Patient-Specific Quality Assurance in Pencil Beam Scanning by 2-Dimensional Array

被引:1
|
作者
Rawiwan, Nuttida [1 ]
Chatchumnan, Nichakan [2 ]
Vimolnoch, Mananchaya [2 ]
Kingkaew, Sakda [2 ]
Oonsiri, Sornjarod [2 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Med Phys Program, Fac Med, Bangkok, Thailand
[2] King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, 1873 Rama 4 Rd, Bangkok 10330, Thailand
关键词
two-dimensional ionization chamber array; pencil beam scanning proton therapy; patient-specific quality assurance;
D O I
10.14338/IJPT-23-00016.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This study aimed to determine the characteristics of 2D ionization chamber array and the confidence limits of the gamma passing rate in pencil beam scanning proton therapy. Materials and Methods: The Varian ProBeam Compact spot-scanning system and the PTW OCTAVIUS 1500XDR array were used as a proton therapy system and detector, respectively. Our methods consisted of 2 parts: (1) the characteristics of the detector were tested and (2) patient-specific quality assurance was performed and evaluated by gamma analysis using dose-difference and distance-to-agreement criteria of 3% and 2 mm, respectively, with 123 treatment plans in head and neck, breast, chest, abdomen, and pelvic regions. Results: The PTW OCTAVIUS 1500XDR array had good reproducibility, uniformity, linearity, repetition rate, and monitor unit per spot within 0.1%, with accuracy, energy dependence, and measurement depth within 0.5%. The overall uncertainty of the PTW OCTAVIUS 1500XDR array was 2.49%. For field size and range shifter, using gamma analysis, the passing rate was 100%. The overall results of patient-specific quality assurance with the gamma evaluation were 98.9% 61.6% in 123 plans and confidence limit was 95.7%. Conclusion: The PTW OTAVIUS 1500XDR offered effective performance in pencil beam scanning proton therapy.
引用
下载
收藏
页码:105 / 110
页数:6
相关论文
共 50 条
  • [21] Evaluation and Validation of IBA I'MatriXX Array for Patient-Specific Quality Assurance of TomoTherapy®
    Sresty, N. V. N. Madhusudhana
    Raju, A. Krishnam
    Reddy, B. Nagarjuna
    Sahithya, V. C.
    Mohmd, Yakub
    Kumar, G. Deleep
    Kumar, T. Anil
    Rushdi, Tasneem
    Aparna, S.
    Prasad, G. Durga
    Bajwa, Harjot Kaur
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) : 222 - 227
  • [22] Patient specific dosimetry and quality assurance for an active uniform scanning proton beam
    Mascia, A. E.
    Wolanski, M.
    Jesseph, R.
    Nichiporov, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S652 - S652
  • [23] The Root-Cause Analysis on Failed Patient-Specific Measurements of Pencil-Beam-Scanning Protons
    Hsi, W. C.
    Ricci, J.
    Su, Z.
    Mund, K.
    Dawson, R.
    Indelicato, D. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E519 - E519
  • [24] Patient-Specific Energy Verify Using the Multilayer Acrylic Disk Detector for Proton Pencil Beam Scanning
    Lah, J.
    Son, J.
    Moon, S.
    Shin, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E693 - E693
  • [25] Applications of a novel detector for pencil beam scanning proton therapy beam quality assurance
    Yang, Pei-Ying
    Hsieh, Yang-Wei
    Kang, Chen-Lin
    Tseng, Chin-Dar
    Lin, Chih-Hsueh
    Shieh, Chin-Shiuh
    Chao, Pei-Ju
    Lee, Tsair-Fwu
    INTERNATIONAL JOURNAL OF MODERN PHYSICS B, 2021, 35 (14N16):
  • [26] Log File-Based Patient Specific Quality Assurance: 5 Years of Experience in Pencil Beam Scanning Pediatric Proton Therapy
    Ates, O.
    Pirlepesov, F.
    Hua, C.
    Merchant, T. E.
    Zhao, L.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [27] Patient-Specific Quality Assurance Techniques for RapidArc Radiotherapy
    O'Daniel, J.
    McMahon, R.
    Chang, Z.
    Yan, H.
    Sakhalkar, H.
    Das, S.
    Wu, Q.
    Yin, F.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [28] Improvement in SRS Commission and Patient-Specific Quality Assurance
    Ding, M.
    Heard, D.
    Ma, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [29] Using an EPID for patient-specific VMAT quality assurance
    Bakhtiari, M.
    Kumaraswamy, L.
    Bailey, D. W.
    de Boer, S.
    Malhotra, H. K.
    Podgorsak, M. B.
    MEDICAL PHYSICS, 2011, 38 (03) : 1366 - 1373
  • [30] Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy
    Stathakis, Sotirios
    Myers, Pamela
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    MEDICAL PHYSICS, 2013, 40 (07)