Imaging Function and Relative Light Transmission of Explanted Opacified Hydrophilic Acrylic Intraocular Lenses

被引:4
|
作者
Naujokaitis, Tadas [1 ]
Khoramnia, Ramin [1 ]
Labuz, Grzegorz [1 ]
Choi, Chul Young [1 ,2 ]
Auffarth, Gerd U. [1 ]
Tandogan, Tamer [1 ,3 ]
机构
[1] Heidelberg Univ, Dept Ophthalmol, David J Apple Ctr Vis Res, D-69120 Heidelberg, Germany
[2] Sungkyunkwan Univ, Sch Med, Kangbuk Samsung Hosp, Dept Ophthalmol, Seoul 03181, South Korea
[3] Augenklin Pallas, CH-4600 Olten, Switzerland
关键词
intraocular lens; hydrophilic; opacification; calcification; optical quality; modulation transfer function; light transmission; OPTICAL-PERFORMANCE; OPACIFICATION; CALCIFICATION; STRAYLIGHT; EXCHANGE; TRANSMITTANCE; SCATTERING; OUTCOMES;
D O I
10.3390/diagnostics13101804
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the influence of intraocular lens (IOL) opacification on the optical performance of explanted hydrophilic acrylic IOLs. We performed a laboratory analysis of 32 Lentis LS-502-1 (Oculentis GmbH, Berlin, Germany) IOLs, explanted due to opacification, in comparison with six clear unused samples of the same IOL model. Using an optical bench setup, we obtained modulation transfer function (MTF), Strehl ratio, two-dimensional MTF, and United States Air Force (USAF) chart images. In addition, we assessed light transmission through the IOLs. The MTF values of opacified IOLs at 3-mm aperture were similar to those of clear lenses, with the median (interquartile range) values of 0.74 (0.01) vs. 0.76 (0.03) at the spatial frequency of 50 line pairs per millimeter in clear and opacified IOLs, respectively. The Strehl ratio of opacified lenses was not lower than that of clear lenses. The USAF-chart analysis showed a considerable reduction in brightness in opacified IOLs. The median (interquartile range) relative light transmission of opacified IOLs in comparison to clear lenses was 55.6% (20.8%) at the aperture size of 3 mm. In conclusion, the explanted opacified IOLs had comparable MTF values to those of clear lenses but significantly reduced light transmission.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Spectroscopic study of explanted opacified hydrophilic acrylic intraocular lenses
    Fodor, Mariann
    Petrovski, Goran
    Moe, Morten C.
    Biro, Zsolt
    Nemeth, Gabor
    Dinya, Zoltan
    Tornai, Ildiko
    Ujvari, Tamas
    Berta, Andras
    Facsko, Andrea
    ACTA OPHTHALMOLOGICA, 2011, 89 (02) : E161 - E166
  • [2] Exchange technique for opacified hydrophilic acrylic intraocular lenses
    Voros, GM
    Strong, NP
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2005, 15 (04) : 465 - 467
  • [3] Light-scattering characteristics of explanted opacified Aquasense intraocular lenses
    van der Meulen, I. J. E.
    Porooshani, H.
    van den Berg, T. J. T. P.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (06) : 830 - 832
  • [4] Late postoperative opacification of hydrophilic acrylic intraocular lenses- A Clinicopathologic Study of Six Opacified Intraocular Lenses
    Ni, Wei
    Ma, Zhizhong
    Bai, Tao
    Wang, Wei
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [5] Opacification of hydrophilic acrylic intraocular lenses: A clinical and ultrastructural analysis of three explanted lenses
    Katsimpris, JM
    Koliakou, K
    Petropoulos, IK
    Konstas, AG
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2003, 220 (03) : 165 - 169
  • [6] Hydrophilic acrylic intraocular lenses
    Mamalis, N
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (09): : 1339 - 1340
  • [7] Light Transmittance of Explanted Hydrophobic Acrylic Intraocular Lenses with Surface Light Scattering
    Werner, Liliana
    Morris, Caleb
    Stallings, Shannon
    Liu, Erica
    Floyd, Anne
    Ollerton, Andrew
    Leishman, Lisa
    Bodnar, Zachary
    Ong, Marcia
    Akinay, Ali
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [8] Laboratory evaluation of higher-order aberrations and light scattering in explanted opacified intraocular lenses
    Grzegorz Łabuz
    Timur M. Yildirim
    Gerd U. Auffarth
    Hyeck-Soo Son
    Ramin Khoramnia
    Eye and Vision, 8
  • [9] Laboratory evaluation of higher-order aberrations and light scattering in explanted opacified intraocular lenses
    Labuz, Grzegorz
    Yildirim, Timur M.
    Auffarth, Gerd U.
    Son, Hyeck-Soo
    Khoramnia, Ramin
    EYE AND VISION, 2021, 8 (01)
  • [10] Calcification of hydrophilic acrylic intraocular lenses
    Werner, Liliana
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 146 (03) : 341 - 343