Comparison of air kerma rate between the S7500 and S7600 xoft axxent sources

被引:2
|
作者
Walter, Autumn E. [1 ,2 ]
Hull, Jennifer L. [1 ]
DeWerd, Larry A. [1 ]
机构
[1] Univ Wisconsin Madison, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, 1111 Highland Ave Madison, Madison, WI 53705 USA
关键词
Electronic brachytherapy; Air-kerma; Well chamber; Dosimetry; WELL CHAMBER RESPONSE; AMBIENT-PRESSURE; BRACHYTHERAPY; DOSIMETRY; STANDARD;
D O I
10.1016/j.brachy.2021.12.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
ABSTRACT PURPOSE: The purpose of this work was to evaluate differences in air-kerma rate of the older, S7500 water-cooled Xoft Axxent source and newer, S7600 Galden-cooled source. used to measure the air-kerma rate at 50 cm for six S7600 Xoft Axxent sources. The average measured air-kerma of the S7600 sources was compared with the measured average air-kerma rate from five S7500 sources. The air-kerma rates of the S7500 sources were measured in a Standard Imaging HDR 1000 + well chamber. The FAC measurements were used to determine a well chamber calibration coefficient for the S7600 source. The S7500 calibration coefficients were incorrectly applied to the S7600 sources to indicate the magnitude of error that can occur if the incorrect calibration coefficient is used. RESULTS: A 10.3% difference was observed between the average air-kerma rates of the two sources although a 17% difference was observed between their calibration coefficients. The application of the S7500 calibration coefficient to the S7600 sources resulted in measured air-kerma rates that were 20% greater than the true value. CONCLUSIONS: This work indicates the need for a new air-kerma rate standard for the S7600 sources, and the results presented in this work are indicative of values that would be obtained at National Institute of Standards and Technology. (c) 2021 The Authors. Published by Elsevier Inc. on behalf of American Brachytherapy Society. This is an open access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ )
引用
收藏
页码:354 / 361
页数:8
相关论文
共 42 条
  • [1] Determination of an air kerma-rate correction factor for the S7600 Xoft Axxent ® source model
    Walter, Autumn E.
    DeWerd, Larry A.
    [J]. BRACHYTHERAPY, 2023, 22 (04) : 512 - 517
  • [2] Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model
    Walter, Autumn E.
    Khan, Ahtesham U.
    DeWerd, Larry A.
    [J]. BRACHYTHERAPY, 2023, 22 (02) : 260 - 268
  • [3] Demonstration of the Need for an Air-Kerma Rate Standard for the Newest Model of the Xoft Axxent TM Source
    Walter, A.
    Anderson, D.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E575 - E575
  • [4] INVESTIGATION OF THE DOSE PROPERTIES AND SOURCE TO SOURCE VARIABILITIES IN XOFT SOURCE MODEL S7500
    Esmaelbeigi, Azin
    Kalinowski, Jonathan
    Vuong, Te
    Enger, Shirin A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 186 : S85 - S85
  • [5] Comparison of Axxent-Xoft, 192Ir and 60Co high-dose-rate brachytherapy sources for image-guided brachytherapy treatment planning for cervical cancer
    Mobit, P. N.
    Packianathan, S.
    He, R.
    Yang, C. C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1052):
  • [6] Csiszar's forward cutoff rate for testing between two arbitrary sources
    Alajaji, F
    Chen, PN
    Rached, Z
    [J]. ISIT: 2002 IEEE INTERNATIONAL SYMPOSIUM ON INFORMATION THEORY, PROCEEDINGS, 2002, : 215 - 215
  • [7] Comparison between the rate of Parkinson's disease progression and patient gender
    Trufanov, Y. O.
    Golovchenko, Y. I.
    [J]. MOVEMENT DISORDERS, 2012, 27 : S143 - S144
  • [8] Comparison between traditional saline versus air expanders: a patient's perspective
    Sokolovska, Sonja
    Grinsell, Damien
    [J]. ANZ JOURNAL OF SURGERY, 2019, 89 (04) : 282 - 283
  • [9] APMP Key comparison report of reference air kerma rate for HDR 192Ir brachytherapy sources (BIPM KCDB: APMP.RI(I)-K8)
    Ishii, J.
    Kurosawa, T.
    Kato, M.
    Toroi, P.
    Chu, W-H
    Yi, C-Y
    Kim, Y. H.
    Seong, Y. M.
    Ngcezu, S. A.
    Mainegra-Hing, E.
    Downtonf, B.
    Dolah, M. T.
    Budiantari, C. T.
    [J]. METROLOGIA, 2021, 58 (1A)
  • [10] Blinking Rate Comparison Between Patients with Chronic Pain and Parkinson's Disease
    Stefanescu, E.
    Pancino, N.
    Graziani, C.
    Lachi, V.
    Sampoli, M.
    Dimitri, G.
    Bargagli, A.
    Zanca, D.
    Bianchini, M.
    Muresanu, D.
    Rufa, A.
    [J]. EUROPEAN JOURNAL OF NEUROLOGY, 2022, 29 : 669 - 669