Immersion ultrasound improves the repeatability of shear-wave elastography for measuring median nerve elasticity

被引:2
|
作者
Zhang, Zhijun [1 ]
Wang, Hui [1 ]
Ding, Xiaoya [1 ]
He, Shu [2 ]
机构
[1] Chongqing Med Univ, Univ Town Hosp, Dept Ultrasound, Chongqing 401331, Peoples R China
[2] Fifth Peoples Hosp Chongqing, Med Affairs Dept, Chongqing 400030, Peoples R China
关键词
Shear wave elastography; Median nerve; Repeatability; POTENTIAL SONOGRAPHIC METHOD; DIAGNOSIS;
D O I
10.1186/s13018-023-04097-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
ObjectiveTo study the factors that influence the measurement of median nerve elasticity, to provide a more reproducible test for the assessment of median nerve elasticity using shear wave elasticity (SWE), and to reduce operator empirical dependence. To compare the repeatability of the median nerve elasticity measurement using immersion SWE with that using contact SWE, analyze the factors affecting SWE measurement, and provide a more repeatable method for doctors without SWE operation experience.MethodsTwo doctors without SWE operation experience measured the median nerve mean elastic modulus (Emean) at the same position and at different time points on the right wrist of 58 healthy volunteers using immersion and contact ultrasound methods. The intraobserver and interobserver repeatability of measurements was assessed using the interclass correlation coefficient (ICC), while the repeatability was assessed using the Bland-Altman diagram.ResultsThe intraobserver and interobserver repeatability of the median nerve elasticity measured via contact SWE by inexperienced operators were classified as good, with ICCs of 0.633 (95% CI 0.380-0.783) and 0.552 (95% CI 0.243-0.735), respectively. The intraobserver and interobserver repeatability of the median nerve elasticity measured by immersion SWE were very good, with ICCs of 0.975 (95% CI 0.958-0.985) and 0.942 (95% CI 0.902-0.966), respectively. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by immersion SWE showed that 98% of the points fell within the 95% limits of agreement. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by contact SWE showed that 94% of the points fell within the 95% limits of agreement.ConclusionImmersion ultrasound can improve the repeatability of median nerve elasticity measurements by inexperienced operators.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Immersion ultrasound improves the repeatability of shear-wave elastography for measuring median nerve elasticity
    Zhijun Zhang
    Hui Wang
    Xiaoya Ding
    Shu He
    Journal of Orthopaedic Surgery and Research, 18
  • [2] Measuring shear-wave speed with point shear-wave elastography and MR elastography: a phantom study
    Kishimoto, Riwa
    Suga, Mikio
    Koyama, Atsuhisa
    Omatsu, Tokuhiko
    Tachibana, Yasuhiko
    Ebner, Daniel K.
    Obata, Takayuki
    BMJ OPEN, 2017, 7 (01):
  • [3] Evaluation of the healthy median nerve elasticity Feasibility and reliability of shear wave elastography
    Zhu, Bihui
    Yan, Feng
    He, Ying
    Wang, Liyun
    Xiang, Xi
    Tang, Yuanjiao
    Yang, Yujia
    Qiu, Li
    MEDICINE, 2018, 97 (43)
  • [4] REGULARIZED TRACKING OF SHEAR-WAVE IN ULTRASOUND ELASTOGRAPHY
    Horeh, Mahmoud Derakhshan
    Asif, Amir
    Rivaz, Hassan
    2017 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2017, : 6264 - 6268
  • [5] Ultrasound elastography: compression elastography and shear-wave elastography in the assessment of tendon injury
    Rui Prado-Costa
    João Rebelo
    João Monteiro-Barroso
    Ana Sofia Preto
    Insights into Imaging, 2018, 9 : 791 - 814
  • [6] Shear-Wave Sources for Soft Tissues in Ultrasound Elastography
    Giammarinaro, B.
    Zorgani, A.
    Catheline, S.
    IRBM, 2018, 39 (04) : 236 - 242
  • [7] Ultrasound elastography: compression elastography and shear-wave elastography in the assessment of tendon injury
    Prado-Costa, Rui
    Rebelo, Joao
    Monteiro-Barroso, Joao
    Preto, Ana Sofia
    INSIGHTS INTO IMAGING, 2018, 9 (05): : 791 - 814
  • [8] Quantitative Assessment of Normal Soft-Tissue Elasticity Using Shear-Wave Ultrasound Elastography
    Arda, Kemal
    Ciledag, Nazan
    Aktas, Elif
    Aribas, Bilgin Kadri
    Kose, Kenan
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) : 532 - 536
  • [9] A Comparative Study of Strain and Shear-Wave Elastography in an Elasticity Phantom
    Carlsen, Jonathan F.
    Pedersen, Malene R.
    Ewertsen, Caroline
    Saftoiu, Adrian
    Lonn, Lars
    Rafaelsen, Soren R.
    Nielsen, Michael B.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (03) : W236 - W242
  • [10] Reliability of ultrasound shear-wave elastography in assessing low back musculature elasticity in asymptomatic individuals
    Koppenhaver, Shane
    Kniss, Joshua
    Lilley, Daniel
    Oates, Michael
    Fernandez-de-las-Penas, Cesar
    Maher, Ruth
    Croy, Theodore
    Shinohara, Minoru
    JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2018, 39 : 49 - 57