Primary activity standardization of 18F for the calibration of the KRISS ionization chamber

被引:0
|
作者
Agusbudiman, Agung [1 ,2 ]
Lee, Kyoung Beom [1 ,2 ]
Lee, Jong Man [1 ,2 ]
Hwang, Sanghoon [1 ,2 ]
Kim, Byoung Chul [1 ]
Heo, Dong Hye [1 ]
机构
[1] Korea Res Inst Stand & Sci KRISS, Ionizing Radiat Metrol Grp, 267 Gajeong Ro, Daejeon 34113, South Korea
[2] Univ Sci & Technol UST, Dept Sci Measurement, 217 Gajeong Ro, Daejeon 34113, South Korea
关键词
Fluorine-18; Activity standardization; 4 pi beta(LS)-gamma coincidence counting; Triple to double coincidence ratio; Ionization chamber;
D O I
10.1007/s10967-023-09099-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A primary activity standardization of F-18 was carried out at KRISS by means of the 4 pi beta(LS)-gamma coincidence counting with a combined standard uncertainty of 0.31%. The activity value was compared with those obtained from the TDCR method, which was performed as a confirmatory measurement of the standardization. The final massic activity value of (13.186 +/- 0.041) MBq.g(-1) at k = 1 was found and used as the reference value to calibrate the KRISS reference ionization chamber employed as the secondary measurement standard of F-18. The calibration result was then compared with the existing calibration factor that was obtained from the efficiency curve of the reference ionization chamber.
引用
收藏
页码:5259 / 5265
页数:7
相关论文
共 50 条
  • [1] Primary activity standardization of 18F for the calibration of the KRISS ionization chamber
    Agung Agusbudiman
    Kyoung Beom Lee
    Jong Man Lee
    Sanghoon Hwang
    Byoung Chul Kim
    Dong Hye Heo
    Journal of Radioanalytical and Nuclear Chemistry, 2023, 332 : 5259 - 5265
  • [2] Calibration of ionization chamber for 18F and 68Ga
    da Silva, Carlos J.
    de Oliveira, Estela M.
    Iwahara, A.
    Delgado, Jose U.
    Poledna, R.
    de Oliveira, Antonio E.
    Moreira, Denise S.
    da Silva, Ronaldo L.
    Gomes, Regio dos Santos
    de Veras, Eduardo V.
    APPLIED RADIATION AND ISOTOPES, 2014, 87 : 188 - 191
  • [3] Activity standardisation of 18F and ionisation chamber calibration for nuclear medicine
    Schrader, H.
    Klein, R.
    Kossert, K.
    APPLIED RADIATION AND ISOTOPES, 2007, 65 (05) : 581 - 592
  • [4] A new NIST primary standardization of 18F
    Fitzgerald, R.
    Zimmerman, B. E.
    Bergeron, D. E.
    Cessna, J. C.
    Pibida, L.
    Moreira, D. S.
    APPLIED RADIATION AND ISOTOPES, 2014, 85 : 77 - 84
  • [5] 18F primary standard at ENEA-INMRI by three absolute techniques and calibration of a well-type IG11 ionization chamber
    Capogni, Marco
    Carconi, Pierluigi
    De Felice, Pierino
    Fazio, Aldo
    APPLIED RADIATION AND ISOTOPES, 2016, 109 : 410 - 413
  • [6] Calibration of KRISS reference ionization chamber for key comparison of 99mTc measurement
    Lee, Jong-Man
    Lee, K. B.
    Lee, S. H.
    Park, Tae Soon
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (09) : 1853 - 1855
  • [7] Calibration of the KRISS reference ionization chamber for certification of 222Rn gaseous sources
    Lee, J. M.
    Lee, K. B.
    Lee, S. H.
    Oh, P. J.
    Park, T. S.
    Kim, B. C.
    Lee, M. S.
    APPLIED RADIATION AND ISOTOPES, 2013, 81 : 230 - 232
  • [8] Standardization of 18F by digital coincidence counting
    Kim, Eun-Joo
    Park, Tae Soon
    Lee, Jong Man
    Hahm, Don Shik
    Lee, K. B.
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (09) : 2103 - 2106
  • [9] Standardization of 18F by coincidence and LSC methods
    Roteta, Miguel
    Garcia-Torano, Eduardo
    Barquero, Leonor Rodriguez
    APPLIED RADIATION AND ISOTOPES, 2006, 64 (10-11) : 1199 - 1202
  • [10] Extended utilization of a 68Ge/68Ga reference source as a mock 18F source in the ionization chamber calibration
    Yamada, Takahiro
    Ishizu, Hidetake
    Arai, Kasumi
    Shiina, Takuya
    APPLIED RADIATION AND ISOTOPES, 2018, 134 : 89 - 94