Real-time monitoring of ultra-high dose rate electron beams using bremsstrahlung photons

被引:0
|
作者
Kim, Hyun [1 ]
Jeong, Dong Hyeok [1 ]
Kang, Sang Koo [1 ]
Lee, Manwoo [1 ]
Lim, Heuijin [1 ]
Lee, Sang Jin [1 ]
Jang, Kyoung Won [1 ]
机构
[1] Dongnam Inst Radiol & Med Sci, Res Ctr, Busan 46033, South Korea
关键词
UHDR electron Beams; Real-time monitoring; Ionization chamber; FLASH radiotherapy; RECOMBINATION; IRRADIATION; COLLECTION; DOSIMETRY;
D O I
10.1016/j.net.2023.05.033
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Recently, as the clinically positive biological effects of ultra-high dose rate (UHDR) radiation beams have been revealed, interest in flash radiation therapy has increased. Generally, FLASH preclinical experiments are performed using UHDR electron beams generated by linear accelerators. Real-time monitoring of UHDR beams is required to deliver the correct dose to a sample. However, it is difficult to use typical transmission-type ionization chambers for primary beam monitoring because there is no suitable electrometer capable of reading high pulsed currents, and collection efficiency is drastically reduced in pulsed radiation beams with ultra-high doses. In this study, a monitoring method using bremsstrahlung photons generated by irradiation devices and a water phantom was proposed. Charges collected in an ionization chamber located at the back of a water phantom were analyzed using the bremsstrahlung tail on electron depth dose curves obtained using radiochromic films. The dose conversion factor for converting a monitored charge into a delivered dose was determined analytically for the Advanced Markus & REG; chamber and compared with experimentally determined values. It is anticipated that the method proposed in this study can be useful for monitoring sample doses in UHDR electron beam irradiation. & COPY; 2023 Korean Nuclear Society, Published by Elsevier Korea LLC. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:3417 / 3422
页数:6
相关论文
共 50 条
  • [21] Determination of the ion collection efficiency of the Razor Nano Chamber for ultra-high dose-rate electron beams
    Cavallone, Marco
    Jorge, Patrik Goncalves
    Moeckli, Raphael
    Bailat, Claude
    Flacco, Alessandro
    Prezado, Yolanda
    Delorme, Rachel
    MEDICAL PHYSICS, 2022, 49 (07) : 4731 - 4742
  • [22] Pulse-By-Pulse Analysis to Evaluate Response of Electrometers to Ultra-High Dose-Rate Electron Beams
    Mancera, M. A. Flores
    Radtke, J.
    Culberson, W. S.
    MEDICAL PHYSICS, 2024, 51 (10) : 7801 - 7802
  • [23] First Characterization of Novel Silicon Carbide Detectors with Ultra-High Dose Rate Electron Beams for FLASH Radiotherapy
    Romano, Francesco
    Milluzzo, Giuliana
    Di Martino, Fabio
    D'Oca, Maria Cristina
    Felici, Giuseppe
    Galante, Federica
    Gasparini, Alessia
    Mariani, Giulia
    Marrale, Maurizio
    Medina, Elisabetta
    Pacitti, Matteo
    Sangregorio, Enrico
    Vanreusel, Verdi
    Verellen, Dirk
    Vignati, Anna
    Camarda, Massimo
    APPLIED SCIENCES-BASEL, 2023, 13 (05):
  • [24] Ultra-High Sensitivity Real-Time Monitoring of Landslide Surface Deformation via Triboelectric Nanogenerator
    Wang, Chao
    Yang, Yu
    Zhang, Xiaosong
    Wang, Pengfei
    Bi, Xiangzhuang
    Li, Hengyu
    Wang, Zhong Lin
    Cheng, Tinghai
    ADVANCED MATERIALS, 2024, 36 (50)
  • [25] Electron Ultra-High Dose-Rate (FLASH) beam monitoring and control through beam current transformers
    Schueler, E.
    Liu, K.
    Palmiero, A.
    Chopra, N.
    Li, Z.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S752 - S753
  • [26] A GPU optimization workflow for real-time execution of ultra-high frame rate computer vision applications
    Nourazar, Mohsen
    Booth, Brian G.
    Goossens, Bart
    JOURNAL OF REAL-TIME IMAGE PROCESSING, 2024, 21 (01)
  • [27] A GPU optimization workflow for real-time execution of ultra-high frame rate computer vision applications
    Mohsen Nourazar
    Brian G. Booth
    Bart Goossens
    Journal of Real-Time Image Processing, 2024, 21
  • [28] Demonstration of ultra-high dose rate electron irradiation at FLASHlab@PITZ
    Li, X-k
    Amirkhanyan, Z.
    Grebinyk, A.
    Gross, M.
    Komar, Y.
    Riemer, F.
    Asoyan, A.
    Boonpornprasert, P.
    Borchert, P.
    Davtyan, H.
    Dmytriiev, D.
    Frohme, M.
    Hoffmann, A.
    Krasilnikov, M.
    Loisch, G.
    Lotfi, Z.
    Mueller, F.
    Schmitz, M.
    Obier, F.
    Oppelt, A.
    Philipp, S.
    Richard, C.
    Vashchenko, G.
    Villani, D.
    Worm, S.
    Stephan, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (05):
  • [29] Radiation safety of ultra-high dose rate electron accelerators for FLASH radiotherapy
    Praestegaard, Lars Hjorth
    MEDICAL PHYSICS, 2024, 51 (09) : 6206 - 6219
  • [30] Characterization of the PTB ultra-high pulse dose rate reference electron beam
    Bourgouin, Alexandra
    Knyziak, Adrian
    Marinelli, Marco
    Kranzer, Rafael
    Schueller, Andreas
    Kapsch, Ralf-Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08):