Estimating size specific dose estimate from computed tomography radiograph localizer with radiation risk assessment

被引:1
|
作者
Burton, Christiane Sarah [1 ]
Al-Ward, Shahad [2 ]
机构
[1] St Jude Childrens Res Hosp, Dept Diagnost Imaging, Memphis, TN 38105 USA
[2] St Jude Childrens Res Hosp, Dept Radiat Oncol, Memphis, TN USA
来源
关键词
CT; localizer; SSDE; water-equivalent diameter; BODY CT; ABDOMEN;
D O I
10.1002/acm2.13989
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundQuantifying radiation burden is necessary for optimizing imaging protocols. The normalized dose coefficient (NDC) is determined from the water-equivalent diameter (WED) and is used to scale the CTDIvol based on body habitus to determine the size specific dose estimate (SSDE). In this study we determine the SSDE prior to the CT scan and how sensitive the SSDE from WED is to the lifetime attributable risk (LAR) from BEIR VII. MethodFor calibration, phantom images are used to relate the mean pixel values along a profile (PPV over bar $\overline {{\rm{PPV}}} $) of the CT localizer to the water-equivalent area (A(W)) of the CT axial scan at the same z-location. Images of the CTDIvol phantoms (32 cm, 16 cm, and similar to 1 cm) and ACR phantom (Gammex 464) were acquired on four scanners. The relationship between the A(W) and PPV over bar $\overline {{\rm{PPV}}} $ was used to calculate the WED from the CT localizer for patient scans. A total of 790 CT examinations of the chest and abdominopelvic regions were used in this study. The effective diameter (ED) was calculated from the CT localizer. The LAR was calculated based on the patient chest and abdomen using the National Cancer Institute Dosimetry System for Computed Tomography (NCICT). The radiation sensitivity index (RSI) and risk differentiability index (RDI) were calculated for SSDE and CTDIvol. ResultsThe WED from CT localizers and CT axials scans show good correlation (R-2 = 0.96) with the maximum percentage difference being 13.45%. The NDC from WED correlates poorly with LAR for lungs (R-2 = 0.18) and stomach (R-2 = 0.19), however that is the best correlation. ConclusionThe SSDE can be determined within 20% as recommended by the report of AAPM TG 220. The CTDIvol and SSDE are not good surrogates for radiation risk, however the sensitivity for SSDE improves when using WED instead of ED.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Understanding and communicating radiation dose and risk from cone beam computed tomography in dentistry
    Aanenson, Jill Weber
    Till, John E.
    Grogan, Helen A.
    JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (03): : 353 - 360
  • [42] DEVELOPMENT OF RADIATION DOSE CALCULATION SOFTWARE USING THE SIZE-SPECIFIC DOSE ESTIMATE
    Shohji, Tomokazu
    Tachibana, Atsushi
    Higuchi, Sousuke
    Nakata, Norio
    Hayashi, Daichi
    Katoh, Yo
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) : 359 - 369
  • [43] Comparative Study of Volume Computed Tomography Dose Index and Size-Specific Dose Estimate Head in Computed Tomography Examination for Adult Patients Based on the Mode of Automatic Tube Current Modulation
    Xu, Jian
    He, Xiaolong
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 71 - 76
  • [44] Volume computed tomography dose index (CTDIvol) and size-specific dose estimate (SSDE) for tube current modulation (TCM) in CT scanning
    Anam, C.
    Haryanto, F.
    Widita, R.
    Arif, I
    Dougherty, G.
    McLean, D.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2018, 16 (03): : 289 - 297
  • [45] ASSESSMENT OF RADIATION DOSE OF MOBILE COMPUTED TOMOGRAPHY IN INTENSIVE CARE UNITS
    Kim, Eunhye
    Choi, Yoonseok
    Park, Hyemin
    Na, Chanyoung
    Kim, Jungmin
    Kim, Jungsu
    Han, Taeho
    RADIATION PROTECTION DOSIMETRY, 2021, 196 (1-2) : 60 - 70
  • [46] ASSESSMENT OF PATIENT DOSE FROM COMPUTED-TOMOGRAPHY
    SHRIMPTON, PC
    WALL, BF
    RADIATION PROTECTION DOSIMETRY, 1992, 43 (1-4) : 205 - 208
  • [47] Application of Aortic Computed Tomography Angiography with Low Radiation Dose: Assessment of Image Quality and Radiation Dose
    Liu, Jianhua
    Li, Long
    Zhang, Yan
    Yuan, Qinghai
    Sun, Yong
    2012 IEEE FIFTH INTERNATIONAL CONFERENCE ON ADVANCED COMPUTATIONAL INTELLIGENCE (ICACI), 2012, : 148 - 151
  • [48] Computed Tomography Dose Monitoring: Is Radiation Dose the Wrong Patient Imaging Risk to Manage?
    Toth, Thomas
    JOURNAL OF PATIENT SAFETY, 2017, 13 (02) : 109 - 109
  • [49] Determination of radiation dose during computed tomography examinations in southern Saudi Arabian hospitals using size-specific dose estimates
    Saeed, Mohammed K.
    Asiri, Abdullah Ali M.
    MAEJO INTERNATIONAL JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 15 (02) : 147 - 158
  • [50] SIMPLE METHOD OF SIZE-SPECIFIC DOSE ESTIMATES CALCULATION FROM PATIENT WEIGHT ON COMPUTED TOMOGRAPHY
    Iriuchijima, Akiko
    Fukushima, Yasuhiro
    Nakajima, Takahito
    Tsushima, Yoshito
    Ogura, Akio
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (02) : 208 - 212