The effect of voxelization in Monte Carlo simulation to validate Bragg peak characteristics for a pencil proton beam

被引:1
|
作者
Price, Gumec M. Gungor [1 ,3 ]
Sarigul, Neslihan [2 ]
机构
[1] Cukurova Univ, Arts Sci Fac, Phys Dept, Adana, Turkiye
[2] Hacettepe Univ, Inst Nucl Sci, Ankara, Turkiye
[3] Cukurova Univ, Arts Sci Fac, Phys Dept, Turkiye, 01330 Saricam, Adana, Turkiye
关键词
voxel size effect; proton therapy; Bragg peak; Monte Carlo simulations; Geant4; DEPTH-DOSE CURVES; THERAPY; GEANT4; ALGORITHM; ENERGY; MODEL; WATER; HETEROGENEITIES; VERIFICATION; RADIOTHERAPY;
D O I
10.5603/RPOR.a2023.0007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The purpose of this research was to show how the Bragg peak (BP) characteristics were affected by changing the voxel size in longitudinal and transverse directions in Monte Carlo (MC) simulations by using Geant4 and to calculate BP characteristics accurately by considering the voxel size effect for 68 MeV and 235.81 MeV.Materials and methods: Different interpolation techniques were applied to simulation data to find the closest results to the experimental data.Results: When the x-size of the voxel was increased 2 times at low energy, the maximum dose increase in the entrance and plateau regions were 17.8% and 17%, respectively, while BP curve shifted to the shallower region, resulting in a 0.5 mm reduction in the curable tumor width (W80pd). At high energy, the maximum dose increase at the entrance and plateau regions were 0.4% and 0.6%, respectively, while it was observed that W80pd did not change. When the y-z sizes of the voxel were in-creased 2 times at low energy, the maximum dose reduction at the entrance and plateau regions was 3.4%, but no change was observed in W80pd. At high energy, when the y-z sizes of the voxel were increased 2.2 times, the maximum dose reduction at the entrance and plateau regions were 8.9% and 9.1%, respectively, while W80pd increased by 0.5 mm. When linear, cubic spline, and Akima interpolations were applied to the simulation data, it was found that the results closest to the experimental data were obtained for Akima interpolations for both energies.Conclusion: it has been shown that the voxel size effect for the longitudinal direction was more effective at low energy than at high energy. However, the voxel size effect for the transverse direction was more effective for high energy.
引用
收藏
页码:102 / 113
页数:12
相关论文
共 50 条
  • [1] Monte Carlo simulation and analysis of proton energy-deposition patterns in the Bragg peak
    Gonzalez-Munoz, Gloria
    Tilly, Nina
    Fernandez-Varea, Jose M.
    Ahnesjo, Anders
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11): : 2857 - 2875
  • [2] The Application of Cloud Computing in Pencil Beam Scanning Proton Therapy Monte Carlo Simulation
    Wang, Z.
    Gao, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 296 - 296
  • [3] Monte Carlo model of a gantry for proton pencil beam scanning
    Winterhalter, Carla
    Safai, Sairos
    Grassberger, Clemens
    Oxley, David
    Weber, Damien C.
    Lomax, Tony
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 837 - 837
  • [4] Validation and Implementation of Proton Pencil Beam Scanning Nozzle Using TOPAS Monte Carlo Simulation
    Lin, C.
    Feng, Y.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E875 - E875
  • [5] Single pencil beam benchmark of a module for Monte Carlo simulation of proton transport in the PENELOPE code
    Verbeek, Nico
    Wulff, Joerg
    Baeumer, Christian
    Smyczek, Sabrina
    Timmermann, Beate
    Brualla, Lorenzo
    [J]. MEDICAL PHYSICS, 2021, 48 (01) : 456 - 476
  • [6] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Monte Carlo simulation tool for commissioning measurements corrections of a proton pencil beam scanning (PBS) mode
    Fracchiolla, F.
    Lorentini, S.
    Schwarz, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S13 - S14
  • [8] Development and validation of the Dynamic Collimation Monte Carlo simulation package for pencil beam scanning proton therapy
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Hyer, Daniel E.
    Geoghegan, Theodore J.
    Patwardhan, Kaustubh A.
    Smith, Blake R.
    Flynn, Ryan T.
    Yu, Jen
    Rana, Suresh
    Gutierrez, Alonso N.
    Hill, Patrick M.
    [J]. MEDICAL PHYSICS, 2021, 48 (06) : 3172 - 3185
  • [9] Pitfalls in the beam modelling process of Monte Carlo calculations for proton pencil beam scanning
    Winterhalter, Carla
    Aitkenhead, Adam
    Oxley, David
    Richardson, Jenny
    Weber, Damien C.
    MacKay, Ranald, I
    Lomax, Antony J.
    Safai, Sairos
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1107):
  • [10] Monte Carlo simulation of the dynamic collimator in a pencil beam scanning system
    Almeida, I. P.
    Sure, M.
    Vilches-Freixas, G.
    Cooley, J.
    Zwart, T.
    Langenegger, A.
    Verhaegen, F.