Dynamic Changes of Capsular-Intraocular Lens Adhesion in Plate-Haptic Hydrophilic and Loop-Haptic Hydrophobic Eyes

被引:0
|
作者
Zhao, Yinying [1 ,2 ]
Zhang, Zhewen [1 ,2 ]
Yang, Fuman [1 ,2 ]
Chang, Pingjun [1 ,2 ]
Wang, Dandan [1 ,2 ]
Huang, Feng [1 ,2 ]
Zhao, Yune [1 ,2 ]
机构
[1] Wenzhou Med Univ, Eye Hosp, 618 East Feng Qi Rd, Hangzhou 310000, Zhejiang, Peoples R China
[2] Natl Clin Res Ctr Ocular Dis, Wenzhou, Zhejiang, Peoples R China
关键词
Intraocular lens; Capsular adhesion; Plate-haptic; Loop-haptic; OPTIC EDGE DESIGN; POSTERIOR CAPSULE; QUANTITATIVE-ANALYSIS; ROTATIONAL STABILITY; OPACIFICATION; SILICONE;
D O I
10.1007/s40123-024-00933-y
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Introduction The aim of this work is to investigate the dynamic changes of capsular-intraocular lens (IOL) adhesion in plate-haptic hydrophilic and loop-haptic hydrophobic eyes.Methods Cataract eyes that met the inclusion criteria were randomly assigned to receive implantation of a plate-haptic hydrophilic or loop-haptic IOL. The anterior capsular adhesion, posterior capsular adhesion, and the configurations of the capsular bend were evaluated using swept-source optical coherence tomography at 1 day, 1 week, 1 month, and 3 months postoperatively.Results In total, 66 eyes of 66 patients were eligible for the analysis: 33 in the plate-haptic group and 33 in the loop-haptic group. The contact between the anterior capsule and IOL in the plate-haptic group was earlier than that in the loop-haptic group upon comparing the measurements taken at 1 day and 1 week (p = 0.001, p = 0.003, respectively). The complete attachment of the posterior capsule and IOL in the plate-haptic group was significantly greater at 1 week, 1 month, and 3 months (p = 0.001, p = 0.000, p = 0.001, respectively). The capsular bend index of the plate-haptic group was significantly greater than that of the loop-haptic group at each time points except at 1 day (p = 0.007, p = 0.049, p = 0.005, respectively). Furthermore, a new type of capsular bend, "cocked adhesion," was observed in the plate-haptic eyes.Conclusions The plate-haptic IOL demonstrated excellent capsular adhesion compared to the loop-haptic IOL, which was probably attributed to haptic compressibility. A special cocked configuration of the capsular bend in plate-haptic IOL was observed for the first time. Further studies are warranted to confirm the effect of the new type of capsular bend.
引用
收藏
页码:1527 / 1535
页数:9
相关论文
共 44 条
  • [21] Rotational stability of plate-haptic toric intraocular lenses in Asian eyes: risk period for intraocular lens rotation and its influencing factors
    Ma, Dongmei
    Shen, Jiying
    Qiu, Tian
    Zhang, Limei
    Hua, Zhixiang
    Cai, Lei
    Yang, Jin
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2023, 49 (03): : 253 - 258
  • [22] Optimal Timing of Repositioning Surgery for a Plate-Haptic Toric Intraocular Lens: A Multicenter Retrospective Study
    Ma, Dongmei
    Yu, Wenjuan
    Cai, Lei
    Shen, Jiying
    Guo, Tao
    Chen, Xu
    Dong, JianHong
    Zeng, Qiulin
    Wang, Aijun
    Hua, Zhixiang
    Zhu, Xiangjia
    Zheng, Tianyu
    Lu, Yi
    Yang, Jin
    JOURNAL OF REFRACTIVE SURGERY, 2023, 39 (02) : 120 - +
  • [23] Associations between anterior segment parameters and rotational stability of a plate-haptic toric intraocular lens
    Yao, Yunqian
    Meng, Jiaqi
    He, Wenwen
    Zhang, Keke
    Wei, Ling
    Cheng, Kaiwen
    Lu, Yi
    Zhu, Xiangjia
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (11): : 1436 - 1440
  • [24] Rotation and decentration of an undersized plate-haptic trifocal toric intraocular lens in an eye with moderate myopia
    Giers, Bert C.
    Khoramnia, Ramin
    Weber, Lea F.
    Tandogan, Tamer
    Auffarth, Gerd U.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (03): : 489 - 493
  • [25] Two-string technique to manage dislocated posterior chamber plate-haptic intraocular lens
    Rahim, Mohammad Fayez
    Malyugin, Boris
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (05): : 722 - 726
  • [26] Clinical and Optical Intraocular Performance of Rotationally Asymmetric Multifocal IOL Plate-Haptic Design Versus C-Loop Haptic Design
    Alio, Jorge L.
    Plaza-Puche, Ana B.
    Javaloy, Jaime
    Jose Ayala, Maria
    Vega-Estrada, Alfredo
    JOURNAL OF REFRACTIVE SURGERY, 2013, 29 (04) : 252 - 259
  • [27] Reply: Associations between anterior segment parameters and rotational stability of a plate-haptic toric intraocular lens
    Yao, Yunqian
    Meng, Jiaqi
    He, Wenwen
    Zhang, Keke
    Wei, Ling
    Cheng, Kaiwen
    Lu, Yi
    Zhu, Xiangjia
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (04): : 514 - 515
  • [28] Comparison of Delivery Forces and Wound Dimensions of New Smaller Incision Injector for Plate-Haptic Intraocular Lens
    Lau, George
    Muchhala, Sushant
    Pilon, Andrew
    Kolesnitchenko, Valeri
    Sadri, Ehsan
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [29] Postoperative intraocular lens rotation -: A randomized comparison of plate and loop haptic implants
    Patel, CK
    Ormonde, S
    Rosen, PH
    Bron, AJ
    OPHTHALMOLOGY, 1999, 106 (11) : 2190 - 2195
  • [30] Influence of a Capsular Tension Ring on Capsular Bag Behavior of a Plate Haptic Intraocular Lens An Intraindividual Randomized Trial
    Schartmueller, Daniel
    Roeggla, Veronika
    Schwarzenbacher, Luca
    Meyer, Elias Laurin
    Abela-Formanek, Claudette
    Leydolt, Christina
    Menapace, Rupert
    OPHTHALMOLOGY, 2024, 131 (04) : 445 - 457