Patient-Specific Quality Assurance in Pencil Beam Scanning by 2-Dimensional Array

被引:1
|
作者
Rawiwan, Nuttida [1 ]
Chatchumnan, Nichakan [2 ]
Vimolnoch, Mananchaya [2 ]
Kingkaew, Sakda [2 ]
Oonsiri, Sornjarod [2 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Med Phys Program, Fac Med, Bangkok, Thailand
[2] King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, 1873 Rama 4 Rd, Bangkok 10330, Thailand
关键词
two-dimensional ionization chamber array; pencil beam scanning proton therapy; patient-specific quality assurance;
D O I
10.14338/IJPT-23-00016.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This study aimed to determine the characteristics of 2D ionization chamber array and the confidence limits of the gamma passing rate in pencil beam scanning proton therapy. Materials and Methods: The Varian ProBeam Compact spot-scanning system and the PTW OCTAVIUS 1500XDR array were used as a proton therapy system and detector, respectively. Our methods consisted of 2 parts: (1) the characteristics of the detector were tested and (2) patient-specific quality assurance was performed and evaluated by gamma analysis using dose-difference and distance-to-agreement criteria of 3% and 2 mm, respectively, with 123 treatment plans in head and neck, breast, chest, abdomen, and pelvic regions. Results: The PTW OCTAVIUS 1500XDR array had good reproducibility, uniformity, linearity, repetition rate, and monitor unit per spot within 0.1%, with accuracy, energy dependence, and measurement depth within 0.5%. The overall uncertainty of the PTW OCTAVIUS 1500XDR array was 2.49%. For field size and range shifter, using gamma analysis, the passing rate was 100%. The overall results of patient-specific quality assurance with the gamma evaluation were 98.9% 61.6% in 123 plans and confidence limit was 95.7%. Conclusion: The PTW OTAVIUS 1500XDR offered effective performance in pencil beam scanning proton therapy.
引用
收藏
页码:105 / 110
页数:6
相关论文
共 50 条
  • [1] Beam Parameters of Pediatric Plans with a Pencil Beam Scanning Proton System and Their Effects On Patient-Specific Quality Assurance
    Al-ward, S.
    Zhao, L.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E710 - E710
  • [2] Development of a three-dimensional scintillation detector for pencil beam verification in proton therapy patient-specific quality assurance
    Frelin, Anne-Marie
    Daviau, Gautier
    Bui, My Hoang Hoa
    Fontbonne, Cathy
    Fontbonne, Jean-Marc
    Lebhertz, Dorothee
    Mainguy, Erwan
    Moignier, Cyril
    Thariat, Juliette
    Vela, Anthony
    [J]. MEDICAL PHYSICS, 2024,
  • [3] Factors influencing the performance of patient specific quality assurance for pencil beam scanning IMPT fields
    Trnkova, P.
    Bolsi, A.
    Albertini, F.
    Weber, D. C.
    Lomax, A. J.
    [J]. MEDICAL PHYSICS, 2016, 43 (11) : 5998 - 6007
  • [4] Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc
    Popple, Richard A.
    Sullivan, Rodney J.
    Yuan, Yuan
    Wu, Xingen
    Covington, Elizabeth L.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 203 - 210
  • [5] Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance
    Lin, Liyong
    Kang, Minglei
    Solberg, Timothy D.
    Mertens, Thierry
    Baumer, Christian
    Ainsley, Christopher G.
    McDonough, James E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 270 - 276
  • [6] Analysis of ion beam teletherapy patient-specific quality assurance
    Liu, Xiaoli
    Deng, Yu
    Schlegel, Nicki
    Huang, Zhijie
    Moyers, Michael F.
    [J]. MEDICAL DOSIMETRY, 2019, 44 (01) : 43 - 50
  • [7] Analytic Dose Verification for Patient-Specific Proton Pencil Beam Scanning Plans
    Chang, C.
    Mah, D.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3359 - 3359
  • [8] 3D Gamma Test for Patient Specific Quality Assurance of Proton Pencil Beam Scanning Dose
    Pirelepesov, F.
    Yao, W.
    Ates, O.
    Lukose, R.
    Moskvin, V.
    Zhao, L.
    Farr, J.
    Merchant, T.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [9] Tornotherapy patient-specific quality assurance
    Chaudhari, S.
    Goddu, S.
    Mutic, S.
    Grigsby, J.
    Low, D.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2498 - 2499
  • [10] Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams
    Chang, Chang
    Poole, Kendra L.
    Teran, Anthony V.
    Luckman, Scott
    Mah, Dennis
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 381 - 388