An effective and optimized patient-specific QA workload reduction for VMAT plans after MLC-modelling optimization

被引:2
|
作者
Marsac, Thomas [1 ]
Camejo, Arianna Batista [2 ]
Chiavassa, Sophie [1 ]
Jan, Simon [1 ]
Lorand, Hugo [1 ]
Moignier, Alexandra [1 ]
机构
[1] Inst Cancreol Ouest, Med Phys Dept, Bd Prof Jacques Monod, F-44805 St Herblain, France
[2] Inst Cancreol Ouest, Med Phys Dept, Rue Andre Bocquel, F-49055 Angers, France
关键词
Radiotherapy; Patient-specific quality assurance; Complexity metrics; MLC modelling; Workload reduction; ARC THERAPY PLANS; QUALITY-ASSURANCE; COMPLEXITY; IMRT; METRICS; DELIVERY; RADIOTHERAPY;
D O I
10.1016/j.ejmp.2023.102548
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Many complexity metrics characterize modulated plans. First, this study aimed at identify the optimal complexity metrics to reduce workload associated to patient-specific quality assurance (PSQA) for our equipment and processes. Second, it intended to optimize our MLC modelling to improve measurement and calculation agreement with expectation of further reducing PSQA workload.Methods: Correlation and sensitivity at specificity equals to 1 were evaluated for PSQA results and different complexity metrics. Thresholds to stop PSQA were determined. After validation of the optimal complexity metric and threshold for our equipment and process, the MLC modelling was reviewed with a recently published methodology. This method is based on measurements with a Farmer-type ionization chamber of synchronous and asynchronous sweeping gap plans. Effect on the PSQA results and the identified threshold was investigated.Results: In our center, the most appropriate complexity metric for reducing our PSQA workload was the Mod-ulation Complexity Score for VMAT (MCSv). The optimization of the MLC modelling significantly reduced the number of controlled plans, specifically for one of our two Varian Clinac. Any plan with a MCSv >= 0.34 is treated without PSQA.Conclusion: This study rationalized and reduced our PSQA workload by approximately 30%. It is a continuing work with new TPS, machine or PSQA equipment. It encourages centers to re-evaluate their MLC modelling as well as assess the benefit of complexity metrics to streamline their PSQA workflow. An easier access, at least for reporting, at best for optimizing plans, into the TPS would be beneficial for the community.
引用
收藏
页数:8
相关论文
共 10 条
  • [1] Patient-specific QA for CyberKnife MLC plans using Monte Carlo
    Mackeprang, P. H.
    Vuong, D.
    Volken, W.
    Henzen, D.
    Schmidhalter, D.
    Malthaner, M.
    Mueller, S.
    Frei, D.
    Aebersold, D. M.
    Fix, M. K.
    Manser, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S791 - S791
  • [2] Anatomy-based, patient-specific VMAT QA using EPID or MLC log files
    Defoor, Dewayne L.
    Vazquez-Quino, Luis A.
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    Stathakis, Sotirios
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 206 - 215
  • [3] Reduction of Systematic Dosimetric Uncertainties in VMAT Treatments Triggered by Patient-Specific QA
    Mijnheer, B.
    Janssen, T.
    van Mourik, A.
    Kaas, J.
    Damen, E.
    Mans, A.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E245 - E245
  • [4] Patient-specific QA using DVH analysis provides better clinical relevance of complex VMAT plans
    Yoosuf, A.
    AlShehri, S.
    AlHadab, A.
    Alqathami, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S711 - S711
  • [5] The Sensitivity of Gamma Index to Detect MLC Positioning Errors Using Varian EPID and ArcCHECK for Patient-Specific VMAT QA
    Maraghechi, B.
    Darko, J.
    Davis, J.
    Mitchell, N.
    Shah, M.
    Fleck, A.
    Osei, E.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2887 - 2887
  • [6] Improvement of deep learning prediction model in patient-specific QA for VMAT with MLC leaf position map and patient's dose distribution
    Tozuka, Ryota
    Kadoya, Noriyuki
    Tomori, Seiji
    Kimura, Yuto
    Kajikawa, Tomohiro
    Sugai, Yuto
    Xiao, Yushan
    Jingu, Keiichi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (10):
  • [7] The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT
    Kim, Jung-In
    Park, So-Yeon
    Kim, Hak Jae
    Kim, Jin Ho
    Ye, Sung-Joon
    Park, Jong Min
    [J]. RADIATION ONCOLOGY, 2014, 9
  • [8] The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT
    Jung-in Kim
    So-Yeon Park
    Hak Jae Kim
    Jin Ho Kim
    Sung-Joon Ye
    Jong Min Park
    [J]. Radiation Oncology, 9
  • [9] Comparison of 2D and 3D Gamma Analysis in Patient-Specific QA for Prostate VMAT Plans
    Clemente, F.
    Perez, C.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 240 - +
  • [10] Investigating the Ability of Patient-Specific QA Devices to Detect Changes in Dose Distribution Due to MLC Positioning Errors in Stereotactic Radiotherapy Plans
    James, S.
    Elder-, E.
    Huh, C.
    Al-Basheer, A.
    Barrett, J.
    Mostafaei, F.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E235 - E235