Bayesian estimation to deconvolute single-particle ICP-MS data with a mixed Poisson distribution

被引:1
|
作者
Suzuki, Yoshinari [1 ]
Kondo, Midori [1 ]
Harimoto, Masae [1 ]
Okamoto, Yusuke [1 ]
Tanaka, Yu-ki [2 ]
Ogra, Yasumitsu [2 ]
Akiyama, Hiroshi [1 ,3 ]
机构
[1] Natl Inst Hlth Sci, Div Foods, 3-25-26 Tonomachi,Kawasaki Ku, Kawasaki, Kanagawa 2109501, Japan
[2] Chiba Univ, Grad Sch Pharmaceut Sci, 1-8-1 Inohana,Chuo Ku, Chiba, Chiba 2608675, Japan
[3] Hoshi Univ, Dept Analyt Chem, 2-4-41 Ebara,Shinagawa Ku, Tokyo 1428501, Japan
关键词
PLASMA-MASS SPECTROMETRY; TRANSPORT EFFICIENCY; DATA-ACQUISITION; NANOPARTICLES; TIME; CAPABILITIES; RESOLUTION;
D O I
10.1039/d3ja00220a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Single-particle ICP-MS (spICP-MS) is an established method for the determination of inorganic nanoparticle (NP) mass distributions and particle number concentrations. However, spICP-MS is not applicable to some cases, especially cases that require distinguishing signals from dissolved ions and signals from relatively small NPs. To deconvolute spICP-MS data, which is obtained by setting the dwell time similar to the particle event duration time, a Bayesian estimation method was developed for spICP-MS analysis using silver (Ag) and silica (SiO2) NPs. The signal distributions of the spICP-MS data were parameterised using a Bayesian estimation method on the assumption that they could be described by mixed Poisson distributions. Analytical results were then compared to results obtained with conventional criteria. When the instrument parameters were set so that the particle-event duration was within 2 readings and hence did not deviate from the assumptions of the current Bayesian model, better estimation results could be obtained with the Bayesian estimation method than with methods based on conventional criteria, especially for a sample with high particle number concentration. Furthermore, applying the specific informative prior distribution enabled us to obtain reasonable estimation results, even when the signal counts were 6 or less and the background counts were high. Because appropriate NP information was obtained for Ag-NP and SiO2-NP, the Bayesian estimation method can be universally adopted with inorganic NPs detectable by ICP-MS. Model parameters (mean particle-event intensity and frequency, etc.) were directly estimated using the Bayesian method without summing particle-event intensities. Model parameters were transformed to NP information.
引用
收藏
页码:190 / 203
页数:14
相关论文
共 50 条
  • [41] Two-Dimensional Cytometry Platform for Single-Particle/Cell Analysis with Laser-Induced Fluorescence and ICP-MS
    Wu, Chengxin
    Wei, Xing
    Men, Xue
    Zhang, Xuan
    Yu, Yong-Liang
    Xu, Zhang-Run
    Chen, Ming-Li
    Wang, Jian-Hua
    ANALYTICAL CHEMISTRY, 2021, 93 (23) : 8203 - 8209
  • [42] Can ultrafast single-particle analysis using ICP-MS affect the detection limit? Case study: Silver nanoparticles
    Mestek, Oto
    Loula, Martin
    Kana, Antonin
    Vosmanska, Magda
    TALANTA, 2020, 210
  • [43] Statistical properties of spikes in single particle ICP-MS time scans
    Peyneau, Pierre-Emmanuel
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (12) : 2683 - 2690
  • [44] Single Particle ICP-MS: Advances toward routine analysis of nanomaterials
    Montano, Manuel D.
    Olesik, John W.
    Barber, Angela G.
    Challis, Katie
    Ranville, James F.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (19) : 5053 - 5074
  • [45] Silver nanoparticle characterization using single particle ICP-MS (SP-ICP-MS) and asymmetrical flow field flow fractionation ICP-MS (AF4-ICP-MS)
    Mitrano, Denise M.
    Barber, Angela
    Bednar, Anthony
    Westerhoff, Paul
    Higgins, Christopher P.
    Ranville, James F.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (07) : 1131 - 1142
  • [46] A systematic evaluation of Flow Field Flow Fractionation and single-particle ICP-MS to obtain the size distribution of organo-mineral iron oxyhydroxide colloids
    Moens, Claudia
    Waegeneers, Nadia
    Fritzsche, Andreas
    Nobels, Peter
    Smolders, Erik
    JOURNAL OF CHROMATOGRAPHY A, 2019, 1599 : 203 - 214
  • [47] Determination of silver nanoparticles in cosmetics using single particle ICP-MS
    Věra Kantorová
    Martin Loula
    Antonín Kaňa
    Oto Mestek
    Chemical Papers, 2021, 75 : 5895 - 5905
  • [48] Density Determination of Au Nanoparticles Using Single Particle ICP-MS
    Joo, Nayoung
    Lim, H. B.
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2019, 40 (11): : 1087 - 1092
  • [49] Peak bordering for ultrafast single particle analysis using ICP-MS
    Kana, Antonin
    Loula, Martin
    Koplik, Richard
    Vosmanska, Magda
    Mestek, Oto
    TALANTA, 2019, 197 : 189 - 198
  • [50] An interactive Python']Python-based data processing platform for single particle and single cell ICP-MS
    Lockwood, Thomas E.
    Gonzalez de Vega, Raquel
    Clases, David
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2021, 36 (11) : 2536 - 2544