Intravascular imaging modalities in coronary intervention: Insights from 3D-printed phantom coronary models

被引:1
|
作者
Oliveira, Catarina [1 ]
Brito, Joana [1 ]
Rodrigues, Tiago [1 ]
Santiago, Helena [1 ]
Ricardo, Daniela [1 ]
Cardoso, Pedro [1 ,2 ]
Pinto, Fausto J. [1 ,2 ]
Marques, Joao Silva [1 ,2 ]
机构
[1] CHULN Hosp Santa Maria, Dept Coracao & Vasos, Serv Cardiol, Lisbon, Portugal
[2] Univ Lisbon, Ctr Cardiovasc Univ Lisboa CCUL RISE, Fac Med, Struct & Coronary Heart Dis Unit, Lisbon, Portugal
关键词
IVUS; OCT; Percutaneous coronary intervention; 3D-printing; OPTICAL-COHERENCE-TOMOGRAPHY; STENT IMPLANTATION; ULTRASOUND; IVUS;
D O I
10.1016/j.repc.2023.03.001
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction and objectives: Several studies comparing optical coherence tomography (OCT) and intravascular ultrasound (IVUS) have revealed that OCT consistently provides smaller area and diameter measurements. However, comparative assessment in clinical practice is difficult. Three-dimensional (3D) printing offers a unique opportunity to assess intravascular imaging modalities.We aim to compare intravascular imaging modalities using a 3D-printed coronary artery in a realistic simulator and to assess whether OCT underestimates intravascular dimensions, exploring potential corrections.Methods: A standard realistic left main anatomy with an ostial left anterior descending artery lesion was replicated using 3D printing. After provisional stenting and optimization, IVI was obtained. Modalities included 20 MHz digital IVUS, 60 MHz rotational IVUS (HD-IVUS) and OCT. We assessed luminal area and diameters at standard locations.Results: Considering all coregistered measurements, OCT significantly underestimated area, minimal diameter and maximal diameter measurements in comparison to IVUS and HD-IVUS (p<0.001). No significant differences were found between IVUS and HD-IVUS.A significant systematic dimensional error was found in OCT auto-calibration by comparing known reference diameter of guiding catheter (1.8 mm) to measured mean diameter (1.68 mm & PLUSMN;0.04 mm). By applying a correction factor based on the reference guiding catheter area to OCT, the luminal areas and diameters were not significantly different compared to IVUS and HD-IVUS.Conclusion: Our findings suggest that automatic spectral calibration method for OCT is inaccurate, with a systematic underestimation of luminal dimensions. When guiding catheter correction is applied the performance of OCT is significantly improved. These results may be clinically relevant and need to be validated.& COPY; 2023 Sociedade Portuguesa de Cardiologia. Published by Elsevier Espan & SIM;a, S.L.U. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/bync-nd/4.0/).
引用
收藏
页码:629 / 639
页数:11
相关论文
共 50 条
  • [21] Intravascular Imaging as a Performance Measure for Percutaneous Coronary Intervention
    Stein, Elliot J.
    Mesenbring, Elise
    Smith, Tracy
    Hebbe, Annika
    Salahuddin, Taufiq
    Waldo, Stephen W.
    Dyal, Michael D.
    Doll, Jacob A.
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2025, 18 (02)
  • [22] Optimization of Coronary Bifurcation Intervention Using Intravascular Imaging
    Ashley M. Darlington
    Muhammad Adnan Balouch
    Adhir R. Shroff
    Current Cardiovascular Imaging Reports, 2019, 12
  • [23] Intravascular imaging and coronary intervention: Is the bond getting stronger?
    Basavarajaiah, Sandeep
    Kalkat, Harkaran
    KARDIOLOGIA POLSKA, 2023, 81 (10) : 942 - 943
  • [24] Optimization of Coronary Bifurcation Intervention Using Intravascular Imaging
    Darlington, Ashley M.
    Balouch, Muhammad Adnan
    Shroff, Adhir R.
    CURRENT CARDIOVASCULAR IMAGING REPORTS, 2019, 12 (07)
  • [25] 3D printed coronary models offer new opportunities for developing optimal coronary CT angiography protocols in imaging coronary stents
    Sun, Zhonghua
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (08) : 1350 - 1355
  • [26] 3D-printed biological cell phantom for testing 3D quantitative phase imaging systems
    Michał Ziemczonok
    Arkadiusz Kuś
    Piotr Wasylczyk
    Małgorzata Kujawińska
    Scientific Reports, 9
  • [27] 3D-printed biological cell phantom for testing 3D quantitative phase imaging systems
    Ziemczonok, Michal
    Kus, Arkadiusz
    Wasylczyk, Piotr
    Kujawinska, Malgorzata
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [28] Radiopharmaceutical imaging based on 3D-CZT Compton camera with 3D-printed mouse phantom
    Tian, Feng
    Geng, Changran
    Yao, Zhiyang
    Wu, Renyao
    Xu, Jianfeng
    Cai, Fei
    Tang, Xiaobin
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 96 : 140 - 148
  • [29] Evaluating 3D-printed models of coronary anomalies: a survey among clinicians and researchers at a university hospital in the UK
    Lee, Matthew
    Moharem-Elgamal, Sarah
    Beckingham, Rylan
    Hamilton, Mark
    Manghat, Nathan
    Milano, Elena Giulia
    Bucciarelli-Ducci, Chiara
    Caputo, Massimo
    Biglino, Giovanni
    BMJ OPEN, 2019, 9 (03):
  • [30] A lesson from intravascular imaging: insights for recognizing a spontaneous coronary artery dissection
    Buccheri, Dario
    Milazzo, Diego
    Geraci, Salvatore
    Vaccaro, Giovanni
    Caramanno, Giuseppe
    JOURNAL OF THORACIC DISEASE, 2017, 9 (12) : 5363 - 5367