Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients

被引:4
|
作者
Khanna, Ashish K. K. [1 ,2 ,3 ]
Garcia, Julio O. [4 ]
Saha, Amit K. [3 ,5 ]
Harris, Lynnette [3 ,5 ]
Baruch, Martin [6 ]
Martin, R. Shayn [7 ]
机构
[1] Wake Forest Univ, Atrium Hlth Wake Forest Baptist Med Ctr, Sch Med, Dept Anesthesiol,Sect Crit Care Med, Winston Salem, NC 27109 USA
[2] Outcomes Res Consortium, Cleveland, OH 44195 USA
[3] Perioperat Outcomes & Informat Collaborat POIC, Winston Salem, NC 27109 USA
[4] Wake Forest Univ, Sch Med, Winston Salem, NC USA
[5] Wake Forest Univ, Atrium Hlth Wake Forest Baptist Med Ctr, Dept Anesthesiol, Sch Med, Winston Salem, NC USA
[6] Caretaker Med, Charlottesville, VA USA
[7] Wake Forest Univ, Atrium Hlth Wake Forest Baptist Med Ctr, Dept Surg, Sch Med, Winston Salem, NC USA
关键词
Pulse decomposition analysis; Cardiac output; Hemodynamics; Wireless; Monitoring; Cardiac surgery; Postoperative; Intensive care unit; ACCURACY; PRESSURE; NEXFIN; CNAP;
D O I
10.1007/s10877-023-01059-5
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
PurposePulse Decomposition Analysis (PDA) uses integration of the systolic area of a distally transmitted aortic pulse as well as arterial stiffness estimates to compute cardiac output. We sought to assess agreement of cardiac output (CO) estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CCO) and a wireless, wearable noninvasive device, (Vitalstream, Caretaker Medical, Charlottesville, VA), that utilizes the Pulse Decomposition Analysis (CO-PDA) method in postoperative cardiac surgery patients in the intensive care unit.MethodsCO-CCO measurements were compared with post processed CO-PDA measurements in prospectively enrolled adult cardiac surgical intensive care unit patients. Uncalibrated CO-PDA values were compared for accuracy with CO-CCO via a Bland-Altman analysis considering repeated measurements and a concordance analysis with a 10% exclusion zone.Results259.7 h of monitoring data from 41 patients matching 15,583 data points were analyzed. Mean CO-CCO was 5.55 L/min, while mean values for the CO-PDA were 5.73 L/min (mean of differences +- SD 0.79 & PLUSMN; 1.11 L/min; limits of agreement - 1.43 to 3.01 L/min), with a percentage error of 37.5%. CO-CCO correlation with CO-PDA was moderate (0.54) and concordance was 0.83.ConclusionCompared with the CO-CCO Swan-Ganz, cardiac output measurements obtained using the CO-PDA were not interchangeable when using a 30% threshold. These preliminary results were within the 45% limits for minimally invasive devices, and pending further robust trials, the CO-PDA offers a noninvasive, wireless solution to complement and extend hemodynamic monitoring within and outside the ICU.
引用
收藏
页码:139 / 146
页数:8
相关论文
共 50 条
  • [31] Comparison of cardiac output and cardiac index values measured by critical care echocardiography with the values measured by pulse index continuous cardiac output (PiCCO) in the pediatric intensive care unit:a preliminary study
    Nagehan Aslan
    Dincer Yildizdas
    Ozden Ozgur Horoz
    Yasemin Coban
    Fadli Demir
    Sevcan Erdem
    Yasar Sertdemir
    Italian Journal of Pediatrics, 46
  • [32] Arterial Blood Pressure Waveform Based Cardiac Output Analysis Correlates with Continuous Pulmonary Artery Thermodilution In Post Cardiac Surgery ICU Patients Including Those With Arrhythmia
    Khanna, Ashish K.
    Nosow, Lillian M.
    Sands, Lauren E.
    Saha, Amit
    Agashe, Harsha
    Harris, Lynnette
    Marchant, Bryan E.
    ANESTHESIA AND ANALGESIA, 2022, 134 : 189 - 190
  • [33] Post-Cardiac Surgery Patients With Hyperlactatemia In The Intensive Care Unit: Outcome And Risk Factors
    Mak, N. T. J. J.
    Iqbal, S.
    Banici, L.
    De Varennes, B.
    Khwaja, K.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [34] Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia
    Dyer, R. A.
    Piercy, J. L.
    Reed, A. R.
    Strathie, G. W.
    Lombard, C. J.
    Anthony, J. A.
    James, M. F.
    BRITISH JOURNAL OF ANAESTHESIA, 2011, 106 (01) : 77 - 81
  • [35] Comparison of non-calibrated pulse-contour analysis with continuous thermodilution for cardiac output assessment in patients with induced hypothermia after cardiac arrest
    Haenggi, Matthias
    Barthelmes, Daniel
    Ulmer, Hanno
    Takala, Jukka
    Jakob, Stephan M.
    RESUSCITATION, 2011, 82 (04) : 423 - 426
  • [36] Flow-Regulated Extracorporeal Arteriovenous Tubing Loop for Cardiac Output Measurements by Ultrasound Velocity Dilution: Validation in Post-Cardiac Surgery Intensive Care Unit Patients
    Eremenko, Alexsandr A.
    Safarov, Perviz N.
    ASAIO JOURNAL, 2010, 56 (06) : 522 - 526
  • [37] Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit
    Siegelaar, Sarah E.
    Barwari, Temo
    Hermanides, Jeroen
    van der Voort, Peter H. J.
    Hoekstra, Joost B. L.
    DeVries, J. Hans
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (05): : 1283 - 1289
  • [38] Microcirculation and its relation with continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit
    Siegelaar, S. E.
    Barwari, T.
    Hermanides, J.
    van der Voort, P. H. J.
    Hoekstra, J. B. L.
    DeVries, J. H.
    DIABETOLOGIA, 2011, 54 : S116 - S117
  • [39] Outcomes of post-cardiac surgery patients with persistent hyperlactatemia in the intensive care unit: a matched cohort study
    Mak, Nicole T. J. J.
    Iqbal, Sameena
    de Varennes, Benoit
    Khwaja, Kosar
    JOURNAL OF CARDIOTHORACIC SURGERY, 2016, 11
  • [40] Outcomes of post-cardiac surgery patients with persistent hyperlactatemia in the intensive care unit: a matched cohort study
    Nicole T. J. J. Mak
    Sameena Iqbal
    Benoit de Varennes
    Kosar Khwaja
    Journal of Cardiothoracic Surgery, 11