Determination of an air kerma-rate correction factor for the S7600 Xoft Axxent ® source model

被引:0
|
作者
Walter, Autumn E. [1 ,2 ]
DeWerd, Larry A. [1 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI 53792 USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Human Oncol, 600 Highland Ave, Madison, WI 53792 USA
关键词
Air-kerma; Xoft; Electronic brachytherapy; Standard; Calibration; WELL CHAMBER RESPONSE; AMBIENT-PRESSURE;
D O I
10.1016/j.brachy.2023.02.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work was to provide guidance for the lack of an air-kerma rate standard for the S7600 Xoft Axxent & REG; source by providing a correction factor to apply to the National Institute of Standards and Technology (NIST) traceable S7500 well chamber (WC) calibration coefficient before the development of an S7600 standard at NIST.METHODS AND MATERIALS: The Attix free air chamber (FAC) at the University of Wis-consin Medical Radiation Research Center was used to measure the air-kerma rate at 50 cm for six S7500 and six S7600 sources. These same sources were then measured using five stan-dard imaging HDR1000+ WCs. The measurements made with the FAC were used to calculate source-specific WC calibration coefficients for the S7500 and S7600 source. These results were compared to the NIST traceable calibration coefficients for the S7500 source. The average results for each WC were then averaged together, and a ratio of the S7600 to S7500 WC calibration coefficients was determined. RESULTS: The average S7600 air-kerma rate measurement with the FAC was 7% lower than the average air-kerma rate measurements of the S7500 source. On average, the S7500 determined WC calibration coefficients agreed within & PLUSMN;1% of the NIST traceable S7500 values. The S7600 WC calibration coefficients were up to 16% less than the NIST traceable S7500 values. The final correction factor determined to be applied to the NIST traceable S7500 value was 0.8415 with an associated uncertainty of & PLUSMN;8.1% at k = 2.CONCLUSIONS: This work provides a suggested correction factor for the S7600 Xoft Axxent source such that the sources can be accurately implemented in the clinical setting. & COPY; 2023 The Authors. Published by Elsevier Inc. on behalf of American Brachytherapy Society. This is an open access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ )
引用
收藏
页码:512 / 517
页数:6
相关论文
共 14 条
  • [1] Comparison of air kerma rate between the S7500 and S7600 xoft axxent sources
    Walter, Autumn E.
    Hull, Jennifer L.
    DeWerd, Larry A.
    [J]. BRACHYTHERAPY, 2022, 21 (03) : 354 - 361
  • [2] Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model
    Walter, Autumn E.
    Khan, Ahtesham U.
    DeWerd, Larry A.
    [J]. BRACHYTHERAPY, 2023, 22 (02) : 260 - 268
  • [3] Demonstration of the Need for an Air-Kerma Rate Standard for the Newest Model of the Xoft Axxent TM Source
    Walter, A.
    Anderson, D.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E575 - E575
  • [4] Determination and Evaluation of a Dose-Rate Conversion Coefficient for the Xoft Axxent® Electronic Brachytherapy Source in the Xoft Titanium Cervical Applicator
    Simiele, S.
    Micka, J.
    Culberson, W.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3709 - 3709
  • [5] Dose-Rate Constant Determination of the Xoft Axxent Electronic Brachytherapy Source Using Spectroscopic Methods
    Pike, T.
    Moga, J.
    Davis, S.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Air-kerma strength determination of a 169Yb high dose rate brachytherapy source
    VanDamme, J. J.
    Culberson, W. S.
    DeWerd, L. A.
    Micka, J. A.
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 3935 - 3942
  • [7] Air-kerma strength determination of a 169Ytterbium high dose rate brachytherapy source
    VanDamme, J.
    DeWerd, L.
    Micka, J.
    Culberson, W.
    Davis, S.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1984 - 1984
  • [8] Experimental Determination of Ionization Chamber Correction Factors for In-Phantom Measurements of Reference Air Kerma Rate and Absorbed Water Dose Rate of Brachytherapy 192Ir Source
    Chan, M.
    Lee, V.
    Wong, M.
    Leung, R.
    Law, G.
    Lee, K.
    Cheung, S.
    Tung, S.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3465 - 3465
  • [9] Determination of air-kerma strength for the 192Ir GammaMedplus iX pulsed-dose-rate brachytherapy source
    Riley, A. D.
    Pike, T. L.
    Micka, J. A.
    Fulkerson, R. K.
    DeWerd, L. A.
    [J]. MEDICAL PHYSICS, 2013, 40 (07)
  • [10] Determination of the Kwall correction factor for a cylindrical ionization chamber to measure air-kerma in 60Co gamma beams
    Laitano, RF
    Toni, MP
    Pimpinella, M
    Bovi, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (14): : 2411 - 2431