Multi-parametric optimization of magnetic resonance imaging sequences for magnetic resonance-guided radiotherapy

被引:0
|
作者
Fahad, Hafiz Muhammad [1 ,2 ,3 ]
Dorsch, Stefan [1 ,3 ,4 ]
Zaiss, Moritz [5 ,6 ]
Karger, Christian P. [1 ,3 ]
机构
[1] German Canc Res Ctr, Med Phys Radiat Oncol, Heidelberg, Germany
[2] Heidelberg Univ, Fac Med, Heidelberg, Germany
[3] Heidelberg Inst Radiat Oncol HIRO, Natl Ctr Radiat Res Oncol NCRO, Heidelberg, Germany
[4] Heidelberg Univ Hosp, Dept Radiat Oncol, Heidelberg, Germany
[5] Friedrich Alexander Univ Erlangen Nurnberg FAU, Univ Hosp Erlangen, Inst Neuroradiol, Erlangen, Germany
[6] Max Planck Inst Biol Cyberrnet, Magnet Resonance Ctr, Tubingen, Germany
关键词
Contrast maximization; MR-guided radiotherapy; Remote control MRI; Supervised learning; MR sequences optimization; MRI SCANNER; CANCER;
D O I
10.1016/j.phro.2023.100497
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Magnetic Resonance Imaging (MRI) is widely used in oncology for tumor staging, treatment response assessment, and radiation therapy (RT) planning. This study proposes a framework for automatic optimization of MRI sequences based on pulse sequence parameter sets (SPS) that are directly applied on the scanner, for application in RT planning.Materials and methods: A phantom with seven in-house fabricated contrasts was used for measurements. The proposed framework employed a derivative-free optimization algorithm to repeatedly update and execute a parametrized sequence on the MR scanner to acquire new data. In each iteration, the mean-square error was calculated based on the clinical application. Two clinically relevant optimization goals were pursued: achieving the same signal and therefore contrast as in a target image, and maximizing the signal difference (contrast) between specified tissue types. The framework was evaluated using two optimization methods: a covariance matrix adaptation evolution strategy (CMA-ES) and a genetic algorithm (GA). Results: The obtained results demonstrated the potential of the proposed framework for automatic optimization of MRI sequences. Both CMA-ES and GA methods showed promising results in achieving the two optimization goals, however, CMA-ES converged much faster as compared to GA.Conclusions: The proposed framework enables for automatic optimization of MRI sequences based on SPS that are directly applied on the scanner and it may be used to enhance the quality of MRI images for dedicated applications in MR-guided RT.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Characterization of kidneys in patients with systemic sclerosis by multi-parametric magnetic resonance quantitative imaging
    Tong, Xinyu
    He, Huilin
    Ning, Zihan
    Shen, Rui
    Du, Chenlin
    Zeng, Xiaofeng
    Wang, Qian
    He, Zuo-Xiang
    Xu, Dong
    Zhao, Xihai
    MAGNETIC RESONANCE IMAGING, 2024, 109 : 203 - 210
  • [42] A Multi-Parametric Investigation on Waterlogged Wood Using a Magnetic Resonance Imaging Clinical Scanner
    Longo, Sveva
    Egizi, Federica
    Stagno, Valeria
    Di Trani, Maria Giovanna
    Marchelletta, Gianni
    Gili, Tommaso
    Fazio, Enza
    Favero, Gabriele
    Capuani, Silvia
    FORESTS, 2023, 14 (02):
  • [43] Prostate cancer biomarkers: new scenarios in the multi-parametric magnetic resonance imaging era
    Porpiglia, Francesco
    De Luca, Stefano
    BJU INTERNATIONAL, 2017, 120 (06) : 745 - 746
  • [44] In room magnetic resonance imaging guided radiotherapy (MRIgRT)
    Lagendijk, J
    Raaymakers, B
    van der Heide, U
    Overweg, J
    Brown, K
    Bakker, C
    Raaijmakers, A
    Vulpen, M
    Welleweerd, J
    Jürgenliemk-Schulz, I
    MEDICAL PHYSICS, 2005, 32 (06) : 2067 - 2067
  • [45] Magnetic Resonance-Guided Focused Ultrasound Surgery
    Al Hilli, Mariam M.
    Stewart, Elizabeth A.
    SEMINARS IN REPRODUCTIVE MEDICINE, 2010, 28 (03) : 241 - 248
  • [46] Magnetic resonance-guided musculoskeletal interventional radiology
    Carrino, John A.
    Blanco, Roberto
    SEMINARS IN MUSCULOSKELETAL RADIOLOGY, 2006, 10 (02) : 159 - 173
  • [47] Magnetic Resonance-guided Active Catheter Tracking
    Wang, Wei
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2015, 23 (04) : 579 - +
  • [48] An efficient chemical shift imaging scheme for magnetic resonance-guided neurosurgery
    Liu, HY
    Hall, WA
    Martin, AJ
    Truwit, CL
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (01) : 1 - 7
  • [49] Magnetic resonance-guided focused ultrasound surgery
    Hudson, Susan B. A.
    Stewart, Elizabeth A.
    CLINICAL OBSTETRICS AND GYNECOLOGY, 2008, 51 (01): : 159 - 166
  • [50] Magnetic resonance-guided radiation therapy: A review
    Chin, Stephen
    Eccles, Cynthia L.
    McWilliam, Alan
    Chuter, Robert
    Walker, Emma
    Whitehurst, Philip
    Berresford, Joseph
    Van Herk, Marcel
    Hoskin, Peter J.
    Choudhury, Ananya
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2020, 64 (01) : 163 - 177