Morphometric Analysis and Linear Measurements of the Scala Tympani and Implications in Cochlear Implant Electrodes

被引:5
|
作者
Fujiwara, Rance J. T. [1 ]
Ishiyama, Gail [2 ]
Lopez, Ivan A. [1 ]
Ishiyama, Akira [1 ,3 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Head & Neck Surg, Los Angeles, CA USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Neurol, Los Angeles, CA USA
[3] 10833 Conte Ave,62-132 CHS, Los Angeles, CA 90095 USA
关键词
3-D reconstruction; Cochlear implant; Electrode design; Scala tympani; Temporal bone; TISSUE FORMATION; FIBROUS TISSUE; DIMENSIONS; BONE; INTRACOCHLEAR; RECOGNITION;
D O I
10.1097/MAO.0000000000003848
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
HypothesisThe objective of this study was to perform detailed height and cross-sectional area measurements of the scala tympani in histologic sections of nondiseased human temporal bones and correlate them with cochlear implant electrode dimensions.BackgroundPrevious investigations in scala tympani dimensions have used microcomputed tomography or casting modalities, which cannot be correlated directly with microanatomy visible on histologic specimens.MethodsThree-dimensional reconstructions of 10 archival human temporal bone specimens with no history of middle or inner ear disease were generated using hematoxylin and eosin histopathologic slides. At 90-degree intervals, the heights of the scala tympani at lateral wall, midscala, and perimodiolar locations were measured, along with cross-sectional area.ResultsThe vertical height of the scala tympani at its lateral wall significantly decreased from 1.28 to 0.88 mm from 0 to 180 degrees, and the perimodiolar height decreased from 1.20 to 0.85 mm. The cross-sectional area decreased from 2.29 (standard deviation, 0.60) mm(2) to 1.38 (standard deviation, 0.13) mm(2) from 0 to 180 degrees (p = 0.001). After 360 degrees, the scala tympani shape transitioned from an ovoid to triangular shape, corresponding with a significantly decreased lateral height relative to perimodiolar height. Wide variability was observed among the cochlear implant electrode sizes relative to scala tympani measurements.ConclusionThe present study is the first to conduct detailed measurements of heights and cross-sectional area of the scala tympani and the first to statistically characterize the change in its shape after the basal turn. These measurements have important implications in understanding locations of intracochlear trauma during insertion and electrode design.
引用
收藏
页码:E343 / E349
页数:7
相关论文
共 42 条
  • [21] Experimental Validation of a Three-Dimensional Heat Transfer Model Within the Scala Tympani With Application to Magnetic Cochlear Implant Surgery
    Esmailie, Fateme
    Francoeur, Mathieu
    Ameel, Tim
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (09) : 2821 - 2832
  • [22] Experimental Measurements of Potentials Generated by the Electrodes of a Cochlear Implant in a Phantom
    Tognola, G.
    Pesatori, A.
    Norgia, M.
    Sibella, F.
    Burdo, S.
    Svelto, C.
    Parazzini, M.
    Paglialonga, A.
    Ravazzani, P.
    11TH MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING 2007, VOLS 1 AND 2, 2007, 16 (1-2): : 390 - +
  • [23] Atraumatic Insertion of a Cochlear Implant Pre-Curved Electrode Array by a Robot-Automated Alignment with the Coiling Direction of the Scala Tympani
    Torres, Renato
    Hochet, Baptiste
    Daoudi, Hannah
    Carre, Fabienne
    Mosnier, Isabelle
    Sterkers, Olivier
    Ferrary, Evelyne
    Nguyen, Yann
    AUDIOLOGY AND NEURO-OTOLOGY, 2022, 27 (02) : 148 - 155
  • [24] Morphometric Analysis of Temporal Bone Radiology for Cochlear Implant Candidacy
    Vibhor Malhotra
    Savith Kumar
    Gayathri S. Menon
    Anand Hejjaji Krishnamurthy
    Shankar Medikeri
    Sunil Narayan Dutt
    Indian Journal of Otolaryngology and Head & Neck Surgery, 2024, 76 : 702 - 711
  • [25] Morphometric Analysis of Temporal Bone Radiology for Cochlear Implant Candidacy
    Malhotra, Vibhor
    Kumar, Savith
    Menon, Gayathri S.
    Krishnamurthy, Anand Hejjaji
    Medikeri, Shankar
    Dutt, Sunil Narayan
    INDIAN JOURNAL OF OTOLARYNGOLOGY AND HEAD & NECK SURGERY, 2024, 76 (01) : 702 - 711
  • [26] BEHAVIOR OF TONE-INDUCED COCHLEAR RESPONSES TO POTASSIUM CONCENTRATION ELEVATED BY SCALA TYMPANI PERFUSION AND MONITORED BY ION-SENSITIVE ELECTRODES
    SALT, AN
    STOPP, PE
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 64 : S132 - S132
  • [27] Assessing Cochlear Implant Insertion Angle From an Intraoperative X-ray Using a Rotating 3D Helical Scala Tympani Model
    Giardina, Christopher K.
    Canfarotta, Michael W.
    Thompson, Nicholas J.
    Fitzpatrick, Douglas C.
    Hodge, Sarah E.
    Baker, Jenna
    O'Connell, Brendan P.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (06) : E686 - E694
  • [28] Computed Tomography-Based Measurements of the Cochlear Duct: Implications for Cochlear Implant Pitch Tuning
    Jiam, Nicole T.
    Gilbert, Melanie
    Mo, Jonathan
    Jiradejvong, Patpong
    Limb, Charles J.
    EAR AND HEARING, 2021, 42 (03): : 732 - 743
  • [29] Miniature cochlea: a study of radiological measurements and its implications during the cochlear implant surgery
    Tarabishi M.N.
    Sarwat A.A.
    Kamal Rabie H.M.
    Nada I.M.
    Hakim E.K.
    Hasaballah M.S.
    El-Sharnouby M.M.
    Aziz T.T.A.
    The Egyptian Journal of Otolaryngology, 2016, 32 (3) : 170 - 177
  • [30] Analysis of electrode impedance and its subcomponents for lateral wall, mid-scala, and perimodiolar electrodes in cochlear implants
    Saoji, Aniket A.
    Graham, Madison
    Stein, Amy
    Koka, Kanthaiah
    COCHLEAR IMPLANTS INTERNATIONAL, 2022, 23 (02) : 87 - 94