Validation of a New Salt-Assisted HS-GC-FID Method for the Determination of Ethanol in the Vitreous Humor

被引:5
|
作者
Musile, Giacomo [1 ]
Pigaiani, Nicola [1 ]
Pasetto, Emma [1 ]
Ballotari, Marco [1 ]
Tagliaro, Franco [1 ,2 ]
Bortolotti, Federica [1 ]
机构
[1] Univ Verona, Dept Diagnost & Publ Hlth, Unit Forens Med, Ple LA Scuro 10, I-37134 Verona, Italy
[2] IM Sechenov First Moscow State Med Univ, Inst Translat Med & Biotechnol, Lab Pharmacokinet & Metabol Anal, Bolshaya Pirogovskaya St, Moscow 119991, Russia
关键词
POSTMORTEM BLOOD; ALCOHOL; ORIGIN;
D O I
10.1093/jat/bkac087
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Headspace gas chromatography with a flame ionization detector (HS-GC-FID) is a well-established approach for determining blood alcohol concentration, including in cadaveric specimens. Although the integrity of blood specimens can be adequately guaranteed after the sampling, the quantification of ethanol in cadaveric blood can be affected by postmortem fermentative phenomena occurring between the time since death and the sampling of biofluids. The vitreous humor is less affected by putrefactive phenomena allowing compound determination and its use as an alternative biological matrix. The present work aimed to develop and validate a method using the salting-out effect and based on HS-GC-FID for the determination of ethanol in the vitreous humor. The reported analytical method is based on a simple vitreous humor pre-treatment consisting of a dilution (1:9) with a solution of 2.5 mol/L K2CO3 and 0.0012 mol/L tert-butanol (internal standard). After 1 min of incubation, part of the specimen evaporated in the headspace (2,000 mu L) is injected into the chromatographic system and analyzed in isothermal mode (40 degrees C), with a chromatographic time of 1.6 min. The method was validated in terms of selectivity, the lowest limit of detection, intraday and total imprecision, and trueness (bias). The determination of ethanol in the vitreous humor and blood was carried out in 75 cases. The correlation between the two matrices was confirmed in 61 cases. However, 14 vitreous humor specimens showed lower ethanol concentrations, and in the related blood specimens, it was possible to identify the signal of n-propanol, a typical product of postmortem microbial fermentation, that justifies the excess of ethanol in the blood specimens.
引用
收藏
页码:E274 / E279
页数:6
相关论文
共 50 条
  • [31] A rapid analytical method based on microwave-assisted extraction for the determination of drugs of abuse in vitreous humor
    Fernandez, Purificacion
    Seoane, Santiago
    Vazquez, Cristina
    Bermejo, Ana M.
    Carro, Antonia M.
    Lorenzo, Rosa A.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 401 (07) : 2177 - 2186
  • [32] A rapid analytical method based on microwave-assisted extraction for the determination of drugs of abuse in vitreous humor
    Purificación Fernández
    Santiago Seoane
    Cristina Vázquez
    Ana M. Bermejo
    Antonia M. Carro
    Rosa A. Lorenzo
    Analytical and Bioanalytical Chemistry, 2011, 401 : 2177 - 2186
  • [33] A salt-assisted graphene oxide aggregation method for the determination of dimethylamine and trimethylamine by ion chromatography with conductivity detection
    Xu, Shu-Xia
    Zhang, Lei-Lei
    Zhang, Hou-Chun
    Zeng, Ying
    Shi, Ze-Ming
    Ni, Shi-Jun
    ANALYTICAL METHODS, 2016, 8 (08) : 1828 - 1835
  • [34] Investigation of Behavior of Forced Degradation of Lidocaine HCl by NMR Spectroscopy and GC-FID Methods: Validation of GC-FID Method for Determination of Related Substance in Pharmaceutical Formulations
    Kadioglu, Yucel
    Atila, Alptug
    Gultekin, Mehmet Serdar
    Alp, Nurdan Alcan
    IRANIAN JOURNAL OF PHARMACEUTICAL RESEARCH, 2013, 12 (04): : 659 - 669
  • [35] METHOD VALIDATION FOR HS-SPME / GC-MS FOR DETERMINATION OF CARBARYL IN BLOOD PLASMA
    de Souza, Francisco Thiago C.
    Camelo, Andre Luiz M.
    Marques, Ricardo de Araujo
    Filho, Luiz Ivando Pires F.
    Pinheiro, Ronald F.
    Longhinotti, Elisane
    Becker, Helena
    QUIMICA NOVA, 2017, 40 (05): : 586 - 593
  • [36] Development and validation of a specific and sensitive GC-FID method for the determination of impurities in 5-chlorovaleroyl chloride
    Tang, Liya
    Kim, Alexandre
    Miller, Scott A.
    Lloyd, David K.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2010, 53 (03) : 309 - 314
  • [37] Determination of Ethanol in Flavor Microcapsules by Ultrasound Assisted Total Vaporization Headspace - Gas Chromatography (TV HS-GC)
    Li, Wei
    Liu, Xue-Yan
    Yu, Kang
    Luo, Ying-Chun
    Nie, Deng-Pan
    Dai, Yi
    Wang, Hao
    ANALYTICAL LETTERS, 2023, 56 (09) : 1478 - 1487
  • [38] Determination of Three Corticosteroids in the Biologic Matrix of Vitreous Humor by HPLC-tandem Mass Spectrometry: Method Development and Validation
    Prieto, Esther
    Vispe, Eugenio
    Otin-Mallada, Sofia
    Garcia-Martin, Elena
    Polo-Llorens, Vicente
    Fraile, Jose M.
    Pablo, Luis E.
    Mayoral, Jose A.
    CURRENT EYE RESEARCH, 2017, 42 (02) : 244 - 251
  • [39] A novel method for the therapeutic drug monitoring of biperiden in plasma by GC-MS using salt-assisted liquid-liquid microextraction
    Kul, Aykut
    Sagirli, Olcay
    CLINICA CHIMICA ACTA, 2023, 543
  • [40] Optimization and Validation of a GC-FID/QuEChERS Method for Quantitative Determination of Spiromesifen Residues in Tomato Fruits, Leaves and Soil Matrices
    Kadhum, Hanan Abdulsalam
    Hammood, Mohammad K.
    Arif, Mokhtar Abdulsattar
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY RESEARCH, 2024, 11 (04): : 447 - 461