Development of a Monte Carlo-based scatter correction method for total-body PET using the uEXPLORER PET/CT scanner

被引:2
|
作者
Bayerlein, Reimund [1 ]
Spencer, Benjamin A. [2 ]
Leung, Edwin K. [3 ]
Omidvari, Negar [2 ]
Abdelhafez, Yasser G. [1 ]
Wang, Qian [2 ]
Nardo, Lorenzo [1 ]
Cherry, Simon R. [1 ,2 ]
Badawi, Ramsey D. [1 ,2 ]
机构
[1] Univ Calif Davis, Dept Radiol & Biomed Engn, Davis, CA 95616 USA
[2] Univ Calif Davis, Biomed Engn, Davis, CA USA
[3] UIH Amer, Houston, TX USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 04期
关键词
scatter correction (SC); total-body PET; Monte Carlo (MC) simulation; SIMULATION; SIMSET; RECONSTRUCTION; VALIDATION;
D O I
10.1088/1361-6560/ad2230
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. This study presents and evaluates a robust Monte Carlo-based scatter correction (SC) method for long axial field of view (FOV) and total-body positron emission tomography (PET) using the uEXPLORER total-body PET/CT scanner. Approach. Our algorithm utilizes the Monte Carlo (MC) tool SimSET to compute SC factors in between individual image reconstruction iterations within our in-house list-mode and time-of-flight-based image reconstruction framework. We also introduced a unique scatter scaling technique at the detector block-level for optimal estimation of the scatter contribution in each line of response. First image evaluations were derived from phantom data spanning the entire axial FOV along with image data from a human subject with a large body mass index. Data was evaluated based on qualitative inspections, and contrast recovery, background variability, residual scatter removal from cold regions, biases and axial uniformity were quantified and compared to non-scatter-corrected images. Main results. All reconstructed images demonstrated qualitative and quantitative improvements compared to non-scatter-corrected images: contrast recovery coefficients improved by up to 17.2% and background variability was reduced by up to 34.3%, and the residual lung error was between 1.26% and 2.08%. Low biases throughout the axial FOV indicate high quantitative accuracy and axial uniformity of the corrections. Up to 99% of residual activity in cold areas in the human subject was removed, and the reliability of the method was demonstrated in challenging body regions like in the proximity of a highly attenuating knee prosthesis. Significance. The MC SC method employed was demonstrated to be accurate and robust in TB-PET. The results of this study can serve as a benchmark for optimizing the quantitative performance of future SC techniques.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Low-Dose Imaging in a New Preclinical Total-Body PET/CT Scanner
    Molinos, Cesar
    Sasser, Todd
    Salmon, Phil
    Gsell, Willy
    Viertl, David
    Massey, James C.
    Minczuk, Krzysztof
    Li, Jie
    Kundu, Bijoy K.
    Berr, Stuart
    Correcher, Carlos
    Bahadur, Ali
    Attarwala, Ali A.
    Stark, Simon
    Junge, Sven
    Himmelreich, Uwe
    Prior, John O.
    Laperre, Kjell
    Van Wyk, Sonica
    Heidenreich, Michael
    [J]. FRONTIERS IN MEDICINE, 2019, 6
  • [32] Total-Body PET Scanner Prototype Due Next Year
    Abbasi, Jennifer
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2017, 318 (02): : 116 - 116
  • [33] Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation
    Zaidi, Habib
    Ay, Mohammad Reza
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2161 - 2165
  • [34] An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT
    Poludniowski, G.
    Evans, P. M.
    Hansen, V. N.
    Webb, S.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (12): : 3847 - 3864
  • [35] GPU-accelerated Monte Carlo Based Scatter Correction in Brain PET/MR
    Gaens, Michaela
    Bert, Julien
    Pietrzyk, Uwe
    Shah, N. Jon
    Visvikis, Dimitris
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [36] GPU-accelerated Monte Carlo based scatter correction in brain PET/MR
    Michaela Gaens
    Julien Bert
    Uwe Pietrzyk
    A Autret
    N Jon Shah
    Dimitris Visvikis
    [J]. EJNMMI Physics, 1 (Suppl 1)
  • [37] Monte Carlo Simulation for Scatter Correction in Brain PET/MRI based on GPU Acceleration
    Ma, B.
    Caldeira, L.
    Tellmann, L.
    Lohmann, P.
    Scheins, J.
    Kops, E.
    Xu, H.
    Lerche, C.
    Shah, N. J.
    Pietrzyk, U.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S505 - S505
  • [38] An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner
    Gao Fei
    Ryoko, Yamada
    Mitsuo, Watanabe
    Liu Hua-Feng
    [J]. CHINESE PHYSICS B, 2009, 18 (07) : 3066 - 3072
  • [39] An effective scatter correction method based on single scatter simulation for a 3D whole-body PET scanner
    高飞
    Yamada Ryoko
    Watanabe Mitsuo
    刘华锋
    [J]. Chinese Physics B, 2009, 18 (07) : 3066 - 3072
  • [40] Development and validation of a Monte Carlo model for a SiPM-based PET scanner
    Merlet, A.
    Presles, B.
    Su, K. H.
    Cochet, A.
    Vrigneaud, J. M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S345 - S345