Lu-177-PSMA dosimetry for kidneys and tumors based on SPECT images at two imaging time points

被引:6
|
作者
Chen, Gefei [1 ]
Lu, Zhonglin [1 ]
Jiang, Han [1 ,2 ]
Afshar-Oromieh, Ali [3 ]
Rominger, Axel [3 ]
Shi, Kuangyu [3 ]
Mok, Greta S. P. [1 ,4 ]
机构
[1] Univ Macau, Fac Sci & Technol, Dept Elect & Comp Engn, Biomed Imaging Lab BIG, Taipa, Macao, Peoples R China
[2] Fujian Med Univ, CT Ctr, PET, Union Hosp, Fuzhou, Fujian, Peoples R China
[3] Univ Bern, Bern Univ Hosp, Dept Nucl Med, Bern, Switzerland
[4] Univ Macau, Minist Educ, Frontiers Sci Ctr Precis Oncol, Taipa, Macao, Peoples R China
关键词
SPECT; Lu-177; PSMA; curve fitting; single time point; dosimetry; THERAPY;
D O I
10.3389/fmed.2023.1246881
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundPersonalized dosimetry for Lu-177-PSMA treatment requires multiple-time-point SPECT/CT scans to calculate time-integrated activity (TIA). This study evaluates two-time-point (TTP) methods for TIA calculation for kidneys and tumors.MethodsA total of 18 patients treated with 3.7-7.4 GBq Lu-177 PSMA-617 were analyzed retrospectively, including 18 sets of left and right kidneys, as well as 45 tumors. Four quantitative SPECT/CT (4TP) were acquired at 2 h, 20 h, 40 h, 60 h (n = 11), or 200 h (n = 7) after treatment, and they were fit bi-exponentially as reference. The TTP method was fitted by a mono-exponential washout function using two selected imaging time points for kidneys. For tumors, one uptake and one washout phase were modeled, assuming linear (type I) and same (type II) uptake phase between 0 h to the first time point and mono-exponential washout thereafter. Two single-time-point (STP) methods were also implemented for comparison. TIA calculated by TTP and STP methods were compared with reference to the 4TP TIA.ResultsFor the kidneys, the TTP methods using 20 h-60 h and 40 h-200 h had smaller mean absolute errors of 8.05 +/- 6.05% and 4.95 +/- 3.98%, respectively, as compared to other combinations of time points and STP methods. For tumors, the type I and type II TTP methods using 20h-60 h and 40-200 h had smaller mean absolute errors of 6.14 +/- 5.19% and 12.22 +/- 4.44%, and 8.31 +/- 7.16% and 4.48 +/- 7.10%, respectively, as compared to other TTP and STP methods.ConclusionThe TTP methods based on later imaging time demonstrated fewer errors than the STP methods in kidney and tumor TIA. Imaging at 20 h-60 h and 40 h-200 h could simplify the dosimetry procedures with fewer TIA estimation errors.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Lu-177-PSMA dosimetry based on two-imaging-time-point
    Chen, Gefei
    Lu, Zhonglin
    Jiang, Han
    Afshar-Oromieh, Ali
    Rominger, Axel
    Shi, Kuangyu
    Mok, Greta
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64
  • [2] Partial volume correction for Lu-177-PSMA SPECT
    Liu, Yibin
    Lu, Zhonglin
    Chen, Gefei
    Shi, Kuangyu
    Mok, Greta S. P.
    EJNMMI PHYSICS, 2024, 11 (01):
  • [3] Reproducibility of SPECT-based MTV for patients treated with Lu-177-PSMA
    Nuttens, V.
    Ooms, D.
    Vermeulen, S.
    De Bondt, P.
    Koole, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S209 - S209
  • [4] Lu-177-PSMA dosimetry using image-based TOPAS Monte Carlo simulations
    Lei, Bowen
    Yang, Fan
    Balaji, Vibha
    Luo, Jie
    Onecha, Victor Valladolid
    Heidari, Pedram
    Li, Quanzheng
    Bertolet, Alejandro
    Dutta, Joyita
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [5] Bone marrow dosimetry in low volume mHSPC patients receiving Lu-177-PSMA therapy using SPECT/CT
    Grob, Dagmar
    Prive, Bastiaan M.
    Muselaers, Constantijn H. J.
    Mehra, Niven
    Nagarajah, James
    Konijnenberg, Mark W.
    Peters, Steffie M. B.
    EJNMMI PHYSICS, 2024, 11 (01)
  • [6] Clinical dosimetry in [177Lu]Lu-PSMA therapy: fewer imaging time points, still useful
    De Vries-Huizing, D.
    Rijkhorst, E.
    Owers, E.
    De Wit-Van der Veen, B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S126 - S126
  • [7] Comparison of Red Marrow Dosimetry Methodologies for Lu-177-PSMA Radionuclide Therapy
    Lehnert, W.
    Apostolova, I.
    Buchert, R.
    Klutmann, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S76 - S77
  • [8] Bone marrow dosimetry in low volume hormone mHSPC receiving Lu-177-PSMA therapy using SPECT/CT
    Grob, D.
    Prive, B. M.
    Nagarajah, J.
    Konijnenberg, M. W.
    Peters, S. M. B.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S126 - S126
  • [9] Bone marrow dosimetry in low volume mHSPC patients receiving Lu-177-PSMA therapy using SPECT/CT
    Dagmar Grob
    Bastiaan M. Privé
    Constantijn H. J. Muselaers
    Niven Mehra
    James Nagarajah
    Mark W. Konijnenberg
    Steffie M. B. Peters
    EJNMMI Physics, 11
  • [10] Dosimetry for Lu-177-PSMA therapy of patients with mCRPC: A comparison of 3D SPECT and 2D planar methods
    Delker, A.
    Gosewisch, A.
    Rieger, S.
    Ilhan, H.
    Fendler, W.
    Gildehaus, F.
    Bartenstein, P.
    Boening, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S213 - S214