Impact of Fingernail Polish on Pulse Oximetry Measurements: A Systematic Review

被引:4
|
作者
Aggarwal, Ashutosh Nath [1 ]
Agarwal, Ritesh [1 ]
Dhooria, Sahajal [1 ]
Prasad, Kuruswamy Thurai [1 ]
Sehgal, Inderpaul Singh [1 ]
Muthu, Valliappan [1 ]
机构
[1] Postgrad Inst Med Educ & Res, Dept Pulm Med, Chandigarh 160012, India
关键词
color; nail polish; oxygen saturation; pulse oximetry; systematic review; NAIL POLISH; OXYGEN-SATURATION; INTERFERENCE; HENNA;
D O I
10.4187/respcare.10399
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The effect of application of fingernail polish on SpO2 measurement remains unclear. We conducted this systematic review to ascertain the impact of fingernail polish on SpO2 measurement. METHODS: We queried PubMed, Embase, and CINAHL databases for publica-tions indexed through December 2022. We included studies providing paired SpO2 data from fin-gertips without and after nail polish application or reporting the number of subjects whose SpO2 could not be measured due to fingernail polish. We used random effects modeling to summarize standardized mean differences (SMDs) and corresponding 95% CI for different nail polish colors from comparative studies. RESULTS: We retrieved 122 studies and included 21 publications, mostly performed on healthy volunteers. Of these, 17 (81.0%) studies had a low risk of bias. We summarized mean SMD for 10 nail polish colors (black, blue, brown, green, orange, pink, purple, red, white, and yellow) from 25 paired data sets on SpO2 across 20 studies. We found small (likely clinically insignificant) but statistically significant differences in mean SpO2 when fingers were coated with black, blue, brown, or purple nail polish (SMD -0.57, -0.47, -0.33, and -0.25, respectively; 95% CI -0.86 to -0.29, -0.84 to -0.10, -0.59 to -0.07, and -0.48 to -0.02, respectively). Only one of 4 studies reported a high proportion of unsuccessful oxim-eter readings from fingers painted with black (88.0%) or brown (36.0%) nail polish. CONCLUSIONS: Although fingernail polish of some colors can marginally reduce SpO2 reading or occasionally impede SpO2 measurement, the variability is clinically insignificant.
引用
收藏
页码:1271 / 1280
页数:10
相关论文
共 50 条
  • [41] Influence of cuff pressures of automatic sphygmomanometers on pulse oximetry measurements
    Sondej, Tadeusz
    Zawadzka, Sylwia
    MEASUREMENT, 2022, 187
  • [42] Effect of routine pulse oximetry measurements on ED triage classification
    Summers, RL
    Anders, RM
    Woodward, LH
    Jenkins, AK
    Galli, RL
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1998, 16 (01): : 5 - 7
  • [43] Variability in estimating shunt from single pulse oximetry measurements
    Karlen, Walter
    Petersen, Christian L.
    Dumont, Guy A.
    Ansermino, J. Mark
    PHYSIOLOGICAL MEASUREMENT, 2015, 36 (05) : 967 - 981
  • [44] Melanin and the Impact to Optical Signals Common in Pulse Oximetry
    Dove, Jake
    McGonigle, Scott
    Gomes, Pedro
    ANESTHESIA AND ANALGESIA, 2024, 138
  • [45] The impact of pulse oximetry on diagnosis, management and outcomes of acute febrile illness in low-income and middle-income countries: a systematic review
    Chew, Rusheng
    Zhang, Meiwen
    Chandna, Arjun
    Lubell, Yoel
    BMJ GLOBAL HEALTH, 2021, 6 (11):
  • [46] Effectiveness and safety of pulse oximetry in remote patient monitoring of patients with COVID-19: a systematic review
    Alboksmaty, Ahmed
    Beaney, Thomas
    Elkin, Sarah
    Clarke, Jonathan M.
    Darzi, Ara
    Aylin, Paul
    Neves, Ana-Luisa
    LANCET DIGITAL HEALTH, 2022, 4 (04):
  • [47] Pulse Oximetry and Perfusion Index Screening for Congenital Heart Defects: A Systematic Review and Meta-analysis
    Jiang, S. L.
    Zhan, Y. J.
    Yan, P.
    Yue, Y.
    Tang, J.
    AMERICAN JOURNAL OF PERINATOLOGY, 2023, 40 (15) : 1611 - 1617
  • [48] Pulse Oximetry: Review of a Potential Aid in Endodontic Diagnosis
    Jafarzadeh, Hamid
    Rosenberg, Paul A.
    JOURNAL OF ENDODONTICS, 2009, 35 (03) : 329 - 333
  • [49] REVIEW OF HEMOGLOBIN VARIANTS WITH ABNORMAL PULSE OXIMETRY READINGS
    Forward, Jodi
    Hogan, Mary-Jane
    PEDIATRIC BLOOD & CANCER, 2022, 69
  • [50] Determining factors of pulse oximetry accuracy: A literature review
    Leon-Valladares, D.
    Barrio-Mateu, L. A.
    Cortes-Carmona, N.
    Fuentes-Lizana, G.
    Cabanas, A. M.
    Latorre-Progulakis, K.
    Guajardo, M. Fuentes
    Martin-Escudero, P.
    REVISTA CLINICA ESPANOLA, 2024, 224 (05): : 314 - 330