CT Reconstruction Algorithm and Low Contrast Detectability of Phantom Study: A Systematic Review and Meta-Analysis

被引:0
|
作者
Yusof, Nur Aimi Adibah [1 ]
Karim, Muhammad Khalis Abdul [1 ]
Asikin, Nursyazalina Mohd [1 ]
Paiman, Suriati [1 ]
Kechik, Mohd Mustafa Awang [1 ]
Rahman, Mohd Amiruddin Abdul [1 ]
Noor, Noramaliza Mohd [2 ]
机构
[1] Univ Putra Malaysia, Fac Sci, Dept Phys, Seri Kembangan 43400, Malaysia
[2] Univ Putra Malaysia, Fac Med, Dept Radiol, Seri Kembangan 43400, Malaysia
关键词
Computed tomography; phantom study; filtered back projection; low contrast detectability; iterative reconstruction; radiation; IMAGE QUALITY; ITERATIVE RECONSTRUCTION; COMPUTED-TOMOGRAPHY; SCAN PARAMETERS; PERFORMANCE; RESOLUTION;
D O I
10.2174/1573405618666220816160544
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background For almost three decades, computed tomography (CT) has been extensively used in medical diagnosis, which led researchers to conduct linking of CT dose exposure with image quality. Methods In this study, a systematic review and a meta-analysis study were conducted on CT phantom for resolution study especially based on the low contrast detectability (LCD). Furthermore, the association between the CT parameter such as tube voltage and the type of reconstruction algorithm, the amount of phantom scanning affecting the image quality and the exposure dose were also investigated in this study. We utilize PubMed, ScienceDirect, Google Scholar and Scopus databases to search related published articles from the year 2011 until 2020. The notable keywords comprise "computed tomography", "CT phantom", and "low contrast detectability". Of 52 articles, 20 articles are within the inclusion criteria in this systematic review. Results The dichotomous outcomes were chosen to represent the results in terms of risk ratio as per meta-analysis study. Notably, the noise in iterative reconstruction (IR) reduced by 24%, 33% and 36% with the use of smooth, medium and sharp filters, respectively. Furthermore, adaptive iterative dose reduction (AIDR 3D) improved image quality and the visibility of smaller less dense objects compared to filtered back-projection. Most of the researchers used 120 kVp tube voltage to scan phantom for quality assurance study. Conclusion Hence, optimizing primary factors such as tube potential reduces the dose exposure significantly, and the optimized IR technique could substantially reduce the radiation dose while maintaining the image quality.
引用
收藏
页码:1105 / 1113
页数:9
相关论文
共 50 条
  • [41] Statins and Contrast Induced-AKI: A Systematic Review and Meta-Analysis
    Chittoor, Shriman
    Kohli, Kripa
    Navaneethan, Sankar D.
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2022, 33 (11): : 353 - 354
  • [42] Statins and Contrast-Induced Nephropathy: A Systematic Review and Meta-Analysis
    Briasoulis, Alexandros
    Pala, Mohan
    Telila, Tesfaye
    Merid, Obsinet
    Akintoye, Emmanuel
    Vogiatzi, Georgia
    Oikonomou, Evangelos
    Tousoulis, Dimitris
    CURRENT PHARMACEUTICAL DESIGN, 2017, 23 (46) : 7141 - 7148
  • [43] Lung Cancer Screening with Low-Dose CT in Smokers: A Systematic Review and Meta-Analysis
    Hunger, Theresa
    Wanka-Pail, Eva
    Brix, Gunnar
    Griebel, Juergen
    DIAGNOSTICS, 2021, 11 (06) : NA
  • [44] Objective assessment of low contrast detectability for real CT phantom and in simulated images using a model observer
    Hernandez-Giron, I.
    Geleijns, J.
    Calzado, A.
    Salvado, M.
    Jaemai, R. M. S.
    Veldkamp, W. J. H.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3477 - 3480
  • [45] Effects of various generations of iterative CT reconstruction algorithms on low-contrast detectability as a function of the effective abdominal diameter: A quantitative task-based phantom study
    Viry, Anais
    Aberle, Christoph
    Racine, Damien
    Knebel, Jean-Francois
    Schindera, Sebastian T.
    Schmidt, Sabine
    Becce, Fabio
    Verdun, Francis R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 48 : 111 - 118
  • [46] Ossicular Chain Reconstruction With Titanium Prostheses: A Systematic Review and Meta-Analysis
    Kortebein, Sarah
    Russomando, Alessandra Cataldo
    Greda, Dominik
    Cooper, Matthew
    Ledbetter, Leila
    Kaylie, David
    OTOLOGY & NEUROTOLOGY, 2023, 44 (02) : 107 - 114
  • [47] Free Flap Reconstruction of the Lower Lip: A Systematic Review and Meta-Analysis
    Murray-Douglas, Alexander
    Romeo, Pascalino
    Fox, Carly
    JOURNAL OF RECONSTRUCTIVE MICROSURGERY, 2024,
  • [48] Posterolateral Corner Reconstruction: A Systematic Review and Meta-analysis of Biomechanical Studies
    Zielinski, Kyle P.
    Wieland, Mark D.
    Sequeira, Sean B.
    Gould, Heath P.
    Dreese, James C.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2025,
  • [49] Isolated medial patellofemoral reconstruction outcomes: A systematic review and meta-analysis
    Castagno, Christopher
    Kneedler, Sterling
    Fares, Austin
    Maier, Michael
    Gontre, Gil
    Weiss, William M.
    KNEE, 2023, 44 : 59 - 71
  • [50] Prevention of infection in peripheral arterial reconstruction: A systematic review and meta-analysis
    Stewart, Andrew H.
    Eyers, Paul S.
    Earnshaw, Jonothan J.
    JOURNAL OF VASCULAR SURGERY, 2007, 46 (01) : 148 - 155