共 50 条
Residual leak following left atrial appendage occlusion with lariat closed with amplatzer PFO occluder device
被引:1
|作者:
Bawa, Danish
[1
]
Ahmed, Adnan
[1
]
Darden, Douglas
[1
]
Lakkireddy, Dhanunjaya
[1
]
机构:
[1] Kansas City Heart Rhythm Inst, Overland Pk, KS USA
来源:
关键词:
amplatzer PFO occluder device;
lariat device;
left atrial appendage closure;
residual leak;
D O I:
10.1111/pace.14669
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
BackgroundLeaks from incomplete ligation of the left atrial appendage (LAA) following closure with Lariat device are not uncommon and associated with higher stroke rate. Clinical VignetteThis clinical vignette highlights a procedure of closure of residual leak following closure with Lariat device. There was a persistent central leak of 3-4 mm with bidirectional flow, confirmed on TEE. The Amplatzer Talisman PFO Occluder was chosen for the closure of the leak because of its asymmetric discs with a smaller distal disc and expandable waist. ConclusionIn situations where the options are limited for PDL closure, such as the one described in this vignette, it is paramount to continue to find innovative solutions to emerging challenges in transcatheter interventions.
引用
收藏
页码:440 / 442
页数:3
相关论文