Comparison of Corneal Higher-order Aberrations Measured by Scheimpflug Camera and Placido Disc-based Topography in Korean Patients

被引:0
|
作者
Lim, Yeon Ju [1 ]
Jung, Do Hee [1 ]
Pak, Kang Yeun [1 ]
Cho, Chan-Ho [1 ,2 ]
机构
[1] Inje Univ, Haeundae Paik Hosp, Dept Ophthalmol, Coll Med, Pusan, South Korea
[2] Inje Univ, Coll Med, Haeundae Paik Hosp, Dept Ophthalmol, 875 Haeun Daero, Busan 48108, South Korea
来源
关键词
Corneal higher-order aberrations; iTrace (TM); Pentacam (R) HR; STATISTICAL-METHODS; PRINCIPLES; THICKNESS; EYES;
D O I
10.3341/jkos.2023.64.12.1141
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: We compared corneal higher-order aberration (HOA) measurements between the Pentacam (R) HR (PC), which uses the Scheimpflug principle, and the iTrace (TM) aberrometer (IT), which evaluates Placido disc topography.Methods: We retrospectively analyzed 109 eyes of 87 patients without a history of ocular surface disease during the period from January 2021 to December 2022; both devices were used on the same day. We calculated the root mean square values (mu m) of corneal total HOA and of the corneal 3rd-and 4th-order HOA at a pupil diameter of 4 mm. Data were compared by Bland-Altman plots for agreement analysis and Pearson correlation coefficient for correlation analysis.Results: There was no significant difference in the total HOA (p = 0.145), coma (p = 0.309), or secondary astigmatism (p = 0.080) between the PC and IT measurements; all other HOAs significantly differed between devices (p < 0.001). In the correlation analysis, the total HOA (r = 0.605, p < 0.001) and coma (r = 0.634, p < 0.001) were moderately correlated between the two devices; the other HOAs showed low degree of correlations. In the Bland-Altman plot analysis, all HOAs showed low agreement between the two devices.Conclusions: Corneal total HOA and coma measured by the two devices were significantly correlated between the two devices, but other HOAs showed significant differences in measurement and low correlations. Therefore, corneal HOA measurements cannot be interpreted interchangeably between the two devices.
引用
收藏
页码:1141 / 1151
页数:11
相关论文
共 50 条
  • [21] Comparison of Corneal Higher-order Aberrations between Miniscleral and Hybrid Lenses in Keratoconus
    Assadpour, Mehrdad
    Nabovati, Payam
    Hashemi, Hassan
    Rafati, Shokoofeh
    Yekta, Abbasali
    Jafarzadehpur, Ebrahim
    Khabazkhoob, Mehdi
    OPTOMETRY AND VISION SCIENCE, 2020, 97 (09) : 749 - 753
  • [22] Characteristics of ocular higher-order aberrations in patients with pellucid marginal corneal degeneration
    Oie, Yoshinori
    Maeda, Naoyuki
    Kosaki, Ryo
    Suzaki, Asaki
    Hirohara, Yoko
    Mihashi, Toshifurni
    Hori, Yuichi
    Inoue, Tornoyuki
    Nishida, Kohji
    Fujikado, Takashi
    Tano, Yasuo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (11): : 1928 - 1934
  • [23] Corneal higher-order aberrations after phacoemulsification: a comparison of 3 different incision sizes
    von Sonnleithner, Christoph
    Pilger, Daniel
    Homburg, David
    Brockmann, Tobias
    Torun, Necip
    Bertelmann, Eckart
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2017, 27 (04) : 402 - 406
  • [24] Comparison between color light-emitting diode corneal topographer and scheimpflug imager in measuring corneal astigmatism and corneal higher order aberrations
    Hwang, Yousook
    Kim, Da Yeoung
    Chung, So-Hyang
    ACTA OPHTHALMOLOGICA, 2019, 97
  • [25] Anterior corneal optical irregularity measured by higher-order aberrations induced by a broad beam excimer laser
    Hsieh, Yi-Ting
    Wang, I-Jong
    Hu, Fung-Rong
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2012, 95 (05) : 522 - 530
  • [26] Comparison of Corneal Higher-Order Aberrations Following Topography-Guided LASIK and SMILE for Myopic Correction: A Propensity Score Matching Analysis
    Kang, Eun Min
    Ryu, Ik Hee
    Lee, In Sik
    Kim, Jin Kuk
    Kim, Sun Woong
    Ji, Yong Woo
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (20)
  • [27] Higher-Order Aberrations of Anterior and Posterior Corneal Surfaces in Patients With Keratectasia After LASIK
    Maeda, Naoyuki
    Nakagawa, Tomoya
    Kosaki, Ryo
    Koh, Shizuka
    Saika, Makoto
    Fujikado, Takashi
    Nishida, Kohji
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (06) : 3905 - 3911
  • [28] Effect of axes of measurements on ocular and corneal higher-order aberrations in patients following LASIK
    Takada, A
    Maeda, N
    Kuroda, T
    Ninomiya, S
    Fujikado, T
    Tano, Y
    Hirohara, Y
    Mihashi, T
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U3 - U3
  • [29] The Impact of Corneal Higher Order Aberrations on the Discrepancy Between Manifest Refractive Astigmatism and Topography-Measured Anterior Corneal Astigmatism in Healthy Candidates
    Wallerstein, Avi
    Ridgway, Chelsea
    Gauvin, Mathieu
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (09)
  • [30] Comparison of total, spherical, and high order aberrations from nidek OPD-Scan, ZyWave, and corneal data derived from placido topography
    Naroo, SA
    Patel, H
    Oudit, D
    Rundle, K
    JOURNAL OF REFRACTIVE SURGERY, 2004, 20 (05) : S758 - S758