Intraoperative electron beam intercomparison of 6 sites using mailed thermoluminescence dosimetry: Absolute dose and energy

被引:0
|
作者
Dries, Wim [1 ]
Petoukhova, Anna [2 ,7 ]
Hertsens, Nicolas [3 ]
Stevens, Piet [4 ]
Jarbinet, Valerie [5 ]
Bimmel-Nagel, Cathryn Huibregtse [1 ]
Weterings, Jan [1 ]
van Wingerden, Ko [2 ]
Bauwens, Charlotte [3 ]
Vanreusel, Verdi [4 ]
Simon, Stephane [6 ]
机构
[1] Catharina Hosp, Eindhoven, Netherlands
[2] Haaglanden Med Ctr, Dept Med Phys, Leidschendam, Netherlands
[3] AZ Groeninge, Dept Radiotherapy, Kortrijk, Belgium
[4] GZA Hosp, Antwerp, Belgium
[5] CHU Tivoli, La Louviere, Belgium
[6] Jules Bordet Inst, Brussels, Belgium
[7] Haaglanden Med Ctr, Dept Med Phys, POB 432, NL-2501 CK The Hague, Netherlands
关键词
IORT; Electron IORT (IOERT); Intraoperative electron beam dosimetry; Intercomparison; Thermoluminescence dosimetry; Absolute dose and energy; TASK FORCE/ACROP RECOMMENDATIONS; ESTRO/ACROP IORT RECOMMENDATIONS; RADIATION-THERAPY; AUDITS; TLD;
D O I
10.1016/j.ejmp.2024.103302
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In 2018, the Netherlands Commission on Radiation Dosimetry subcommittee on IORT initiated a limited intercomparison of electron IORT (IOERT) in Belgium and The Netherlands. The participating institutions have enough variability in age, type of equipment, and in dose calibration protocols. Methods: In this study, three types of IOERT-dedicated mobile accelerators were represented: Mobetron 2000, LIAC HWL and LIAC. Mobetron produces electron beams with energies of 6, 9 and 12 MeV, while LIAC HWL and LIAC can deliver 6, 8, 10 and 12 MeV electron beams. For all energies, the reference beam (10 cm diameter, 0 incidence) and 5 cm diameter beams were measured, as these smaller beams are used more frequently in clinic. The mailed TLD service from the Radiation Dosimetry Services (RDS, Houston, USA) has been used. Following RDS' standard procedures, each beam was irradiated to 300 cGy at d(max )with TLDs around d(max) and around depth of 50 % dose (R50). Absolute dose at 100 % and beam energy, expressed as R50, could be verified in this way. Results: All absolute doses and energies under reference conditions were well within RDS-stated uncertainties: dose deviations were <5 % and deviations in R50 were <5 mm. For the small 5 cm beams, all results were also within acceptance levels except one absolute dose value. Deviations were not significantly dependent on manufacturer, energy, diameter and calibration protocol. Conclusions: All absolute dose values, except one of a non-reference beam, and all energy values were well within the measurement accuracy of RDS TLDs.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Assessment of out-of-field radiation doses for high dose per pulse intraoperative electron beam radiotherapy using TLD-100
    Aghdam, Seyed Rashid Hosseini
    Aghamiri, Seyed Mahmoud Reza
    Siavashpour, Zahra
    Malekie, Shahryar
    Dashtipoor, Mohammad Reza
    RADIATION PHYSICS AND CHEMISTRY, 2023, 204
  • [32] Absolute dosimetry of a 1.5 T MR-guided accelerator-based high-energy photon beam in water and solid phantoms using Aerrow
    Renaud, James
    Sarfehnia, Arman
    Bancheri, Julien
    Seuntjens, Jan
    MEDICAL PHYSICS, 2020, 47 (03) : 1291 - 1304
  • [33] Monte carlo calculation of the energy spectrum of a 6 MeV electron beam using penetration and energy loss of positrons and electrons code
    Apaza Veliz, Danny Giancarlo
    Wilches Visbal, Jorge Homero
    Chen Abrego, Felipe
    Vega Ramirez, Jose Luis
    JOURNAL OF MEDICAL PHYSICS, 2020, 45 (02) : 116 - 122
  • [34] A new add-on electron multi-leaf collimation (EMLC): Dosimetry and dose calculation using a new beam model with Voxel Monte Carlo
    Vatanen, T.
    Traneus, E.
    Lahtinen, T.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S166 - S166
  • [35] Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer
    Ciocca, M
    Piazzi, V
    Lazzari, R
    Vavassori, A
    Luini, A
    Veronesi, P
    Galimberti, V
    Intra, M
    Guido, A
    Tosi, G
    Veronesi, U
    Orecchia, R
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 (02) : 213 - 216
  • [36] Contribution of Cerenkov radiation in high-energy x-ray and electron beam film dosimetry using water-substitute phantoms
    Fujisaki, T
    Saitoh, H
    Hiraoka, T
    Kuwabara, A
    Abe, S
    Inada, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (06): : N105 - N109
  • [37] INTRAOPERATIVE ELECTRON-BEAM RADIATION-THERAPY - TECHNIQUE, DOSIMETRY, AND DOSE SPECIFICATION - REPORT OF TASK-FORCE 48 OF THE RADIATION-THERAPY COMMITTEE, AMERICAN-ASSOCIATION-OF-PHYSICISTS-IN-MEDICINE
    PALTA, JR
    BIGGS, PJ
    HAZLE, JD
    HUQ, MS
    DAHL, RA
    OCHRAN, TG
    SOEN, J
    DOBELBOWER, RR
    MCCULLOUGH, EC
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (03): : 725 - 746
  • [38] Study the sensitivity of Dose Calculation in Prism Treatment Planning System Using Monte Carlo Simulation of 6 MeV Electron Beam
    Hardiansyah, D.
    Male, S.
    Haryanto, F.
    4TH INTERNATIONAL CONFERENCE ON ADVANCES IN NUCLEAR SCIENCE AND ENGINEERING (ICANSE 2013), 2014, 1615 : 213 - 223
  • [39] In vivo dosimetry using EBT3 and EBT-XD radiochromic films for high-energy electron beam in hypofractionation keloid radiotherapy
    Lin, Yu -Fang
    Hsieh, Chen-Hsi
    Tien, Hui-Ju
    Lai, Lu-Han
    Hsu, Shih-Ming
    Shueng, Pei-Wei
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [40] Landau distribution-based regularized algorithm for reconstruction of electron beam energy spectrum using depth dose distributions in targeted materials
    Zolotov, Sergei A.
    Bliznyuk, Ulyana A.
    Antipina, Natalia A.
    Borshchegovskaya, Polina Yu
    Chernyaev, Alexander P.
    Ipatova, Victoria S.
    Nikitchenko, Alexander D.
    Toropygina, Maria, I
    Nikolaeva, Anna A.
    Studenikin, Felix R.
    PHYSICA SCRIPTA, 2025, 100 (04)