Mucosal Impedance Spectroscopy for Objective Real-time Assessment of Mucosal Health

被引:1
|
作者
Arora, Priyanka [1 ]
Singh, Jaspreet [2 ]
Jena, Anuraag [1 ]
Kumar, Surinder [1 ]
Sardana, Viren [3 ]
Sarkar, Siddhartha [3 ]
Kaman, Lileswar [4 ]
Behera, Arunanshu [4 ]
Dahiya, Divya [4 ]
Nada, Ritambhra [5 ]
Tandup, Cherring [4 ]
Jatana, H. S. [2 ]
Dutta, Usha [1 ,6 ]
机构
[1] Post Grad Inst Med Educ & Res, Dept Gastroenterol, Chandigarh, India
[2] Dept MEMS, Semicond Lab, Mohali, India
[3] Cent Sci Instrument Org, Chandigarh, India
[4] Post Grad Inst Med Educ & Res, Dept Surg, Chandigarh, India
[5] Post Grad Inst Med Educ & Res, Dept Histopathol, Chandigarh, India
[6] Postgrad Inst Med Educ & Res, Dept Gastroenterol, Chandigarh 160012, India
来源
TECHNIQUES AND INNOVATIONS IN GASTROINTESTINAL ENDOSCOPY | 2023年 / 25卷 / 03期
关键词
Impedance; Malignancy; Inflammation; Endoscopic catheter; TIGHT JUNCTION PROTEINS; DIAGNOSIS;
D O I
10.1016/j.tige.2023.03.007
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
BACKGROUND AND AIMS: There is a need for a real-time objective tool to assess the mucosal health of the gastrointestinal (GI) tract during endoscopy. Our aim was to develop, design, and validate a lowcost mucosal impedance (MI) device and determine its role in differentiating diseased mucosa from normal mucosa. METHODS: A biocompatible catheter was designed and developed after multiple iterations. It was validated with a commercially available catheter and histopathological analysis in a blinded manner. Patients undergoing resection of the GI tract were recruited after consent, and the resected specimens were analyzed ex vivo for MI within 10 minutes of resection. An average of 3 MI readings of the visually diseased segment and adjacent normal segment were analyzed. MI values of diseased and non-diseased mucosa were compared. RESULTS: The in-house catheter was validated with a commercially available impedance measuring device. It showed a high degree of positive correlation (rho = 0.616; P < 0.001). Two hundred and thirtytwo patients (age 46 </n> 15 years; 68% females) (180 inflammatory and 52 malignant pathology) who were undergoing abdominal surgery were enrolled. The median impedance value of diseased segments was significantly lower than that of the adjacent normal segments of gut in 130 paired samples [1832 (727)Q vs 2604(1295)Q; P < 0.001]. The MI value of segments containing malignant tissue (n = 50) and inflamed tissue (n = 80) was significantly lower than the MI value of adjacent normal segments of the GI tract [1880(977)Q vs 2583(1431)Q; P < 0.001 and 1787(557)Q vs 2515(1244)Q; P < 0.001, respectively]. There was a 24% reduction in visually diseased segments [median reduction 712(661)Q] from adjacent normal segments. A biocompatible endoscopic catheter (3-mm diameter) has been developed and tested in 3 patients and was found to differentiate diseased from normal mucosa. CONCLUSION: Impedance spectroscopy is an effective real-time, simple, objective tool to differentiate diseased gut mucosa from healthy mucosa.
引用
收藏
页码:228 / 235
页数:8
相关论文
共 50 条
  • [41] Five mucosal transcripts of interest in ulcerative colitis identified by quantitative real-time PCR: a prospective study
    Anders Eriksson
    Carl-Fredrik Flach
    Anders Lindgren
    Eva Kvifors
    Stefan Lange
    BMC Gastroenterology, 8
  • [42] Bio-Impedance Sensor for Real-Time Artery Diameter Waveform Assessment
    Al-harosh, Mugeb
    Yangirov, Marat
    Kolesnikov, Dmitry
    Shchukin, Sergey
    SENSORS, 2021, 21 (24)
  • [43] Real-time assessment of cytotoxicity by impedance measurement on a 96-well plate
    Ceriotti, L.
    Ponti, J.
    Broggi, F.
    Kob, A.
    Drechsler, S.
    Thedinga, E.
    Colpo, P.
    Sabbioni, E.
    Ehret, R.
    Rossi, F.
    SENSORS AND ACTUATORS B-CHEMICAL, 2007, 123 (02) : 769 - 778
  • [44] Five mucosal transcripts of interest in ulcerative colitis identified by quantitative real-time PCR: a prospective study
    Eriksson, Anders
    Flach, Carl-Fredrik
    Lindgren, Anders
    Kvifors, Eva
    Lange, Stefan
    BMC GASTROENTEROLOGY, 2008, 8 (1)
  • [45] Real-time monitoring of nasal mucosal pH during carbon dioxide stimulation: Implications for stimulus dynamics
    Shusterman, D
    Avila, PC
    CHEMICAL SENSES, 2003, 28 (07) : 595 - 601
  • [46] Real-Time In Vivo Imaging of Early Mucosal Changes during Ischemia-Reperfusion in Human Jejunum
    Grootjans, Joep
    Hameeteman, Wim
    Masclee, Ad A.
    van Dam, Ronald M.
    Buurman, Wim A.
    Dejong, Cornelis H. C.
    PLOS ONE, 2012, 7 (06):
  • [47] Zinc Oxidation in Dilute Alkaline Solutions Studied by Real-Time Electrochemical Impedance Spectroscopy
    Ko, Younghoon
    Park, Su-Moon
    JOURNAL OF PHYSICAL CHEMISTRY C, 2012, 116 (13): : 7260 - 7268
  • [48] Simultaneous impedance spectroscopy and fluorescence microscopy for the real-time monitoring of the response of cells to drugs
    Parviz, M.
    Gaus, K.
    Gooding, J. J.
    CHEMICAL SCIENCE, 2017, 8 (03) : 1831 - 1840
  • [49] Real-Time Electrochemical Impedance Spectroscopy Diagnosis of the Marine Solid Oxide Fuel Cell
    Nakajima, Hironori
    Kitahara, Tatsumi
    7TH EUROPEAN THERMAL-SCIENCES CONFERENCE (EUROTHERM2016), 2016, 745
  • [50] Real-time monitoring of scale formation in reverse osmosis using electrical impedance spectroscopy
    Hu, Zhixin
    Antony, Alice
    Leslie, Greg
    Le-Clech, Pierre
    JOURNAL OF MEMBRANE SCIENCE, 2014, 453 : 320 - 327