Cone beam computed tomography (CBCT); paranasal sinuses; sinusitis; radiation dosage; nasal surgical procedures; BEAM COMPUTED-TOMOGRAPHY; CT; MULTIDETECTOR; HEAD; RISK;
D O I
10.4193/Rhin22.385
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Background: This study evaluates the clinical image quality (IQ) and usability of a sinonasal ultra-low-dose (ULD) cone-beam computed tomography (CBCT) scan. The results are compared to those of a high resolution (HR) CBCT scan to identify the streng-ths and weaknesses of a ULD CBCT protocol.Methodology: Sixty-six anatomical sites in 33 subjects were imaged twice using two imaging modalities: HR CBCT (Scanora 3Dx scanner; Soredex, Tuusula, Finland) and ULD CBCT (Promax 3D Mid scanner; Plandent, Helsinki, Finland). IQ, opacification and obstruction, structural features and operative usability were assessed.Results: The overall IQ in subjects with "no or minor opacification" was excellent: 100% (HR CBCT) and 99% (ULD CBCT) of ratings were evaluated as sufficient for every structure. Increased opacification reduced the quality of both imaging modalities, resulting in sufficient IQ in 97% (HR CBCT) and 92% (ULD CBCT) of evaluations. With greater opacification, the sufficient-insufficient ratio was better (p< 0.05) for HR CBCT regarding the anterior ethmoidal artery, lateral lamella, frontal recess, cribriform plate, superior concha and posterior ethmoidal artery. ULD CBCT's operative applicability was generally good with moderate restraints in fron-toethmoidectomy, frontal sinusotomy, sphenotomy and posterior ethmoidectomy in cases with greater opacification.Conclusions: IQ of paranasal ULD CBCT is sufficient for clinical diagnostics and should be considered for surgical planning. We recommend it as the primary imaging protocol for all patients who meet imaging criteria due to recurrent or chronic nasal symptoms. Additional or conventional imaging might be needed for patients with extensive chronic rhinosinusitis and/or indicati-ons of frontal sinus involvement.
引用
收藏
页码:221 / 230
页数:10
相关论文
共 50 条
  • [1] Evaluation of the Image Quality of Ultra-Low-Dose Paranasal Sinus CT
    Aksamoglu, Melih
    Menzilcioglu, Mehmet Sait
    EUROPEAN JOURNAL OF THERAPEUTICS, 2023, 29 (02): : 143 - 148
  • [2] Reliability of High-Pitch Ultra-Low-Dose Paranasal Sinus Computed Tomography for Evaluating Paranasal Sinus Anatomy and Sinus Disease
    Aksoy, Elif Ayanoglu
    Ozden, Sila Ulus
    Karaarslan, Ercan
    Unal, Omer Faruk
    Tanyeri, Hasan
    JOURNAL OF CRANIOFACIAL SURGERY, 2014, 25 (05) : 1801 - 1804
  • [3] Ultra-Low-Dose Photon-Counting CT Imaging of the Paranasal Sinus With Tin Prefiltration How Low Can We Go?
    Grunz, Jan-Peter
    Petritsch, Bernhard
    Luetkens, Karsten Sebastian
    Kunz, Andreas Steven
    Lennartz, Simon
    Ergun, Suleyman
    Bley, Thorsten Alexander
    Huflage, Henner
    INVESTIGATIVE RADIOLOGY, 2022, 57 (11) : 728 - 733
  • [4] Tin-filtered 100 kV ultra-low-dose CT of the paranasal sinus: Initial clinical results
    Petritsch, Bernhard
    Kosmala, Aleksander
    Weng, Andreas Max
    Bley, Thorsten Alexander
    PLOS ONE, 2019, 14 (05):
  • [5] Ultra-low-dose CBCT scan: rational map for ear surgery
    Tamminen, Pekka
    Jarnstedt, Jorma
    Lehtinen, Antti
    Numminen, Jura
    Lehtimaki, Lauri
    Rautiainen, Markus
    Kivekas, Ilkka
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2023, 280 (03) : 1161 - 1168
  • [6] Ultra-low-dose CBCT scan: rational map for ear surgery
    Pekka Tamminen
    Jorma Järnstedt
    Antti Lehtinen
    Jura Numminen
    Lauri Lehtimäki
    Markus Rautiainen
    Ilkka Kivekäs
    European Archives of Oto-Rhino-Laryngology, 2023, 280 : 1161 - 1168
  • [7] Application of a full model-based iterative reconstruction (MBIR) in 80 kVp ultra-low-dose paranasal sinus CT imaging of pediatric patients
    Jihang Sun
    Qifeng Zhang
    Xiaomin Duan
    Chengyue Zhang
    Pengpeng Wang
    Chenguang Jia
    Yong Liu
    Yun Peng
    La radiologia medica, 2018, 123 : 117 - 124
  • [8] Application of a full model-based iterative reconstruction (MBIR) in 80 kVp ultra-low-dose paranasal sinus CT imaging of pediatric patients
    Sun, Jihang
    Zhang, Qifeng
    Duan, Xiaomin
    Zhang, Chengyue
    Wang, Pengpeng
    Jia, Chenguang
    Liu, Yong
    Peng, Yun
    RADIOLOGIA MEDICA, 2018, 123 (02): : 117 - 124
  • [9] Comparison of CBCT and CT in Terms of Dose Value of Organs at Risk In Paranasal Sinus Imaging
    Pak, Farideh
    Pooya, Seyed Mahdi Hosseini
    Shabani, Hamed
    Baradaran, Samaneh
    Sadeq, Ula Salim
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (04) : 208 - 213
  • [10] ULTRA-LOW-DOSE HEPARIN
    不详
    LANCET, 1980, 1 (8174): : 907 - 908