Sedation/agitation monitoring using a wearable device for ventilated children

被引:1
|
作者
Isaka, Kanako [1 ]
Hatachi, Takeshi [1 ,2 ]
Morita, Kanako [1 ]
Shimizu, Yoshiyuki [1 ]
Takeuchi, Muneyuki [1 ]
机构
[1] Osaka Womens & Childrens Hosp, Dept Intens Care Med, Izumi, Japan
[2] Osaka Womens & Childrens Hosp, Dept Intens Care Med, 840 Murodocho, Izumi, Osaka 5941101, Japan
关键词
actigraphy; deep sedation; pediatrics; physiologic monitoring; wearable electronic device; SEDATION; DELIRIUM; PROTOCOL; INFANTS; SLEEP;
D O I
10.1111/ped.15660
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: Developing continuous and labor-saving sedation/agitation monitoring methods in ventilated children is important to avoid undesirable events such as unplanned extubation. The existing scales are often challenging to use. We therefore aimed to evaluate the feasibility of sedation/agitation monitoring using a wearable device with a built-in accelerometer for ventilated children.Methods: This prospective observational pilot study included children aged 15 years or less, admitted to the pediatric intensive care unit on mechanical ventilation after cardiac catheterization between December 2021 and April 2022. The wearable device with a built-in accelerometer was attached to either of the upper limbs, and accelerations due to upper limb movements were measured for 2 h after admission or until extubation, whichever was earliest. Accelerations were measured at 0.02 s intervals, with the mean acceleration calculated for each 1 min interval. The State Behavioral Scale (SBS) was completed at 1 min intervals, with the SBS score (-1, 0, 1, or 2) compared with the mean acceleration.Results: The study included 20 children with a median age of 12 months. The mean accelerations and SBS scores were positively correlated (Kendall's tau, 0.22; p < 0.001), with an increase in the median (interquartile range) acceleration from an SBS score of -1 through 2, as follows: SBS -1, 0.200 (0.151-0.232) m/s(2); SBS 0, 0.202 (0.190-0.235) m/s(2); SBS, 1, 0.312 (0.236-0.427) m/s(2); SBS 2, 0.455 (0.332-0.517) m/s(2). No adverse events were observed.Conclusions: This study showed that continuous, labor-saving sedation/agitation monitoring of ventilated children was feasible using a wearable device with a built-in accelerometer.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Bispectral index correlates well with Richmond agitation sedation scale in mechanically ventilated critically ill patients
    Karamchandani, Kunal
    Rewari, Vimi
    Trikha, Anjan
    Batra, Ravinder Kumar
    JOURNAL OF ANESTHESIA, 2010, 24 (03) : 394 - 398
  • [42] Correlation between the Sedation-Agitation Scale and the Bispectral Index in ventilated patients in the intensive care unit
    Arbour, Richard
    Waterhouse, Julie
    Seckel, Maureen A.
    Bucher, Linda
    HEART & LUNG, 2009, 38 (04): : 336 - 345
  • [43] Bispectral index correlates well with Richmond agitation sedation scale in mechanically ventilated critically ill patients
    Kunal Karamchandani
    Vimi Rewari
    Anjan Trikha
    Ravinder Kumar Batra
    Journal of Anesthesia, 2010, 24 : 394 - 398
  • [44] A plea for enhanced monitoring of depth of sedation in patients who are intubated and ventilated
    Romagnoli, Stefano
    Matta, Basil
    Driver, Brian E.
    Evered, Lisbeth
    CRITICAL CARE, 2024, 28 (01)
  • [45] The impact of Bispectral Index monitoring on sedation administration in mechanically ventilated patients
    Weatherburn, C.
    Endacott, R.
    Tynan, P.
    Bailey, M.
    ANAESTHESIA AND INTENSIVE CARE, 2007, 35 (02) : 204 - 208
  • [46] EVALUATION OF SEDATION USING THE RICHMOND AGITATION SEDATION SCALE IN PATIENTS WITH PROLONGED MECHANICAL VENTILATION
    Raurell Torreda, M.
    Burgos Cordon, A.
    Conejo Caravaca, A.
    Corcuera Pastor, I.
    REVISTA ROL DE ENFERMERIA, 2015, 38 (7-8): : 533 - 538
  • [47] Wearable Auscultation Device Validation in Children
    Schmidt, H.
    Spielberg, D. R.
    Lakser, O. J.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [48] Games Applied to Children with Motor Impairment using the Myo Wearable Device
    Fernandes, Flavia G.
    Cardoso, Alexandre
    Lopes, Renato A.
    ANAIS DA ACADEMIA BRASILEIRA DE CIENCIAS, 2020, 92 (01):
  • [49] Monitoring sedation, agitation, analgesia, neuromuscular blockade, and delirium in adult ICU patients
    Riker, RR
    Fraser, GL
    SEMINARS IN RESPIRATORY AND CRITICAL CARE MEDICINE, 2001, 22 (02) : 189 - 198
  • [50] Signal Processing for a Wearable Device for Activity Monitoring
    Teodorescu, Horia-Nicolai L.
    Hagan, Marius
    2017 IEEE INTERNATIONAL CONFERENCE ON E-HEALTH AND BIOENGINEERING CONFERENCE (EHB), 2017, : 1 - 4