Immersion ultrasound improves the repeatability of shear-wave elastography for measuring median nerve elasticity

被引:2
|
作者
Zhang, Zhijun [1 ]
Wang, Hui [1 ]
Ding, Xiaoya [1 ]
He, Shu [2 ]
机构
[1] Chongqing Med Univ, Univ Town Hosp, Dept Ultrasound, Chongqing 401331, Peoples R China
[2] Fifth Peoples Hosp Chongqing, Med Affairs Dept, Chongqing 400030, Peoples R China
关键词
Shear wave elastography; Median nerve; Repeatability; POTENTIAL SONOGRAPHIC METHOD; DIAGNOSIS;
D O I
10.1186/s13018-023-04097-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
ObjectiveTo study the factors that influence the measurement of median nerve elasticity, to provide a more reproducible test for the assessment of median nerve elasticity using shear wave elasticity (SWE), and to reduce operator empirical dependence. To compare the repeatability of the median nerve elasticity measurement using immersion SWE with that using contact SWE, analyze the factors affecting SWE measurement, and provide a more repeatable method for doctors without SWE operation experience.MethodsTwo doctors without SWE operation experience measured the median nerve mean elastic modulus (Emean) at the same position and at different time points on the right wrist of 58 healthy volunteers using immersion and contact ultrasound methods. The intraobserver and interobserver repeatability of measurements was assessed using the interclass correlation coefficient (ICC), while the repeatability was assessed using the Bland-Altman diagram.ResultsThe intraobserver and interobserver repeatability of the median nerve elasticity measured via contact SWE by inexperienced operators were classified as good, with ICCs of 0.633 (95% CI 0.380-0.783) and 0.552 (95% CI 0.243-0.735), respectively. The intraobserver and interobserver repeatability of the median nerve elasticity measured by immersion SWE were very good, with ICCs of 0.975 (95% CI 0.958-0.985) and 0.942 (95% CI 0.902-0.966), respectively. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by immersion SWE showed that 98% of the points fell within the 95% limits of agreement. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by contact SWE showed that 94% of the points fell within the 95% limits of agreement.ConclusionImmersion ultrasound can improve the repeatability of median nerve elasticity measurements by inexperienced operators.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Agreement and accuracy of shear-wave techniques (point shear-wave elastography and 2D-shear-wave elastography) using transient elastography as reference
    Fernandes, Flavia
    Piedade, Juliana
    Freitas, Gabriela
    Area, Philippe
    Santos, Ricardo
    Grinsztejn, Beatriz
    Veloso, Valdilea
    Pereira, Gustavo
    Perazzo, Hugo
    EUROPEAN JOURNAL OF GASTROENTEROLOGY & HEPATOLOGY, 2022, 34 (08) : 873 - 881
  • [32] Muscle elasticity in patients with neonatal brachial plexus palsy using shear-wave ultrasound elastography. Preliminary results
    Riquier-Le Chatelier, Morgane
    Giai, Joris
    Lallement-Dudek, Pauline
    Herisson, Olivier
    Fitoussi, Frank
    JOURNAL OF PEDIATRIC ORTHOPAEDICS-PART B, 2021, 30 (04): : 385 - 392
  • [33] Assessment of thyroid gland elasticity with shear-wave elastography in Hashimoto's thyroiditis patients
    Kara, Turgay
    Ates, Fatih
    Durmaz, Mehmet Sedat
    Akyurek, Nesibe
    Durmaz, Funda Gokgoz
    Ozbakir, Bora
    Ozturk, Mehmet
    JOURNAL OF ULTRASOUND, 2020, 23 (04) : 543 - 551
  • [34] Determination of Normal Hepatic Elasticity by Using Real-time Shear-wave Elastography
    Suh, Chong Hyun
    Kim, So Yeon
    Kim, Kyoung Won
    Lim, Young-Suk
    Lee, So Jung
    Lee, Moon-Gyu
    Lee, JungBok
    Lee, Sung-Gyu
    Yu, Eunsil
    RADIOLOGY, 2014, 271 (03) : 895 - 900
  • [35] Evaluation of ultrasound point shear wave elastography reliability in an elasticity phantom
    Sultan, Salahaden R.
    Alghamdi, Amer
    Abdeen, Rawan
    Almutairi, Fahad
    ULTRASONOGRAPHY, 2022, 41 (02) : 291 - 297
  • [36] Effect of Depth on Ultrasound Point Shear Wave Elastography in an Elasticity Phantom
    Almutairi, Fahad F.
    Abdeen, Rawan
    Alyami, Jaber
    Sultan, Salahaden R.
    APPLIED SCIENCES-BASEL, 2022, 12 (13):
  • [37] Shear Wave Elastography Improves Specificity of Ultrasound for Parotid Nodules
    Liu, Guangqing
    Wu, Size
    Liang, Xian
    Cui, Xiaojing
    Zuo, Dongshen
    ULTRASOUND QUARTERLY, 2018, 34 (02) : 62 - 66
  • [38] Quantitative Assessment of Thyroid, Submandibular, and Parotid Glands Elasticity With Shear-Wave Elastography in Children
    Habibi, Hatice Arioz
    Durmaz, Emine Sebnem Memis
    Qarayeva, Vusale
    Kandemirli, Sedat Giray
    Ucar, Ayse Kalyoncu
    Aslan, Mine
    Apaydin, Gozde
    Kurugoglu, Sebuh
    Adaletli, Ibrahim
    ULTRASOUND QUARTERLY, 2018, 34 (02) : 58 - 61
  • [39] Correlations Between Range of Motion and Elasticity of the Coracohumeral Ligament Evaluated With Shear-Wave Elastography
    Kanazawa, Kenji
    Hagiwara, Yoshihiro
    Sekiguchi, Takuya
    Fujita, Ryo
    Suzuki, Kazuaki
    Koide, Masashi
    Ando, Akira
    Yabe, Yutaka
    JOURNAL OF SPORT REHABILITATION, 2021, 30 (01) : 9 - 15
  • [40] Using shear-wave elastography in skeletal muscle: A repeatability and reproducibility study on biceps femoris muscle
    Sarabon, Nejc
    Kozinc, Zige
    Podrekar, Nastja
    PLOS ONE, 2019, 14 (08):