SEM Analysis and Micro-CT Evaluation of Four Dental Implants after Three Different Mechanical Requests-In Vitro Study

被引:2
|
作者
Vinhas, Ana Sofia [1 ]
Salazar, Filomena [1 ]
Mendes, Jose Manuel [1 ]
Silva, Antonio Sergio [1 ]
Rios-Carrasco, Blanca [2 ]
Rios-Santos, Jose Vicente [2 ]
Gil, Javier [3 ]
Herrero-Climent, Mariano [4 ]
Aroso, Carlos [1 ]
机构
[1] Univ Inst Hlth Sci IUCS, Oral Pathol & Rehabil Res Unit UNIPRO, CESPU, Rua Cent Gandra, P-4585116 Gandra, Portugal
[2] Univ Seville, Sch Dent, Dept Periodontol, C Avicena S-N, Seville 41009, Spain
[3] Univ Int Catalunya, Bioengn Inst Technol, Fac Med & Hlth Sci, St Cugat Barcelona 08195, Spain
[4] Porto Dent Inst, P-4150518 Porto, Portugal
关键词
implant-abutment connection; implant-abutment axial displacement; implant-abutment wear; fatigue-test; mechanical stresses; SEM analysis; Micro-CT evaluation; EXTERNAL HEXAGON; ABUTMENT CONNECTION; INTERNAL CONNECTION; AXIAL DISPLACEMENT; FRACTURE-BEHAVIOR; REVERSE TORQUE; FATIGUE; TITANIUM; ROUGHNESS; PERFORMANCE;
D O I
10.3390/ma17020434
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Statement of problem: Implant-supported rehabilitations are an increasingly frequent practice to replace lost teeth. Before clinical application, all implant components should demonstrate suitable durability in laboratory studies, through fatigue tests. Objective: The purpose of this in vitro study was to evaluate the integrity and wear of implant components using SEM, and to assess the axial displacement of the implant-abutment assembly by Micro-CT, in different implant connections, after three distinct mechanical requests. Materials and methods: Four KLOCKNER implants (external connection SK2 and KL; and internal connection VEGA and ESSENTIAL) were submitted to three different mechanical requests: single tightening, multiple tightening, and multiple tightening and cyclic loading (500 N x 100 cycles). A total of 16 samples were evaluated by SEM, by the X-ray Bragg-Brentano method to obtain residual stresses, and scratch tests were realized for each surface and Micro-CT (4 control samples; 4 single tightening; 4 multiple tightening; 4 multiple tightening and cyclic loading). All dental implants were fabricated with commercially pure titanium (grade 3 titanium). Surface topography and axial displacement of abutment into the implant, from each group, were evaluated by SEM and Micro-CT. Results: In the manufacturing state, implants and abutments revealed minor structural changes and minimal damage from the machining process. The application of the tightening torque and loading was decisive in the appearance and increase in contact marks on the faces of the hexagon of the abutment and the implant. Vega has the maximum compressive residual stress and, as a consequence, higher scratch force. The abutment-implant distances in SK2 and KL samples did not show statistically significant differences, for any of the mechanical demands analyzed. In contrast, statistically significant differences were observed in abutment-implant distance in the internal connection implants Vega and Essential. Conclusions: The application of mechanical compression loads caused deformation and contact marks in all models tested. Only internal connection implants revealed an axial displacement of the abutment into the implant, but at a general level, a clear intrusion of the abutment into the implant could only be confirmed in the Essential model, which obtained its maximal axial displacement with cyclic loading.
引用
收藏
页数:21
相关论文
共 50 条
  • [31] Evaluation of Stresses Occurring on Three Different Zirconia Dental Implants: Three-Dimensional Finite Element Analysis
    Caglar, Alper
    Bal, Bilge Turhan
    Aydin, Cemal
    Yilmaz, Handan
    Ozkan, Secil
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2010, 25 (01) : 95 - 103
  • [32] Evaluation of positional accuracy in multiple implants using four different splinting materials: An in vitro study
    Joseph, Thara Maria
    Ravichandran, R.
    Lylajam, S.
    Viswabharan, Prasanth
    Janardhanan, Kavitha
    Rajeev, Smitha
    JOURNAL OF INDIAN PROSTHODONTIC SOCIETY, 2018, 18 (03): : 239 - 247
  • [33] Evaluation with micro-CT of different anticoagulation strategies during hemodialysis in patients with thrombocytopenia: A randomized crossover study
    Vanommeslaeghe, Floris
    De Somer, Filip
    Josipovic, Ivan
    Boone, Matthieu
    Dhondt, Annemieke
    Van Biesen, Wim
    Eloot, Sunny
    ARTIFICIAL ORGANS, 2019, 43 (08) : 756 - 763
  • [34] Histological and micro-CT analysis of peri-implant soft and hard tissue healing on implants with different healing abutments configurations
    Souza, Andre Barbisan
    Alshihri, AbdulMonem
    Kaemmerer, Peer W.
    Araujo, Mauricio G.
    Gallucci, German O.
    CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 (10) : 1007 - 1015
  • [35] Macrophagic Inflammatory Response Next to Dental Implants with Different Macro- and Micro-Structure: An In Vitro Study
    Menini, Maria
    Delucchi, Francesca
    Baldi, Domenico
    Pera, Francesco
    Bagnasco, Francesco
    Pesce, Paolo
    APPLIED SCIENCES-BASEL, 2021, 11 (12):
  • [36] Mineral density changes in bovine carious dentin after treatment with bioactive dental cements: a comparative micro-CT study
    Neves, Aline Borburema
    Bergstrom, Tamiris Gomes
    Fonseca-Goncalves, Andrea
    Pires dos Santos, Thais Maria
    Lopes, Ricardo Tadeu
    Neves, Aline de Almeida
    CLINICAL ORAL INVESTIGATIONS, 2019, 23 (04) : 1865 - 1870
  • [37] Mineral density changes in bovine carious dentin after treatment with bioactive dental cements: a comparative micro-CT study
    Aline Borburema Neves
    Tamiris Gomes Bergstrom
    Andréa Fonseca-Gonçalves
    Thais Maria Pires dos Santos
    Ricardo Tadeu Lopes
    Aline de Almeida Neves
    Clinical Oral Investigations, 2019, 23 : 1865 - 1870
  • [38] Efficacy of three irrigation methods in removing calcium hydroxide from curved root canals: An in vitro micro-CT study
    Radeva, Elka Nikolaeva
    Mironova, Jasmina, V
    Marinova-Takorova, Mirela B.
    JOURNAL OF INTERNATIONAL SOCIETY OF PREVENTIVE AND COMMUNITY DENTISTRY, 2023, 13 (04): : 342 - 348
  • [39] Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro study
    Jorio, Isabelle C.
    Stawarczyk, Bogna
    Attin, Thomas
    Schmidlin, Patrick R.
    Sahrmann, Philipp
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (08) : 881 - 892
  • [40] Biomechanical evaluation of dental implants with three different designs: Removal torque and resonance frequency analysis in rabbits
    Gehrke, Sergio Alexandre
    Marin, Giovanni Wiel
    ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2015, 199 : 30 - 35