Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone

被引:0
|
作者
Tomazini, Zan [1 ]
Gersak, Gregor [1 ]
Begus, Samo [1 ]
机构
[1] Univ Ljubljana, Fac Elect Engn, Ljubljana 1000, Slovenia
关键词
RF microphone; blood pressure; low frequency; oscillometric; BP simulator; CALIBRATION; HYPERTENSION;
D O I
10.3390/s23198340
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards.