Cochlear coverage with lateral wall cochlear implant electrode arrays affects post-operative speech recognition

被引:7
|
作者
Weller, Tobias [1 ,2 ,3 ]
Timm, Max Eike [1 ]
Lenarz, Thomas [1 ,2 ,3 ]
Buechner, Andreas [1 ,2 ,3 ]
机构
[1] Hannover Med Sch, Dept Otorhinolaryngol, Hannover, Germany
[2] German Hearing Ctr DHZ, Hannover, Germany
[3] Cluster Excellence Hearing4All, Oldenburg, Germany
来源
PLOS ONE | 2023年 / 18卷 / 07期
关键词
INSERTION DEPTH; WORD RECOGNITION; PERFORMANCE; OUTCOMES; MORPHOLOGY; LOCATION; POSITION; LENGTH;
D O I
10.1371/journal.pone.0287450
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
ObjectivesThe goal was to investigate the relationship between the insertion angle/cochlear coverage of cochlear implant electrode arrays and post-operative speech recognition scores in a large cohort of patients implanted with lateral wall electrode arrays. MethodsPre- and post-operative cone beam computed tomography scans of 154 ears implanted with lateral wall electrode arrays were evaluated. Traces of lateral wall and electrode arrays were combined into a virtual reconstruction of the implanted cochlea. This reconstruction was used to measure insertion angles and proportional cochlear coverage. Word recognition scores and sentence recognition scores measured 12 months after implantation using electric-only stimulation were used to examine the relationship between cochlear coverage/insertion angle and implantation outcomes. ResultsPost-operative word recognition scores and the difference between post- and pre-operative word recognition scores were positively correlated with both cochlear coverage and insertion angle, however sentence recognition scores were not. A group-wise comparison of word recognition scores revealed that patients with cochlear coverage below 70% performed significantly worse than patients with coverage between 79%-82% (p = 0.003). Performance of patients with coverage above 82% was on average poorer than between 79%-82, although this finding was not statistically significant (p = 0.84). Dividing the cohort into groups based on insertion angle quadrants revealed that word recognition scores were highest above 450 & DEG; insertion angle, sentence recognition scores were highest between 450 & DEG; and 630 & DEG; and the difference between pre- and post-operative word recognition scores was largest between 540 & DEG; and 630 & DEG;, however none of these differences reached statistical significance. ConclusionsThe results of this study show that cochlear coverage has an effect on post-operative word recognition abilities and the benefit patients receive from their implant. Generally, higher coverage led to better outcomes, however there were results indicating that insertion past 82% cochlear coverage may not provide an additional benefit for word recognition. These findings can be useful for choosing the optimal electrode array and thereby improving cochlear implantation outcomes on a patient-individual basis.
引用
收藏
页数:22
相关论文
共 50 条
  • [1] Automatic Identification of Cochlear Implant Electrode Arrays for Post-Operative Assessment
    Noble, Jack H.
    Schuman, Theodore A.
    Wright, Charles G.
    Labadie, Robert F.
    Dawant, Benoit M.
    MEDICAL IMAGING 2011: IMAGE PROCESSING, 2011, 7962
  • [2] Effects of cochlear implant surgical technique on post-operative electrode impedance
    Gu, Ping
    Jiang, Yi
    Gao, Xue
    Huang, Shasha
    Yuan, Yongyi
    Wang, Guojian
    Li, Beicheng
    Xi, Xin
    Dai, Pu
    ACTA OTO-LARYNGOLOGICA, 2016, 136 (07) : 677 - 681
  • [3] Preoperative prediction of insertion depth of lateral wall cochlear implant electrode arrays
    Khan, Mohammad M. R.
    Labadie, Robert F.
    Noble, Jack H.
    MEDICAL IMAGING 2020: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2021, 11315
  • [4] COCHLEAR IMPLANT ELECTRODE LOCALIZATION IN POST-OPERATIVE CT USING A SPHERICAL MEASURE
    Braithwaite, Benjamin
    Kjer, H. Martin
    Fagertun, Jens
    Gonzalez Ballester, Miguel A.
    Dhanasingh, Anandhan
    Mistrik, Pavel
    Gerber, Nicolas
    Paulsen, Rasmus R.
    2016 IEEE 13TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2016, : 1329 - 1333
  • [5] Preoperative prediction of angular insertion depth of lateral wall cochlear implant electrode arrays
    Khan, Mohammad M. R.
    Labadie, Robert F.
    Noble, Jack H.
    JOURNAL OF MEDICAL IMAGING, 2020, 7 (03)
  • [6] Influence of electrode to cochlear duct length ratio on post-operative speech understanding outcomes
    Cooperman, Shayna P.
    Aaron, Ksenia A.
    Fouad, Ayman
    Tran, Emma
    Blevins, Nikolas H.
    Fitzgerald, Matthew B.
    COCHLEAR IMPLANTS INTERNATIONAL, 2022, 23 (02) : 59 - 69
  • [7] Comparative Performance of Lateral Wall and Perimodiolar Cochlear Implant Arrays
    Sturm, Joshua J.
    Patel, Vir
    Dibelius, Greg
    Kuhlmey, Megan
    Kim, Ana H.
    OTOLOGY & NEUROTOLOGY, 2021, 42 (04) : 532 - 539
  • [8] Research Advances in Post-Operative Rehabilitation Following Cochlear Implant
    LI Yong-xin1
    JournalofOtology, 2007, (02) : 92 - 96
  • [9] Preoperative angular insertion depth prediction in case of lateral wall cochlear implant electrode arrays
    Khan, Mohammad M. R.
    Labadie, Robert F.
    Noble, Jack H.
    MEDICAL IMAGING 2020: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2021, 11315
  • [10] Speech Recognition Outcomes in Adults With Slim Straight and Slim Modiolar Cochlear Implant Electrode Arrays
    MacPhail, Margaret E.
    Connell, Nathan T.
    Totten, Douglas J.
    Gray, Mitchell T.
    Pisoni, David
    Yates, Charles W.
    Nelson, Rick F.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2022, 166 (05) : 943 - 950