Imaging Function and Relative Light Transmission of Explanted Opacified Hydrophilic Acrylic Intraocular Lenses

被引:4
|
作者
Naujokaitis, Tadas [1 ]
Khoramnia, Ramin [1 ]
Labuz, Grzegorz [1 ]
Choi, Chul Young [1 ,2 ]
Auffarth, Gerd U. [1 ]
Tandogan, Tamer [1 ,3 ]
机构
[1] Heidelberg Univ, Dept Ophthalmol, David J Apple Ctr Vis Res, D-69120 Heidelberg, Germany
[2] Sungkyunkwan Univ, Sch Med, Kangbuk Samsung Hosp, Dept Ophthalmol, Seoul 03181, South Korea
[3] Augenklin Pallas, CH-4600 Olten, Switzerland
关键词
intraocular lens; hydrophilic; opacification; calcification; optical quality; modulation transfer function; light transmission; OPTICAL-PERFORMANCE; OPACIFICATION; CALCIFICATION; STRAYLIGHT; EXCHANGE; TRANSMITTANCE; SCATTERING; OUTCOMES;
D O I
10.3390/diagnostics13101804
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the influence of intraocular lens (IOL) opacification on the optical performance of explanted hydrophilic acrylic IOLs. We performed a laboratory analysis of 32 Lentis LS-502-1 (Oculentis GmbH, Berlin, Germany) IOLs, explanted due to opacification, in comparison with six clear unused samples of the same IOL model. Using an optical bench setup, we obtained modulation transfer function (MTF), Strehl ratio, two-dimensional MTF, and United States Air Force (USAF) chart images. In addition, we assessed light transmission through the IOLs. The MTF values of opacified IOLs at 3-mm aperture were similar to those of clear lenses, with the median (interquartile range) values of 0.74 (0.01) vs. 0.76 (0.03) at the spatial frequency of 50 line pairs per millimeter in clear and opacified IOLs, respectively. The Strehl ratio of opacified lenses was not lower than that of clear lenses. The USAF-chart analysis showed a considerable reduction in brightness in opacified IOLs. The median (interquartile range) relative light transmission of opacified IOLs in comparison to clear lenses was 55.6% (20.8%) at the aperture size of 3 mm. In conclusion, the explanted opacified IOLs had comparable MTF values to those of clear lenses but significantly reduced light transmission.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Effect of types of hydrophilic acrylic monomers in reducing glistenings of hydrophobic acrylic intraocular lenses
    Kim, Tae Hyoung
    Song, Ki Chang
    OPTICAL MATERIALS, 2021, 119
  • [42] Light scattering and light transmittance in intraocular lenses explanted because of optic pacification
    Michelson, Jennifer
    Werner, Liliana
    Ollerton, Andrew
    Leishman, Lisa
    Bodnar, Zachary
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (08): : 1476 - 1485
  • [43] Opacification of AcriFlex 50CSE hydrophilic acrylic intraocular lenses
    Lim, Andrew Keat Eu
    Goh, Pik Pin
    Azura, Ramlee
    Mariam, Ismail
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (04): : 655 - 659
  • [44] Irradiation of hydrophilic acrylic intraocular lenses by a 365 nm UV lamp
    Torrisi, Alfio
    Roszkowska, Anna Maria
    Cutroneo, Mariapompea
    Torrisi, Lorenzo
    OPEN PHYSICS, 2024, 22 (01):
  • [45] Secondary opacification of hydrophilic acrylic intraocular lenses: Report of two cases
    Portmann, A.
    Miri, A.
    Genevois, C.
    Chevalier, L.
    Genty, D.
    Francois, A.
    Muraine, M.
    Genevois, O.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2013, 36 (07): : 561 - 566
  • [46] Light scattering and light transmittance in a series of calcified single-piece hydrophilic acrylic intraocular lenses of the same design
    Barra, Daniel
    Werner, Liliana
    Pacini Costa, Joao Luiz
    Morris, Caleb
    Ribeiro, Tiago
    Ventura, Bruna Vieira
    Dornelles, Fabio
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01): : 121 - 128
  • [47] Localized calcification of hydrophilic acrylic intraocular lenses in association with intracameral injection of gas
    Werner, Liliana
    Wilbanks, Garth
    Ollerton, Andrew
    Michelson, Jennifer
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04): : 720 - 721
  • [48] Opacification of two hydrophilic acrylic intraocular lenses 3 months after implantation
    Macky, TA
    Werner, L
    Soliman, MM
    Pandey, SK
    Izak, AM
    Apple, DJ
    OPHTHALMIC SURGERY LASERS & IMAGING, 2003, 34 (03): : 197 - 202
  • [49] Localized calcification of hydrophilic acrylic intraocular lenses after posterior segment procedures
    Balendiran, Vaishnavi
    MacLean, Kyle
    Mamalis, Nick
    Tetz, Manfred
    Werner, Liliana
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (12): : 1801 - 1807
  • [50] Glistenings 9 years after phacoemulsification in hydrophobic and hydrophilic acrylic intraocular lenses
    Chang, Anthony