Agreement between cardiac output estimation by multi-beat analysis of arterial blood pressure waveforms and continuous thermodilution in post cardiac surgery intensive care unit patients

被引:6
|
作者
Khanna, Ashish K. [1 ,2 ,3 ]
Nosow, Lillian [4 ]
Sands, Lauren [5 ]
Saha, Amit K. [3 ,6 ]
Agashe, Harshavardhan [7 ]
Harris, Lynnette [3 ,6 ]
Martin, R. Shayn [8 ]
Marchant, Bryan [9 ]
机构
[1] Atrium Hlth Wake Forest Baptist Med Ctr, Wake Forest Sch Med, Dept Anesthesiol, Sect Crit Care Med, Winston Salem, NC 27157 USA
[2] Outcomes Res Consortium, Cleveland, OH 44112 USA
[3] Perioperat Outcomes & Informat Collaborat POIC, Winston Salem, NC 27101 USA
[4] Wake Forest Sch Med, Winston Salem, NC 27101 USA
[5] Univ Maryland, Sch Med, Baltimore, MD 21201 USA
[6] Atrium Hlth Wake Forest Baptist Med Ctr, Wake Forest Sch Med, Dept Anesthesiol, Winston Salem, NC USA
[7] Retia Med, Valhalla, NY USA
[8] Atrium Hlth Wake Forest Baptist Med Ctr, Wake Forest Sch Med, Dept Surg, Winston Salem, NC 27157 USA
[9] Atrium Hlth Wake Forest Baptist Med Ctr, Wake Forest Sch Med, Dept Anesthesiol, Sect Crit Care Med,Sect Cardiac Anesthesiol, Winston Salem, NC USA
关键词
Pulse wave analysis; Cardiac output; Hemodynamics; Monitoring; Cardiac surgery; Post-operative; Intensive care unit; METAANALYSIS; STATISTICS;
D O I
10.1007/s10877-022-00924-z
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We sought to assess agreement of cardiac output estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CTD) and a novel pulse wave analysis (PWA) method that performs an analysis of multiple beats of the arterial blood pressure waveform (CO-MBA) in post-operative cardiac surgery patients. PAC obtained CO-CTD measurements were compared with CO-MBA measurements from the Argos monitor (Retia Medical; Valhalla, NY, USA), in prospectively enrolled adult cardiac surgical intensive care unit patients. Agreement was assessed via Bland-Altman analysis. Subgroup analysis was performed on data segments identified as arrhythmia, or with low CO (less than 5 L/min). 927 hours of monitoring data from 79 patients was analyzed, of which 26 had arrhythmia. Mean CO-CTD was 5.29 +/- 1.14 L/min (bias +/- precision), whereas mean CO-MBA was 5.36 +/- 1.33 L/min, (4.95 +/- 0.80 L/min and 5.04 +/- 1.07 L/min in the arrhythmia subgroup). Mean of differences was 0.04 +/- 1.04 L/min with an error of 38.2%. In the arrhythmia subgroup, mean of differences was 0.14 +/- 0.90 L/min with an error of 35.4%. In the low CO subgroup, mean of differences was 0.26 +/- 0.89 L/min with an error of 40.4%. In adult patients after cardiac surgery, including those with low cardiac output and arrhythmia CO-MBA is not interchangeable with the continuous thermodilution method via a PAC, when using a 30% error threshold.
引用
收藏
页码:559 / 565
页数:7
相关论文
共 50 条
  • [1] Agreement between cardiac output estimation by multi-beat analysis of arterial blood pressure waveforms and continuous thermodilution in post cardiac surgery intensive care unit patients
    Ashish K. Khanna
    Lillian Nosow
    Lauren Sands
    Amit K. Saha
    Harshavardhan Agashe
    Lynnette Harris
    R. Shayn Martin
    Bryan Marchant
    Journal of Clinical Monitoring and Computing, 2023, 37 : 559 - 565
  • [2] Cardiac Output Estimation by Multi-Beat Analysis of Arterial Blood Pressure Waveform versus Continuous Pulmonary Artery Thermodilution in Post Cardiac Surgery Intensive Care Unit Patients
    Khanna, Ashish K.
    Sands, Lauren E.
    Nosow, Lillian M.
    Saha, Amit
    Marchant, Bryan E.
    ANESTHESIA AND ANALGESIA, 2021, 132 (5S_SUPPL): : 340 - 341
  • [3] Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients
    Khanna, Ashish K. K.
    Garcia, Julio O.
    Saha, Amit K.
    Harris, Lynnette
    Baruch, Martin
    Martin, R. Shayn
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2024, 38 (01) : 139 - 146
  • [4] Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients
    Ashish K. Khanna
    Julio O Garcia
    Amit K Saha
    Lynnette Harris
    Martin Baruch
    R. Shayn Martin
    Journal of Clinical Monitoring and Computing, 2024, 38 : 139 - 146
  • [5] Lack of agreement between bioimpedance and continuous thermodilution measurement of cardiac output in intensive care unit patients
    Ben N Barry
    Abhiram Mallick
    Andrew R Bodenham
    Michael Vucevic
    Critical Care, 1 (2):
  • [6] Cardiac output estimation by multi-beat analysis of the radial arterial blood pressure waveform versus intermittent pulmonary artery thermodilution: a method comparison study in patients treated in the intensive care unit after off-pump coronary artery bypass surgery
    Gillis Greiwe
    Valerie Peters
    Alexander Hapfelmeier
    Stefano Romagnoli
    Mathias Kubik
    Bernd Saugel
    Journal of Clinical Monitoring and Computing, 2020, 34 : 643 - 648
  • [7] Cardiac output estimation by multi-beat analysis of the radial arterial blood pressure waveform versus intermittent pulmonary artery thermodilution: a method comparison study in patients treated in the intensive care unit after off-pump coronary artery bypass surgery
    Greiwe, Gillis
    Peters, Valerie
    Hapfelmeier, Alexander
    Romagnoli, Stefano
    Kubik, Mathias
    Saugel, Bernd
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2020, 34 (04) : 643 - 648
  • [8] Lack of agreement between pulmonary arterial thermodilution cardiac output and the pressure recording analytical method in postoperative cardiac surgery patients
    Paarmann, H.
    Groesdonk, H. V.
    Sedemund-Adib, B.
    Hanke, T.
    Heinze, H.
    Heringlake, M.
    Schoen, J.
    BRITISH JOURNAL OF ANAESTHESIA, 2011, 106 (04) : 475 - 481
  • [9] Arterial Blood Pressure Waveform Based Cardiac Output Analysis Correlates with Continuous Pulmonary Artery Thermodilution In Post Cardiac Surgery ICU Patients Including Those With Arrhythmia
    Khanna, Ashish K.
    Nosow, Lillian M.
    Sands, Lauren E.
    Saha, Amit
    Agashe, Harsha
    Harris, Lynnette
    Marchant, Bryan E.
    ANESTHESIA AND ANALGESIA, 2022, 134 : 189 - 190
  • [10] Continuous cardiac output measurement: arterial pressure analysis versus thermodilution technique during cardiac surgery with cardiopulmonary bypass
    Lorsomradee, S.
    Lorsomradee, S. R.
    Cromheecke, S.
    De Hert, S. G.
    ANAESTHESIA, 2007, 62 (10) : 979 - 983