Preload and friction in an implant-abutment-screw complex including a carbon-coated titanium alloy abutment screw: an in vitro study

被引:2
|
作者
Sagheb, K. [1 ]
Goergen, C. -I. [1 ]
Doell, S. [1 ]
Schmidtmann, I. [2 ]
Wentaschek, S. [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Univ Med Ctr, Dept Prosthodont & Mat Sci, Augustuspl 2, D-55131 Mainz, Germany
[2] Johannes Gutenberg Univ Mainz, Univ Med Ctr, Inst Med Biostat Epidemiol & Informat, Obere Zahlbacher Str 69, D-55131 Mainz, Germany
关键词
Dental implant; Abutment screw; Preload; Friction; TORQUE;
D O I
10.1186/s40729-023-00473-3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose An experimental approach was designed to measure the preload force, the coefficient of friction and the component of the tightening torque that is needed to surmount the thread-friction in an implant-abutment-screw complex that includes a carbon-coated screw. With the determined preload values the coefficient of friction was calculated. Methods 25 unused complexes, containing an implant, an abutment and a carbon-coated titanium alloy abutment screw, were tested. A custom load frame with two load cells and associated electronics was used. The threads were not lubricated. All abutment screws were torqued ten times to 25 Ncm. The produced preload values and a force that was proportional to the thread-friction component of the tightening torque were recorded. Results Mean preload values decreased significantly with the number of repetitions (p < 0.0001) from initially 329.9 N +/- 33.3 (range 255.7 to 383.9) to 253.7 N +/- 36.8 (range 200.1 to 332.5) for the last tightening procedure. The corresponding change in the calculated coefficient of friction was 0.33 +/- 0.04 (range 0.28 to 0.43) to 0.44 +/- 0.07 (range 0.32 to 0.56). For the thread-friction no corresponding trend for consecutive tightening repetitions could be noticed. Conclusions In the investigated implant-abutment units, repeated use of a coated abutment screw appears to increase the friction of the screw head and thereby decrease the preload. These results indicate that a pre-used coated implant-abutment-screw will fail reaching optimal screw preload.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Fatigue fracture of implant system using TiN and WC coated abutment screw
    Department of Dental Materials, School of Dentistry, Chosun University, Gwangju 501-759, Korea, Republic of
    不详
    Procedia Eng., (680-685):
  • [22] Fatigue Fracture of Implant System Using TiN and WC Coated Abutment Screw
    Choe, H. C.
    Lee, C. H.
    Jeong, Y. H.
    Ko, Y. M.
    Son, M. K.
    Chung, C. H.
    11TH INTERNATIONAL CONFERENCE ON THE MECHANICAL BEHAVIOR OF MATERIALS (ICM11), 2011, 10
  • [23] Survival of Four Conical Implant Abutment Connections After Removal of the Abutment Screw and Simulated Cyclic Loading: An In Vitro Comparative Study
    Yao, Kuang-Ta
    Chang, Tsai-Yu
    Huang, Guan-Jhong
    Fang, Hsu-Wei
    Wang, Ding-Han
    Hsu, Ming-Lun
    JOURNAL OF ORAL IMPLANTOLOGY, 2023, 49 (04) : 393 - 400
  • [24] The effect of the digital manufacturing technique of cantilevered implant-supported frameworks on abutment screw preload
    Altuwaijri, Shahad Mohammmed
    Alotaibi, Hanan Nejer
    Alnassar, Talal Mughaileth
    JOURNAL OF ADVANCED PROSTHODONTICS, 2022, 14 (01): : 22 - 31
  • [25] Effects of abutment screw preload in two implant connection systems: A 3D finite element study
    Jung, Won-kyung
    Lee, Won-sup
    Kwon, Ho-Beom
    JOURNAL OF PROSTHETIC DENTISTRY, 2019, 122 (05): : 474.e1 - 474.e8
  • [26] Influence of Diamondlike Carbon-Coated Screws on the Implant-Abutment Interface
    Vanoni Diez, Juliana Socas
    Brigagao, Vinicius Carvalho
    Cunha, Leonardo Goncalves
    Claro Neves, Ana Christina
    da Silva-Concilio, Lais Regiane
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (05) : 1055 - 1060
  • [27] In Vitro Study of Preload Loss in Different Implant Abutment Connection Designs
    Vinhas, Ana Sofia
    Aroso, Carlos
    Salazar, Filomena
    Relvas, Marta
    Braga, Ana Cristina
    Rios-Carrasco, Blanca
    Gil, Javier
    Rios-Santos, Jose Vicente
    Fernandez-Palacin, Ana
    Herrero-Climent, Mariano
    MATERIALS, 2022, 15 (04)
  • [28] Effect of Abutment Screw Design and Crown/Implant Ratio on Preload Maintenance of Single-Crown Screw-Retained Implant-Supported Prostheses
    Rodrigues, Isabela
    Zanardi, Piero
    Sesma, Newton
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2019, 34 (06) : 1397 - 1403
  • [29] Peri-implant bone alterations under the influence of abutment screw preload stress. A preclinical vivo study
    Al-Omari, Farah A.
    Nagasawa, Masako
    Suliman, Mubarak
    Khraisat, Ameen
    Uoshima, Katsumi
    CLINICAL ORAL IMPLANTS RESEARCH, 2020, 31 (12) : 1232 - 1242
  • [30] Analysis of Dimensional Changes in the Screw and the Surface Topography at the Interface of a Titanium Screw and a Zirconia Abutment Under Cyclic Loading: An In Vitro Study
    Dhingra, Ajay
    Weiner, Saul
    Luke, Allyn C.
    Ricci, John L.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (03) : 661 - 669