Comparison of Corneal Higher-order Aberrations Measured by Scheimpflug Camera and Placido Disc-based Topography in Korean Patients

被引:0
|
作者
Lim, Yeon Ju [1 ]
Jung, Do Hee [1 ]
Pak, Kang Yeun [1 ]
Cho, Chan-Ho [1 ,2 ]
机构
[1] Inje Univ, Haeundae Paik Hosp, Dept Ophthalmol, Coll Med, Pusan, South Korea
[2] Inje Univ, Coll Med, Haeundae Paik Hosp, Dept Ophthalmol, 875 Haeun Daero, Busan 48108, South Korea
来源
关键词
Corneal higher-order aberrations; iTrace (TM); Pentacam (R) HR; STATISTICAL-METHODS; PRINCIPLES; THICKNESS; EYES;
D O I
10.3341/jkos.2023.64.12.1141
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: We compared corneal higher-order aberration (HOA) measurements between the Pentacam (R) HR (PC), which uses the Scheimpflug principle, and the iTrace (TM) aberrometer (IT), which evaluates Placido disc topography.Methods: We retrospectively analyzed 109 eyes of 87 patients without a history of ocular surface disease during the period from January 2021 to December 2022; both devices were used on the same day. We calculated the root mean square values (mu m) of corneal total HOA and of the corneal 3rd-and 4th-order HOA at a pupil diameter of 4 mm. Data were compared by Bland-Altman plots for agreement analysis and Pearson correlation coefficient for correlation analysis.Results: There was no significant difference in the total HOA (p = 0.145), coma (p = 0.309), or secondary astigmatism (p = 0.080) between the PC and IT measurements; all other HOAs significantly differed between devices (p < 0.001). In the correlation analysis, the total HOA (r = 0.605, p < 0.001) and coma (r = 0.634, p < 0.001) were moderately correlated between the two devices; the other HOAs showed low degree of correlations. In the Bland-Altman plot analysis, all HOAs showed low agreement between the two devices.Conclusions: Corneal total HOA and coma measured by the two devices were significantly correlated between the two devices, but other HOAs showed significant differences in measurement and low correlations. Therefore, corneal HOA measurements cannot be interpreted interchangeably between the two devices.
引用
收藏
页码:1141 / 1151
页数:11
相关论文
共 50 条
  • [41] Comparison of corneal aberrations from anterior segment swept source OCT versus Placido-topography combined spectral domain OCT in cataract patients
    Georgiev, Stefan
    Ruiss, Manuel
    Dana-Fisus, Andreea
    Leitgeb, Rainer A.
    Findl, Oliver
    EYE AND VISION, 2023, 10 (01)
  • [42] Response to: WaveLight (R) Contoura topography-guided planning: contribution of anterior corneal higher-order aberrations and posterior corneal astigmatism to manifest refractive astigmatism Reply
    Wallerstein, Avi
    Gauvin, Mathieu
    Cohen, Mark
    CLINICAL OPHTHALMOLOGY, 2018, 12 : 2003 - 2004
  • [43] Erratum to: Effects of Reference Axes Used During Measurements of Ocular and Corneal Higher-Order Aberrations in Patients Following LASIK
    Atsuko Takehara
    Naoyuki Maeda
    Sayuri Ninomiya
    Takashi Fujikado
    Yoko Hiroha
    Toshifumi Mihashi
    Japanese Journal of Ophthalmology, 2006, 50 : 489 - 489
  • [44] Residual higher-order aberrations caused by clinically measured cyclotorsional misalignment or decentration during wavefront-guided excimer laser corneal ablation
    Wang, Li
    Koch, Douglas D.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (12): : 2057 - 2062
  • [45] Toric Artisan after transepithelial topography-guided photorefractive keratectomy for higher-order aberrations following intrastromal corneal ring segments in keratoconus - Trioptics
    Ferreira, Gabriel de Almeida
    Ghanem, Vinicius Coral
    de Pinho Tavares, Renata Leite
    Ghanem, Ramon Coral
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2020, 68 (11) : 2564 - +
  • [46] Long-term changes in contrast-sensitivity, corneal topography and higher-order aberrations after upper eyelid blepharoplasty: A prospective interventional study
    Bhattacharjee, Kasturi
    Misra, Divakant
    Singh, Manpreet
    Deori, Nilutparna
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2020, 68 (12) : 2906 - 2910
  • [47] Comparison of Corneal Surface Higher-Order Aberrations After Endothelial Keratoplasty, Femtosecond Laser-Assisted Keratoplasty, and Conventional Penetrating Keratoplasty
    Chamberlain, Winston
    Omid, Nika
    Lin, Amy
    Farid, Marjan
    Gaster, Ronald N.
    Steinert, Roger F.
    CORNEA, 2012, 31 (01) : 6 - 13
  • [48] Changes in Corneal Higher-Order Aberrations and Ocular Biometric Measurements after Phacoemulsification Combined with Goniosynechialysis in Primary Angle Closure/Glaucoma Patients
    Xia, Jiali
    Guo, Siqi
    Hu, Fei
    Fan, Liqi
    Yu, Ling
    Ye, Jian
    JOURNAL OF OPHTHALMOLOGY, 2024, 2024
  • [49] Corneal Higher-Order Aberrations in Descemet Membrane Endothelial Keratoplasty versus Ultrathin DSAEK in the Descemet Endothelial Thickness Comparison Trial A Randomized Clinical Trial
    Duggan, Matthew J.
    Rose-Nussbaumer, Jennifer
    Lin, Charles C.
    Austin, Ariana
    Labadzinzki, Paula C.
    Chamberlain, Winston D.
    OPHTHALMOLOGY, 2019, 126 (07) : 946 - 957
  • [50] Effects of reference axes used during measurements of ocular and corneal higher-order aberrations in patients following LASIK (vol 50, pg 318, 2006)
    Takehara, Atsuko
    Maeda, Naoyuki
    Ninomiya, Sayuri
    Fujikado, Takashi
    Hirohara, Yoko
    Mihashi, Toshifumi
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2006, 50 (05) : 489 - 489