摘要:
A primary activity standardization of F-18 was carried out at KRISS by means of the 4 pi beta(LS)-gamma coincidence counting with a combined standard uncertainty of 0.31%. The activity value was compared with those obtained from the TDCR method, which was performed as a confirmatory measurement of the standardization. The final massic activity value of (13.186 +/- 0.041) MBq.g(-1) at k = 1 was found and used as the reference value to calibrate the KRISS reference ionization chamber employed as the secondary measurement standard of F-18. The calibration result was then compared with the existing calibration factor that was obtained from the efficiency curve of the reference ionization chamber.