Lu-177-PSMA dosimetry for kidneys and tumors based on SPECT images at two imaging time points

被引:6
|
作者
Chen, Gefei [1 ]
Lu, Zhonglin [1 ]
Jiang, Han [1 ,2 ]
Afshar-Oromieh, Ali [3 ]
Rominger, Axel [3 ]
Shi, Kuangyu [3 ]
Mok, Greta S. P. [1 ,4 ]
机构
[1] Univ Macau, Fac Sci & Technol, Dept Elect & Comp Engn, Biomed Imaging Lab BIG, Taipa, Macao, Peoples R China
[2] Fujian Med Univ, CT Ctr, PET, Union Hosp, Fuzhou, Fujian, Peoples R China
[3] Univ Bern, Bern Univ Hosp, Dept Nucl Med, Bern, Switzerland
[4] Univ Macau, Minist Educ, Frontiers Sci Ctr Precis Oncol, Taipa, Macao, Peoples R China
关键词
SPECT; Lu-177; PSMA; curve fitting; single time point; dosimetry; THERAPY;
D O I
10.3389/fmed.2023.1246881
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundPersonalized dosimetry for Lu-177-PSMA treatment requires multiple-time-point SPECT/CT scans to calculate time-integrated activity (TIA). This study evaluates two-time-point (TTP) methods for TIA calculation for kidneys and tumors.MethodsA total of 18 patients treated with 3.7-7.4 GBq Lu-177 PSMA-617 were analyzed retrospectively, including 18 sets of left and right kidneys, as well as 45 tumors. Four quantitative SPECT/CT (4TP) were acquired at 2 h, 20 h, 40 h, 60 h (n = 11), or 200 h (n = 7) after treatment, and they were fit bi-exponentially as reference. The TTP method was fitted by a mono-exponential washout function using two selected imaging time points for kidneys. For tumors, one uptake and one washout phase were modeled, assuming linear (type I) and same (type II) uptake phase between 0 h to the first time point and mono-exponential washout thereafter. Two single-time-point (STP) methods were also implemented for comparison. TIA calculated by TTP and STP methods were compared with reference to the 4TP TIA.ResultsFor the kidneys, the TTP methods using 20 h-60 h and 40 h-200 h had smaller mean absolute errors of 8.05 +/- 6.05% and 4.95 +/- 3.98%, respectively, as compared to other combinations of time points and STP methods. For tumors, the type I and type II TTP methods using 20h-60 h and 40-200 h had smaller mean absolute errors of 6.14 +/- 5.19% and 12.22 +/- 4.44%, and 8.31 +/- 7.16% and 4.48 +/- 7.10%, respectively, as compared to other TTP and STP methods.ConclusionThe TTP methods based on later imaging time demonstrated fewer errors than the STP methods in kidney and tumor TIA. Imaging at 20 h-60 h and 40 h-200 h could simplify the dosimetry procedures with fewer TIA estimation errors.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Simplified dosimetry for kidneys and tumors in 177Lu-labeled peptide receptor radionuclide therapy
    Oscar Ardenfors
    Joachim N. Nilsson
    Daniel Thor
    Cecilia Hindorf
    EJNMMI Physics, 9
  • [22] Harmonisation of Serial 177Lu-PSMA SPECT/CT Images in Patients with Metastatic Castrate-Resistant Prostate Cancer Treated with 177Lu-PSMA-617: A Prospective Multicentre 177Lu-PSMA SPECT/CT harmonisation (ENZA-p)
    Ayati, N.
    Hioki, T.
    van Oorschodt, J. C. J.
    McGill, G.
    Marquis, H.
    Francis, R.
    Bailey, D.
    Eslick, E.
    Sandhu, S.
    Davis, I. D.
    Stockler, M.
    Subramaniam, S.
    Crumbaker, M.
    Subhash, V.
    Sharma, S.
    Ayers, M.
    Jackson, P.
    Emmett, L.
    Willowson, K.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S280 - S281
  • [23] Quantitative 177Lu-SPECT/CT imaging and validation of a commercial dosimetry software
    D'ambrosio, L.
    Aloj, L.
    Madesani, D.
    Morisco, A.
    Aurilio, M.
    Prisco, A.
    Di Gennaro, F.
    Lastoria, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S263 - S263
  • [24] Mixed model time-activity fitting to reduce SPECT/CT imaging time points for dosimetry in Lu-177 peptide receptor radionuclide therapy (PRRT)
    Devasia, Theresa
    Dewaraja, Yuni
    Schipper, Matthew
    Wong, Ka Kit
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [25] Artifact-free partial volume correction based on spatially variant point spread function with non-negativity constrain for Lu-177-PSMA SPECT
    Liu, Yibin
    Chen, Gefei
    Lu, Zhonglin
    Shi, Kuangyu
    Mok, Greta
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [26] Technical Note: Optimal sampling schedules for kidney dosimetry based on the hybrid planar/SPECT method in 177Lu-PSMA therapy
    Rinscheid, Andreas
    Kletting, Peter
    Eiber, Matthias
    Beer, Ambros J.
    Glatting, Gerhard
    MEDICAL PHYSICS, 2019, 46 (12) : 5861 - 5866
  • [27] 3D Whole-Body SPECT images for dosimetry of 177Lu-PSMA treatment using a 360° CZT gamma camera
    Vergnaud, L.
    Giraudet, A.
    Paquet, E.
    Baudier, T.
    Badel, J.
    Sarrut, D.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S367 - S367
  • [28] Toward Single-Time-Point Image-Based Dosimetry of 177Lu-PSMA-617 Therapy
    Brosch-Lenz, Julia
    Delker, Astrid
    Voelter, Friederike
    Unterrainer, Lena M.
    Kaiser, Lena
    Bartenstein, Peter
    Ziegler, Sibylle
    Rahmim, Arman
    Uribe, Carlos
    Boening, Guido
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64 (05) : 767 - 774
  • [29] Validation of a simplified single time point image-based dosimetry approach for 177Lu-PSMA therapy
    Brosch-Lenz, J.
    Gosewisch, A.
    Voelter, F.
    Kaiser, L.
    Bartenstein, P.
    Ziegler, S.
    Rahmim, A.
    Uribe, C.
    Boening, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S73 - S74
  • [30] Lesion-wise dosimetry for multiple bone metastases in mCRPC patients undergoing Lu-177-PSMA therapy: comparison of five dosimetric methods
    Brosch, J.
    Uribe, C.
    Gosewisch, A.
    Kaiser, L.
    Bartenstein, P.
    Todica, A.
    Ilhan, H.
    Rahmim, A.
    Celler, A.
    Ziegler, S.
    Boening, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S378 - S379