Biomechanical Analysis of Combined Medial Calcar and Lateral Locked Plating Versus Isolated Lateral Locked Plating of Proximal Humerus Fractures

被引:1
|
作者
Kotler, Joshua A. [1 ,5 ]
Zuppke, Julia N. [2 ]
Abraham, Vivek M. [2 ]
Sanville, Jennifer A. [2 ]
Nabet, Austin C. [2 ]
Carofino, Bradley [3 ]
Gardner, Michael J. [4 ]
Smith, Christopher S. [2 ]
机构
[1] III Marine Expeditionary Force, Marine Logist Grp 3, Med Battal 3, Okinawa, Japan
[2] Naval Med Readiness & Training Ctr, Bone & Joint Sports Med Inst, Portsmouth, VA USA
[3] Atlantic Orthopaed Specialists, Virginia Beach, VA USA
[4] Stanford Univ, Dept Orthopaed Surg, Redwood City, CA USA
[5] III MEF, MLG 3, Med Battal 3, Unit 38404, FPO,AP 96604-8404, Okinawa, Japan
关键词
proximal humerus fracture; internal fixation; biomechanics; locking plates; bending; torsion; medial calcar plating; FIXATION; LOCKING; SUPPORT; EPIDEMIOLOGY; REDUCTION; PARALLEL; FRAGMENT;
D O I
10.1097/BOT.0000000000002619
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective:Medial calcar buttress plating combined with lateral locked plating is biomechanically tested against isolated lateral locked plating in synthetic humeri models for the treatment of proximal humerus fractures.Methods:Proximal humerus fractures (OTA/AO type 11-A2.1) were manufactured in 10 pairs of Sawbones humeri models (Sawbones, Pacific Research Laboratories, Vashon Island, WA). Specimen were randomly assigned and instrumented with either medial calcar buttress plating combined with lateral locked plating (CP) or isolated lateral locked plating (LP). Nondestructive torsional and axial load tests were performed to evaluate construct stiffness. Large-cycle axial tests were conducted followed by destructive ramp-to-failure tests. Cyclic stiffness was compared in both nondestructive and ultimate failure loads. Failure displacement was recorded and compared between groups.Results:The addition of medial calcar buttress plating to lateral locked plating constructs significantly increased the axial (P < 0.01) and torsional (P < 0.01) stiffness of the construct compared with isolated lateral locked plating by 95.56% and 37.46%, respectively. All models demonstrated greater axial stiffness (P < 0.01) after 5000 cycles of axial compression, not dependent on the fixation method. During destructive testing, the CP construct withstood 45.35% larger load (P < 0.01) and congruently exhibited 58% less humeral head displacement (P = 0.02) before failure when compared with the LP construct.Conclusion:This study demonstrates the biomechanical superiority of medial calcar buttress plating when combined with lateral locked plating as compared with isolated lateral locked plating of OTA/AO type 11-A2.1 proximal humerus in synthetic humeri models.
引用
收藏
页码:E355 / E360
页数:6
相关论文
共 50 条
  • [41] Reliability of locked plating in tibial plateau fractures with a medial component
    Ehlinger, M.
    Rahme, M.
    Moor, B. -K.
    Di Marco, A.
    Brinkert, D.
    Adam, P.
    Bonnomet, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2012, 98 (02) : 173 - 179
  • [42] Is retrograde nailing superior to lateral locked plating for complete articular distal femur fractures?
    Vemulapalli, K. Chandra
    Pechero, Guillermo R.
    Warner, Stephen J.
    Achor, Timothy S.
    Gary, Joshua L.
    Munz, John W.
    Choo, Andrew M.
    Prasarn, Mark L.
    Routt, Milton L. Chip, Jr.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (02): : 640 - 644
  • [43] Nailing of proximal ulna fractures: biomechanical comparison of a new locked nail with angular stable plating
    Johannes Christof Hopf
    Tobias Eckhard Nowak
    Dorothea Mehler
    Charlotte Arand
    Dominik Gruszka
    Pol Maria Rommens
    European Journal of Trauma and Emergency Surgery, 2021, 47 : 795 - 802
  • [44] Treatment of Peritrochanteric Femur Fractures With Proximal Femur Locked Plating
    Medda, Suman
    Sullivan, Rashad J.
    Marquez-Lana, Alejandro
    Araiza, Edgar T.
    Pilson, Holly T.
    Halvorson, Jason J.
    Carroll, Eben A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2019, 33 (07) : 341 - 345
  • [45] Non-union in lateral locked plating for distal femoral fractures: A systematic review
    Wang, Michael Tianfeng
    An, Vincent V. G.
    Sivakumar, Brahman Shankar
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2019, 50 (11): : 1790 - 1794
  • [46] Nailing of proximal ulna fractures: biomechanical comparison of a new locked nail with angular stable plating
    Hopf, Johannes Christof
    Nowak, Tobias Eckhard
    Mehler, Dorothea
    Arand, Charlotte
    Gruszka, Dominik
    Rommens, Pol Maria
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2021, 47 (03) : 795 - 802
  • [47] The results of tension band rotator cuff suture fixation of locked plating of displaced proximal humerus fractures
    Shukla, Dave R.
    McAnany, Steven
    Pean, Christian
    Overley, Samuel
    Lovy, Andrew
    Parsons, Bradford O.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (02): : 474 - 480
  • [48] Dorsal Plating, Lateral Plating, and Intramedullary Screw Fixation of Extra-Articular Proximal Phalangeal Fractures: A Cadaveric Biomechanical
    Sivakumar, Brahman
    Graham, David J.
    Clarke, Elizabeth
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2024, 49 (03): : 247 - 252
  • [49] Biomechanical evaluation of locked plating fixation for unstable femoral neck fractures
    Bliven, E.
    Sandriesser, S.
    Augat, P.
    von Rueden, C.
    Hackl, S.
    BONE & JOINT RESEARCH, 2020, 9 (06): : 314 - 321
  • [50] Biomechanical analysis of a novel plating for intra-articular distal humerus fractures: combined anteromedial and anterolateral plating
    Wei, Libiao
    Ling, Ming
    An, Zhiquan
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)