Accuracy evaluation of two portable blood glucose meters in feline patients using whole blood samples

被引:0
|
作者
Moresco, Mauricio Bianchini [1 ]
Matesco, Viviana Cauduro [2 ]
de Moura Martins, Francisco Savio [3 ]
Carvalho de Carvalho, Guilherme Luiz [3 ]
Schaefer, Gabriela da Cruz [3 ]
Nunes, Nilson Junior da Silva [3 ,4 ]
Valle, Stella de Faria [3 ,4 ,5 ]
Poppl, Alan Gomes [3 ,6 ]
机构
[1] Univ Fed Rio Grande Sul UFRGS, Hosp Clin Vet, Porto Alegre, RS, Brazil
[2] Ctr Univ Ritter Reis UniRitter, Hosp Vet, Porto Alegre, RS, Brazil
[3] Univ Fed Rio Grande Sul UFRGS, Fac Vet FaVet, Programa Posgrad Ciencias Vet, Porto Alegre, RS, Brazil
[4] Univ Fed Rio Grande Sul UFRGS, Fac Vet FaVet, Lab Anal Clin Vet, Porto Alegre, RS, Brazil
[5] Univ Fed Rio Grande Sul UFRGS, Fac Vet FaVet, Dept Patol Clin Vet, Porto Alegre, RS, Brazil
[6] Univ Fed Rio Grande Sul UFRGS, Fac Vet FaVet, Dept Med Anim, BR-91540000 Porto Alegre, RS, Brazil
来源
CIENCIA RURAL | 2023年 / 53卷 / 10期
关键词
cats; glycemia; hematocrit; portable blood glucose meters; CATS; DOGS;
D O I
10.1590/0103-8478cr20220415
中图分类号
S3 [农学(农艺学)];
学科分类号
0901 ;
摘要
Using portable blood glucose meters (PBGMs) to measure blood glucose (BG) concentration is a common procedure in veterinary practice. Our objective was to evaluate the analytical and clinical accuracy of a human PBGM (Accu-Chek Performa (R)), (AC) and a veterinary PBGM (GlucoCalea (R)), (GC) in feline patients. Central venous blood samples were collected from 48 cats at a Brazilian Veterinary teaching hospital. Two devices from each model were used and compared to a reference method (RM). Analytical accuracy was assessed according to ISO 15197:2013 requirements for human PBGMs. Data were compared using Wilcoxon's nonparametric test and represented by Bland-Altman plots. Hematocrit's effect on BG measurements was evaluated by the Spearman correlation coefficient. Clinical accuracy was determined using error grid analysis (EGA). Values of BG were significantly higher in all PBGMs compared to the RM. Although ISO's analytical accuracy requirements could not be met by any of the devices, AC meters were more accurate than GC meters. All AC measurements -but not GC ones -were within zones A and B of the EGA, meeting ISO requirements for clinical accuracy. Significant hematocrit interference was observed in all devices. Therefore, AC showed greater accuracy compared to GC using feline whole blood samples.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Performance of two portable meters and a benchtop analyzer for blood glucose concentration measurement in rabbits
    Selleri, Paolo
    Di Girolamo, Nicola
    Novari, Gianluca
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2014, 245 (01): : 87 - 98
  • [22] CAPILLARY BLOOD-GLUCOSE MEASUREMENTS IN-HOSPITAL INPATIENTS USING PORTABLE GLUCOSE METERS
    PHILLIPOU, G
    SEABORN, CJ
    HOOPER, J
    PHILLIPS, PJ
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1993, 23 (06): : 667 - 671
  • [23] Assessment of five portable blood glucose meters for use in cats
    Wess, G
    Reusch, C
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2000, 61 (12) : 1587 - 1592
  • [24] A Perspective on the Accuracy of Blood Glucose Meters During Pregnancy
    Immanuel, Jincy
    Simmons, David
    DIABETES CARE, 2018, 41 (10) : 2053 - 2058
  • [25] Evaluating an Accuracy of Blood Glucose Meters in the Practice: Is this possible?
    Ziegler, R.
    Freckmann, G.
    Heinemann, L.
    DIABETES STOFFWECHSEL UND HERZ, 2014, 23 (06): : 345 - 347
  • [26] Accuracy of Handheld Blood Glucose Meters at High Altitude
    de Mol, Pieter
    Krabbe, Hans G.
    de Vries, Suzanna T.
    Fokkert, Marion J.
    Dikkeschei, Bert D.
    Rienks, Rienk
    Bilo, Karin M.
    Bilo, Henk J. G.
    PLOS ONE, 2010, 5 (11):
  • [27] Accuracy of blood glucose meters in pregnant subjects with diabetes
    Stenger, P
    Allen, ME
    Lisius, L
    DIABETES CARE, 1996, 19 (03) : 268 - 269
  • [28] Comparison of the accuracy of seven home blood glucose meters
    Koulas, Georgios
    Sampanis, Christos
    Tziomalos, Konstantinos
    Semertzidis, Panagiotis
    Athanasiadou, Zoe
    Dogramatzi, Fani
    Machera, Konstantina
    Panagiotidou, Dimitra
    Zamboulis, Chrysanthos
    DIABETES, 2006, 55 : A96 - A96
  • [29] Quality guidelines and lack of accuracy of blood glucose meters
    Miedema, K
    CLINICA CHIMICA ACTA, 2005, 355 : S78 - S78
  • [30] Accuracy of home blood glucose meters during hypoglycemia
    Trajanoski, Z
    Brunner, GA
    Gfrerer, RJ
    Wach, P
    Pieber, TR
    DIABETES CARE, 1996, 19 (12) : 1412 - 1415