Calibration of ionization chamber for 18F and 68Ga

被引:4
|
作者
da Silva, Carlos J. [1 ]
de Oliveira, Estela M. [1 ]
Iwahara, A. [1 ]
Delgado, Jose U. [1 ]
Poledna, R. [1 ]
de Oliveira, Antonio E. [1 ]
Moreira, Denise S. [1 ]
da Silva, Ronaldo L. [1 ]
Gomes, Regio dos Santos [1 ]
de Veras, Eduardo V. [1 ]
机构
[1] Comissao Nacl Energia Nucl, Inst Radioprotecao & Dosimetria, Lab Nacl Metrol Radiacoes Ionizantes, BR-22783127 Rio De Janeiro, Brazil
关键词
Standardization; Anti-coincidence counting; Extending dead time; EXTENDING DEAD-TIME; STANDARDIZATION;
D O I
10.1016/j.apradiso.2013.11.137
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In order to maintain the results of primary activity standardizations carried out in 2011 the LNMRI has determined the calibration factors for a pressurized 4 pi-ionization chamber for the nuclides F-18 and Ga-68. This ionization chamber is coupled to a 6517A Keithley electrometer which is controlled by a homemade LabVIEW program. This paper will describe the main issues related to the calibration of an ionization chamber system for positron emitters and short half-life radionuclides such as timing, current measurement, background, decay, and Ra-226 check source measurements. (c) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:188 / 191
页数:4
相关论文
共 50 条
  • [1] Extended utilization of a 68Ge/68Ga reference source as a mock 18F source in the ionization chamber calibration
    Yamada, Takahiro
    Ishizu, Hidetake
    Arai, Kasumi
    Shiina, Takuya
    [J]. APPLIED RADIATION AND ISOTOPES, 2018, 134 : 89 - 94
  • [2] 68Ga or 18F for Prostate Cancer Imaging
    Kesch, Claudia
    Kratochwil, Clemens
    Mier, Walter
    Kopka, Klaus
    Giesel, Frederik L.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (05) : 687 - 688
  • [3] Performance evaluation of Cerenkov luminescence imaging: a comparison of 68Ga with 18F
    J. olde Heuvel
    B. J. de Wit-van der Veen
    K. N. Vyas
    D. S. Tuch
    M. R. Grootendorst
    M. P. M. Stokkel
    C. H. Slump
    [J]. EJNMMI Physics, 6
  • [4] New 18F and 68Ga tracers for PET myocardial perfusion imaging
    Ioppolo, J.
    Dallen, R.
    Hill, L.
    Rendina, L.
    [J]. INTERNAL MEDICINE JOURNAL, 2022, 52 : 9 - 9
  • [5] Primary activity standardization of 18F for the calibration of the KRISS ionization chamber
    Agusbudiman, Agung
    Lee, Kyoung Beom
    Lee, Jong Man
    Hwang, Sanghoon
    Kim, Byoung Chul
    Heo, Dong Hye
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2023, 332 (12) : 5259 - 5265
  • [6] Primary activity standardization of 18F for the calibration of the KRISS ionization chamber
    Agung Agusbudiman
    Kyoung Beom Lee
    Jong Man Lee
    Sanghoon Hwang
    Byoung Chul Kim
    Dong Hye Heo
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 2023, 332 : 5259 - 5265
  • [7] Performance evaluation of Cerenkov luminescence imaging: a comparison of 68Ga with 18F
    Heuvel, J. Olde
    de Wit-van der Veen, B. J.
    Vyas, K. N.
    Tuch, D. S.
    Grootendorst, M. R.
    Stokkel, M. P. M.
    Slump, C. H.
    [J]. EJNMMI PHYSICS, 2019, 6 (01)
  • [8] Development of a Traceable Calibration Methodology for Solid 68Ge/68Ga Sources Used as a Calibration Surrogate for 18F in Radionuclide Activity Calibrators
    Zimmerman, Brian E.
    Cessna, Jeffrey T.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (03) : 448 - 453
  • [9] 18F, 11C and 68Ga in small animal PET imaging
    Muellauer, J.
    Willimayer, R.
    Goertzen, A. L.
    Wanek, T.
    Langer, O.
    Birkfellner, W.
    Kuntner, C.
    [J]. NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2013, 52 (06): : 250 - +
  • [10] Target activity repeatability in a clinically relevant phantom: 18F vs 68Ga
    Silosky, Michael
    Patten, Luke
    Chin, Bennett
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2021, 62