.
    Thierry-Chef, I.
    Chapple, C-L
    Baysson, H.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (04) : R131 - R144
  • [35] An audit of patient radiation doses during interventional cardiology procedures in Uttarakhand, India, and establishment of local diagnostic reference levels
    Uniyal, Satish C.
    Singh, Vikram
    Rawat, Anurag
    Gururani, Kunal
    Belwal, Chandra Mohan
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) : 476 - 487
  • [36] Radiation risks in interventional radiology
    Dendy, P. P.
    BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (961): : 1 - 7
  • [37] Radiation protection in interventional radiology
    Johnson, DR
    Kyriou, J
    Morton, EJ
    Clifton, A
    Fitzgerald, M
    MacSweeney, E
    CLINICAL RADIOLOGY, 2001, 56 (02) : 99 - 106
  • [38] Radiation protection in interventional radiology
    Faulkner, K
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832): : 325 - 326
  • [39] Interventional radiology and radiation therapy
    Janaki, Hadjiev
    ORVOSI HETILAP, 2015, 156 (17) : 706 - 710
  • [40] Radiation exposure in interventional radiology
    Pinto, N. G. V.
    Braz, D.
    Vallim, M. A.
    Filho, L. G. P.
    Azevedo, F. S.
    Barroso, R. C.
    Lopes, R. T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 586 - 590