Treatment for Stanford type B aortic dissection with insufficient anchoring region using castor integrated branched aortic stent graft

被引:1
|
作者
Chen, Weiqing [1 ]
Liu, Dabing [2 ]
Chen, Tao [1 ]
Liu, Jian [1 ]
Guo, Yi [1 ]
Ye, Bo [1 ]
机构
[1] Ganzhou Peoples Hosp, Dept Vasc Surg, Ganzhou, Jiangxi, Peoples R China
[2] Peoples Hosp Ganxian Dist, Dept Gen Surg, Ganzhou, Jiangxi, Peoples R China
来源
关键词
integrated branched stent; aortic dissection; thoracic endovascular aortic repair (TEVAR); anchoring region; castor; LEFT SUBCLAVIAN ARTERY; ENDOVASCULAR REPAIR; CHIMNEY TECHNIQUE; REVASCULARIZATION; OUTCOMES; MANAGEMENT; SOCIETY; STROKE; ZONE;
D O I
10.3389/fcvm.2024.1351342
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: To investigate the clinical efficacy of Castor integrated branched aortic stent graft for the treatment of Stanford type B aortic dissection with insufficient anchoring area. Methods: Retrospective analysis of clinical data of 26 patients with Stanford type B aortic dissection with insufficient anchoring region (<15 mm) treated by Castor branched aortic stent graft from September 2018 to June 2022 at Ganzhou People's Hospital, including 23 acute cases and 3 chronic cases. Results: Surgical procedures were successfully performed in all 26 patients, and during the perioperative period no complications occurred, such as cerebrovascular accident, stenosis or occlusion of left subclavian artery, progression of reverse avulsion of aortic dissection, and paraplegia. During the operation 2 patients had a small amount of type I endoleak, which disappeared during the postoperative follow-up. The other patients had good postoperative follow-up results. Review of the aortic CTA indicated good stent morphology with patency of the left subclavian artery. Conclusions: The Castor integrated branched aortic stent graft expanded the indications for endoluminal treatment for Stanford type B aortic dissection, which can avoid open surgery and has good clinical outcomes.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Treatment of chronic type B aortic dissection with endovascular stent-graft placement
    Noriyuki Kato
    Tadanori Hirano
    Takatsugu Shimono
    Yoshiyuki Nomura
    Miwako Goto
    Hajime Sakuma
    Isao Yada
    Kan Takeda
    CardioVascular and Interventional Radiology, 2000, 23 : 60 - 62
  • [32] Stent-graft treatment of type B aortic dissection involving the right aortic arch - Case report
    Yamazaki, I
    Imoto, K
    Ichikawa, Y
    Kondo, J
    Takanashi, Y
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 2000, 64 (09): : 727 - 728
  • [33] Treatment of chronic type B aortic dissection with endovascular stent-graft placement
    Kato, N
    Hirano, T
    Shimono, T
    Nomura, Y
    Goto, M
    Sakuma, H
    Yada, I
    Takeda, K
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 23 (01) : 60 - 62
  • [34] Endovascular Stent-Graft Placement in Patients with Stanford Type B Aortic Dissection in China: A Systematic Review
    Wang, Junwei
    Li, Yonghui
    Li, Yongxin
    Ren, Zefang
    Chen, Peng
    Qian, Xueke
    Wang, Shenming
    Wang, Jinsong
    ANNALS OF VASCULAR SURGERY, 2016, 36 : 298 - 309
  • [35] Penetrating Aortic Ulceration Treated with Castor Branched Aortic Stent Graft-A Case Series
    Rizza, Antonio
    Negro, Francesco
    Palermi, Stefano
    Palmieri, Cataldo
    Murzi, Michele
    Credi, Giovanni
    Berti, Sergio
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2022, 19 (08)
  • [36] Stent graft-induced new entry after endovascular repair for Stanford type B aortic dissection
    Dong, Zhihui
    Fu, Weiguo
    Wang, Yuqi
    Wang, Chunsheng
    Yan, Zhiping
    Guo, Daqiao
    Xu, Xin
    Chen, Bin
    JOURNAL OF VASCULAR SURGERY, 2010, 52 (06) : 1450 - 1457
  • [37] Emergent endovascular stent-graft placement to treat ruptured Stanford type B acute aortic dissection
    Kawamura, SK
    Nishimaki, H
    Lin, ZB
    Machii, M
    Isobe, Y
    Hayakawa, K
    Soma, K
    JOURNAL OF VASCULAR SURGERY, 2004, 39 (03) : 668 - 671
  • [38] Surgical treatment for retrograde type A aortic dissection after endovascular stent graft placement for type B dissection
    Lu, Shuyang
    Lai, Hao
    Wang, Chunsheng
    Sun, Xiaoning
    Hong, Tao
    Song, Kai
    Yuan, Zhenkai
    Liu, Xueling
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2012, 14 (05) : 538 - 542
  • [39] Correction to: Aortic remodeling with frozen elephant trunk technique for Stanford type A aortic dissection using Japanese J-graft open stent graft
    Masato Tochii
    Yoshiyuki Takami
    Hiroshi Ishikawa
    Michiko Ishida
    Yoshiro Higuchi
    Yusuke Sakurai
    Kentaro Amano
    Yasushi Takagi
    Heart and Vessels, 2019, 34 : 316 - 317
  • [40] Perioperative aortic dissection rupture after endovascular stent graft placement for treatment of type B dissection
    Huang Wen-hui
    Luo Song-yuan
    Luo Jian-fang
    Liu Yuan
    Fan Rui-xin
    Xue Ling
    Yang Fang
    Kang Hui-yuan
    Gu Meng-nan
    Liu Zhen
    Xie Nian-jin
    Dong Hao-jian
    Ni Zhong-han
    Huang Mei-ping
    Chen Ji-yan
    CHINESE MEDICAL JOURNAL, 2013, 126 (09) : 1636 - 1641