Device-device interaction between cardiac implantable electronic devices and continuous-flow left ventricular assist devices

被引:7
|
作者
Tanawuttiwat, Tanyanan [1 ,2 ]
Das, Mithilesh K. [1 ]
Miller, John M. [1 ]
Guglin, Maya E. [1 ]
机构
[1] Indiana Univ, Div Cardiovasc Med, Indianapolis, IN 46202 USA
[2] Indiana Univ, 1800 N Capitol Ave,E300B, Indian, IN 46202 USA
关键词
Cardiac implantable electronic device; Continuous-flow left ventricular assist device; Implantable cardioverter-defibrillator; Leadless pacemaker; Subcutaneous implantable cardioverter-defi-brillator; ELECTROMAGNETIC-INTERFERENCE; CARDIOVERTER-DEFIBRILLATOR; PATIENT; SYSTEM; LVAD;
D O I
10.1016/j.hrthm.2023.02.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The current design of an innovative left ventricular assist device (LVAD) makes use of magnetic levitation technology, which enables the rotors of the device to be completely suspended by magnetic force, reducing friction and blood or plasma damage. However, this electromagnetic field can result in electromagnetic interference (EMI), which can interfere with proper functioning of another car-diac implantable electronic device (CIED) in its direct proximity. Approximately 80% of patients with an LVAD have a CIED, most frequently an implantable cardioverter-defibrillator (ICD). Several device-device interactions have been reported, including EMI -induced inappropriate shocks, inability to establish telemetry connection, EMI-induced premature battery depletion, undersens-ing by the device, and other CIED malfunctions. Unfortunately, additional procedures, including generator exchange, lead adjust-ment, and system extraction, are frequently required because of these interactions. In some circumstances, the additional procedure might be preventable or avoidable with appropriate solutions. In this article, we describe how EMI from the LVAD impacts the func-tionality of the CIED and provide possible management options, including manufacturer-specific information, for the current CIEDs (eg, transvenous and leadless pacemakers, transvenous and subcu-taneous ICDs, and transvenous cardiac resynchronization therapy pacemakers and ICDs).
引用
收藏
页码:918 / 926
页数:9
相关论文
共 50 条
  • [1] Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review
    Gordon, Jonathan S.
    Maynes, Elizabeth J.
    O'Malley, Thomas J.
    Pavri, Behzad B.
    Tchantchaleishvili, Vakhtang
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 61 (01) : 1 - 10
  • [2] Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review
    Jonathan S. Gordon
    Elizabeth J. Maynes
    Thomas J. O’Malley
    Behzad B. Pavri
    Vakhtang Tchantchaleishvili
    Journal of Interventional Cardiac Electrophysiology, 2021, 61 : 1 - 10
  • [3] Cardiac Implantable Electronic Device Infections in Patients with Left Ventricular Assist Devices
    Faerber, G.
    Kirov, H.
    Doenst, T.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2020, 39 (04): : S490 - S490
  • [4] COMPLICATIONS OF CARDIAC IMPLANTABLE ELECTRONIC DEVICE INTERVENTION IN PATIENTS WITH LEFT VENTRICULAR ASSIST DEVICES
    Kiani, Soroosh
    Merchant, Faisal
    Hoskins, Michael
    Cole, Robert T.
    Black, George
    Lloyd, Michael
    Westerman, Stacy
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 485 - 485
  • [5] Complications of cardiac implantable electronic device intervention in patients with left ventricular assist devices
    Kiani, S.
    Black, G. B.
    Lloyd, M. S.
    Merchant, F. M.
    El-Chami, M. F.
    Cole, R. T.
    Hoskins, M. H.
    Westerman, S. B.
    EUROPEAN HEART JOURNAL, 2019, 40 : 1826 - 1826
  • [6] Incidence of Cardiac Implantable Electronic Device Complications in Patients With Left Ventricular Assist Devices
    Gilge, Jasen L.
    Sbircea, Nicole
    Tong, Michael Z.
    Bauza, Karolis
    Cantillon, Daniel J.
    Srivastava, Ajay
    Khedraki, Rola
    Sheikh, Farooq H.
    Ahmed, Sara
    Herr, Jared
    Hsiao, Stephanie
    Patel, Parin J.
    D'Souza, Benjamin
    Ravichandran, Ashwin K.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (04) : 494 - 501
  • [7] Surgical Outcomes of Centrifugal Continuous-Flow Implantable Left Ventricular Assist Devices: Heartmate 3 versus Heartware Ventricular Assist Device
    Shin, Kinam
    Cho, Won Chul
    Shin, Nara
    Kim, Hong Rae
    Kim, Min-Seok
    Chung, Cheol Hyun
    Jung, Sung-Ho
    JOURNAL OF CHEST SURGERY, 2024, 57 (02): : 184 - 194
  • [8] Device Thrombosis With Continuous-Flow Left Ventricular Assist Devices: Is Kounis Syndrome Involved?
    Kounis, Nicholas G.
    Soufras, George D.
    Tsigkas, Grigorios
    Hahalis, George
    ARTIFICIAL ORGANS, 2014, 38 (06) : 521 - 523
  • [9] Continuous-Flow Left Ventricular Assist Device
    Landzaat, Lindy H.
    Sinclair, Christian T.
    Rosielle, Drew A.
    NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (12): : 1149 - 1149
  • [10] Device thrombosis in continuous-flow left ventricular assist devices: A new manifestation of Kounis syndrome?
    Kounis, Nicholas G.
    Soufras, George D.
    Davlouros, Periklis