Comparison of Thames Medical CAT plus Doppler and SunTech Vet 20 oscillometric devices for non-invasive blood pressure measurement in conscious cats

被引:0
|
作者
Casas, Clara [1 ,2 ]
Dye, Charlotte [1 ]
机构
[1] IVC Evidensia, Pride Vet Ctr, Dept Small Anim Med, Derby DE24 8HX, England
[2] Pride Vet Referrals, Derby DE24 8HX, England
关键词
Blood pressure; hypertension; SunTech Vet20; CAT plus Doppler; oscillometric; SYSTEMIC HYPERTENSION; HEALTHY CATS; MONITOR; DOGS; AGREEMENT;
D O I
10.1177/1098612X231216350
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objectives A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats. Methods Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data. Results A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats. Conclusions and relevance Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Beyond the Doppler: The Role of Non-Invasive Blood Pressure Measurement in Patients With Continuous Flow Left Ventricular Assist Devices
    Toma, Mustafa
    Kroll, Nicholas
    Gschnell, Heather
    Hostutler, Sheryl
    Raspovic-Hopper, Susie
    Vacha, Catherine
    Mountis, Maria M.
    JOURNAL OF CARDIAC FAILURE, 2010, 16 (08) : S103 - S103
  • [32] Review of non-invasive blood pressure measurement in animals: Part 2-Evaluation of the performance of non-invasive devices
    Skelding, Alicia
    Valverde, Alexander
    CANADIAN VETERINARY JOURNAL-REVUE VETERINAIRE CANADIENNE, 2020, 61 (05): : 481 - 498
  • [33] A Comparison of Non-Invasive Blood Pressure Measurement Strategies with Intra-Arterial Measurement
    Rebesco, Matthew R.
    Pinkston, M. Cornelia
    Smyrnios, Nicholas A.
    Weisberg, Stacy N.
    PREHOSPITAL AND DISASTER MEDICINE, 2020, 35 (05) : 516 - 523
  • [34] Comparison of non-invasive and invasive blood pressure measurements in horses during anaesthesia using the oscillometric blood pressure monitor S/5 Datex Ohmeda
    Fouche, Nathalie
    Auer, Ulrike
    Iff, Isabelle
    PFERDEHEILKUNDE, 2016, 32 (05): : 479 - 484
  • [35] Continuous non-invasive measurement of mean blood pressure in fingers by volume-clamp and differential oscillometric method
    Jagomagi, K
    Talts, J
    Raamat, R
    Lansimies, E
    CLINICAL PHYSIOLOGY, 1996, 16 (05): : 551 - 560
  • [36] From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: does easiness come with reliability?
    Benmira, A.
    Perez-Martin, A.
    Schuster, I.
    Aichoun, I.
    Coudray, S.
    Bereksi-Reguig, F.
    Dauzat, M.
    EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (02) : 179 - 189
  • [37] Comparison of blood pressure variability by means of non-invasive oscillometric and intra-arterial blood pressure monitoring in acute stroke patients
    Manios, E
    Tsivgoulis, G
    Spengos, K
    Spiliopoulou, J
    Dolianitis, K
    Tseke, P
    Xynos, K
    Dimitriou, A
    Vemmos, K
    Zakopoulos, N
    Mavrikakis, M
    JOURNAL OF HYPERTENSION, 2005, 23 : S125 - S125
  • [38] Comparison between the Parks Medical Doppler and the Mano M?dical Vet BP Doppler for recording indirect systemic blood pressure in conscious dogs
    Domingues, Marina
    Archontakis, Panagiotis E.
    Bouvard, Jonathan
    Winter, Luke
    Gunn-Moore, Danielle A.
    OPEN VETERINARY JOURNAL, 2022, 12 (04) : 562 - 566
  • [39] Influence of intermittent pneumatic compression devices on non-invasive blood pressure measurement of the ankle
    Parikh B.R.
    Simon A.M.
    Kouvaras J.N.
    Ciolino R.B.
    Suthar T.P.
    Dorian R.S.
    Journal of Clinical Monitoring and Computing, 2007, 21 (6) : 381 - 386
  • [40] Comparison of Non-invasive and Invasive Arterial Blood Pressure Measurement for Assessment of Dynamic Cerebral Autoregulation
    Nils H. Petersen
    Santiago Ortega-Gutierrez
    Andres Reccius
    Arjun Masurkar
    Amy Huang
    Randolph S. Marshall
    Neurocritical Care, 2014, 20 : 60 - 68