Interference of bovine hemoglobin-based oxygen carrier-201 (Hemopure) on four hematology analyzers

被引:1
|
作者
Bronkhorst-van Der Helm, Marinke W. [1 ]
Weerkamp, Floor [2 ]
Huisman, Albert [3 ]
Huisman, Elise J. [4 ,5 ]
Russcher, Henk [1 ,6 ]
机构
[1] Erasmus MC, Dept Clin Chem, Univ Med Ctr, Rotterdam, Netherlands
[2] Maasstad Hosp, Dept Clin Chem, Rotterdam, Netherlands
[3] Univ Med Ctr Utrecht, Cent Diagnost Lab, Utrecht, Netherlands
[4] Erasmus MC, Sophia Childrens Hosp, Dept Pediat Hematol, Rotterdam, Netherlands
[5] Sanquin Blood Bank, Unit Transfus Med, Amsterdam, Netherlands
[6] Erasmus MC, Dept Clin Chem, Dr Molewaterplein 40, NL-3015 GD Rotterdam, Netherlands
关键词
extracellular hemoglobin; hematology analyzer; hemoglobin-based oxygen carrier; interference; intracellular hemoglobin; HBOC-201;
D O I
10.1111/ijlh.14146
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Hemoglobin-based oxygen carriers, for example HBOC-201 (Hemopure), are aimed to bridge acute anemia when blood transfusion is not available or refused by the patient. However, since HBOC-201 appears free in plasma, it interferes with laboratory tests. This study presents an overview of HBOC-201 interference on four commonly used hematology analyzers and suggests treatment monitoring possibilities.Methods: Blood samples were spiked with therapeutic doses of HBOC-201 and nine hematology parameters were measured with the Sysmex XN-20, Siemens Advia 2120i, Abbott Alinity Hq and Abbot Cell Dyn Sapphire hematology analyzers. The results were compared to control samples and the bias was determined.Results: Most parameters, including all cell counts, hematocrit and MCV, showed a non-significant bias compared to control. However, the standard, total hemoglobin (Hb) measurement as well as MCH and MCHC showed poor agreement with control, as HBOC-201 was included in this measurement. Yet, the flow cytometry-based Hb method quantified intracellular Hb in spiked samples, excluding HBOC-201.Conclusion: Of all included hematology parameters, only total Hb and the associated MCH and MCHC suffered from interference. In contrast, the flow cytometry-based Hb measurement provided an accurate measure of intracellular Hb. The difference between total Hb and cellular Hb represents the HBOC-201 concentration and can be used to monitor HBOC-201 treatment.
引用
收藏
页码:869 / 874
页数:6
相关论文
共 50 条
  • [41] Hemoglobin-Based Oxygen Carrier and Convection Enhanced Oxygen Transport in a Hollow Fiber Bioreactor
    Chen, Guo
    Palmer, Andre F.
    BIOTECHNOLOGY AND BIOENGINEERING, 2009, 102 (06) : 1603 - 1612
  • [42] Hematology patterns after hemoglobin-based oxygen carrier resuscitation from severe controlled hemorrhage with prolonged delayed definitive care
    Arnaud, Francoise
    Fasipe, Daniel
    Philbin, Nora
    Rice, Jennifer
    Flournoy, W. Shannon
    Ahlers, Stephen
    McCarron, Richard
    Freilich, Daniel
    TRANSFUSION, 2007, 47 (11) : 2098 - 2109
  • [43] In vitro effects of a novel hemoglobin-based oxygen carrier on routine chemistry, therapeutic drug, coagulation, hematology, and blood bank assays
    Callas, DD
    Clark, TL
    Moreira, PL
    Lansden, C
    Gawryl, MS
    Kahn, S
    Bermes, EW
    CLINICAL CHEMISTRY, 1997, 43 (09) : 1744 - 1748
  • [44] RESUSCITATION WITH THE HEMOGLOBIN-BASED OXYGEN CARRIER, HBOC-201, IN A SWINE MODEL OF SEVERE UNCONTROLLED HEMORRHAGE AND TRAUMATIC BRAIN INJURY
    Stern, Susan
    Rice, Jennifer
    Philbin, Nora
    McGwin, Gerald
    Arnaud, Francoise
    Johnson, Todd
    Flournoy, W. Shannon
    Ahlers, Stephen
    Pearce, L. Bruce
    McCarron, Richard
    Freilich, Daniel
    SHOCK, 2009, 31 (01): : 64 - 79
  • [45] Oxygen transport by erythrocyte/hemoglobin solution mixtures in an in vitro capillary as a model of hemoglobin-based oxygen carrier performance
    Page, TC
    Light, WR
    McKay, CB
    Hellums, JD
    MICROVASCULAR RESEARCH, 1998, 55 (01) : 54 - 64
  • [46] Transfusion requirements in phase III study of HBOC-201 (hemopure®, hemoglobin glutamer-250, bovine), a bovine hemoglobin based blood substitute.
    Williams, JP
    Tseuda, K
    Vandermeersch, E
    Bourke, DL
    Jahr, JS
    TRANSFUSION, 2002, 42 (09) : 14S - 14S
  • [47] BIOT 65-Polymersome encapsulated hemoglobin: A novel type of hemoglobin-based oxygen carrier
    Palmer, Andre F.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2007, 234
  • [48] Blood Component Requirements and Erythrocyte Transfusion and Mortality Related to Hemoglobin Deficit in Phase III Trial of Hemoglobin-Based Oxygen Carrier: HBOC-201
    Jahr, Jonathan S.
    Williams, John P.
    AMERICAN JOURNAL OF THERAPEUTICS, 2022, 29 (03) : E279 - E286
  • [49] Extracorporeal membrane oxygenation in piglets using a polymerized bovine hemoglobin-based oxygen-carrying solution (HBOC-201)
    York, GB
    DiGeronimo, RJ
    Wilson, BJ
    Cofer, BR
    Breuer, CK
    Josephs, JD
    Smith, DL
    Sorrells, DL
    JOURNAL OF PEDIATRIC SURGERY, 2002, 37 (10) : 1387 - 1392
  • [50] Use of a hemoglobin-based oxygen carrier as an adjunct in a therapeutic regimen for severe anemia
    Shander, A
    Lui, J
    Alalawi, R
    Bennett, H
    ANESTHESIOLOGY, 2003, 99 (03) : B52 - B52