Comparison of Thames Medical CAT plus Doppler and SunTech Vet 20 oscillometric devices for non-invasive blood pressure measurement in conscious cats

被引:0
|
作者
Casas, Clara [1 ,2 ]
Dye, Charlotte [1 ]
机构
[1] IVC Evidensia, Pride Vet Ctr, Dept Small Anim Med, Derby DE24 8HX, England
[2] Pride Vet Referrals, Derby DE24 8HX, England
关键词
Blood pressure; hypertension; SunTech Vet20; CAT plus Doppler; oscillometric; SYSTEMIC HYPERTENSION; HEALTHY CATS; MONITOR; DOGS; AGREEMENT;
D O I
10.1177/1098612X231216350
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objectives A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats. Methods Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data. Results A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats. Conclusions and relevance Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] A comparison of CAT Doppler and oscillometric Memoprint machines for non-invasive blood pressure measurement in conscious cats
    Jepson, RE
    Hartley, V
    Mendl, M
    Caney, SM
    Gould, DJ
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2005, 7 (03) : 147 - 152
  • [2] Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats
    Cerna, Petra
    Archontakis, Panos E.
    Cheuk, Hester O. K.
    Gunn-Moore, Danielle A.
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2021, 23 (02) : 121 - 130
  • [3] Non-invasive blood pressure measurement in conscious rabbits: A comparison of Doppler ultrasonic and oscillometric devices
    Gonzalez-Gonzalez, Enrique
    Gonzalez-Alonso-Alegre, Elisa
    Montesinos-Barcelo, Andres
    Caro-Vadillo, Alicia
    VETERINARY RECORD, 2024, 195 (07)
  • [4] Validation of the oscillometric blood pressure monitor Vet20 SunTech in anesthetized healthy cats
    Cremer, Jeannette
    da Cunha, Anderson
    Aulakh, Karanvir
    Liu, Chin-Chi
    Acierno, Mark J.
    VETERINARY ANAESTHESIA AND ANALGESIA, 2020, 47 (03) : 309 - 314
  • [5] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Jiang, Ziqing
    Li, Shaoying
    Wang, Lin
    Yu, Feng
    Zeng, Yanping
    Li, Hongbo
    Li, Jun
    Zhang, Zhanfeng
    Zuo, Junling
    JOURNAL OF ANESTHESIA, 2024, 38 (02) : 222 - 231
  • [6] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Ziqing Jiang
    Shaoying Li
    Lin Wang
    Feng Yu
    Yanping Zeng
    Hongbo Li
    Jun Li
    Zhanfeng Zhang
    Junling Zuo
    Journal of Anesthesia, 2024, 38 : 222 - 231
  • [7] Comparison of Doppler ultrasonic sphygmomanometry, oscillometry and high-definition oscillometry for non-invasive blood pressure measurement in conscious cats
    Knies, Marieke
    Teske, Erik
    Kooistra, Hans
    JOURNAL OF FELINE MEDICINE AND SURGERY, 2024, 26 (03)
  • [8] Calibration of oscillometric non-invasive devices for monitoring blood pressure
    Doh, Il
    Lim, Hyun Kyoon
    Ahn, Bongyoung
    METROLOGIA, 2015, 52 (02) : 291 - 296
  • [9] Comparison of oscillometric, Doppler and invasive blood pressure measurement in anesthetized goats
    Szalus-Jordanow, Olga
    Czopowicz, Michel
    Moroz, Agata
    Mickiewicz, Marcin
    Garncarz, Magdalena
    Bagnicka, Emilia
    Frymus, Tadeusz
    Kaba, Jaroslaw
    PLOS ONE, 2018, 13 (05):
  • [10] A clinical evaluation of blood pressure through non-invasive measurement using the oscillometric procedure in conscious dogs
    Mishina, M
    Watanabe, T
    Fujii, K
    Maeda, H
    Wakao, Y
    Takahashi, M
    JOURNAL OF VETERINARY MEDICAL SCIENCE, 1997, 59 (11): : 989 - 993