Performance Evaluation of Urine Osmolality Measurement on Sysmex UF-5000 and the Effect of Molecules and Particles in Urine

被引:0
|
作者
Yis, Ozgur M. [1 ]
Alisik, Murat [1 ]
Bugdayci, Guler [1 ]
Sert, Mehmet A. [1 ]
Erdogan, Utku G. [1 ]
Ates, Merve [1 ]
机构
[1] Bolu Abant Izzet Baysal Univ, Fac Med, Dept Med Biochem, Bolu, Turkiye
关键词
urine osmolality; conductivity; urine; freezing point depression; FREEZING-POINT DEPRESSION; CONDUCTIVITY;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Urine osmolality determines the concentration ability of the kidney. Therefore, it is used to assess the body's hydration status, electrolyte levels, and acid-base disturbances. We aimed to evaluate the analytical perfor-mance of osmolality measurement of the Sysmex UF-5000 (UF-5000), to examine the effect of different molecules and particles in the urine on the osmolality measurement. Methods: Complete urinalysis and conductivity-based osmolality analysis with UF-5000 and osmolality analysis with Advanced (R) Model 3320 Micro-Osmometer (AI-3320) by freezing point reduction method were performed in the urine samples. Samples were grouped as negative, glucosuria, proteinuria, hematuria, pyuria, crystalluria, and urobilinogen.Results: Total impressions were calculated as < 5% and accuracy values were < 1.66% in both analyzers. The re-gression equation was found to be y =-12.54 + 0.956x and the relative difference between the analyzers was 8.7% in 586 samples. When patients with Glucose > 2+ were excluded the regression equation of 507 samples was found as y = 5.10 + 0.948x and the relative difference was 4.6%. The percentages of samples with a difference greater than the allowable difference were 18.8%, 11.6%, 35.9%, 13.7%, 18.7%, and < 12.2% in all samples, all samples without glucosuria > 2+, glucosuria, glucosuria < 3+, proteinuria, and other subgroups, respectively. Conclusions: Considering the good accessibility of the automated routine complete urine analyzer, UF-5000 can be considered to determine whether urine osmolality is within reference or should be measured by methods based on colligative properties. Thus, referral of patients to a clinic that uses the colligative measurement method may be used more effectively.(Clin. Lab. 2023;69:543-552. DOI: 10.7754/Clin.Lab.2022.220536)
引用
收藏
页码:543 / 552
页数:10
相关论文
共 50 条
  • [41] Evaluation of the Sysmex UF-1000i® urine flow cytometer in the diagnostic work-up of suspected urinary tract infection in a Dutch general hospital
    van der Zwet, Wil C.
    Hessels, Jan
    Canbolat, Fatih
    Deckers, Martine M. L.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (12) : 1765 - 1771
  • [42] Performance Evaluation of VeriCheck Drug Test Dip Device for Measurement of Drugs of Abuse in the Urine
    Butorin, Yury
    Hinton, Lorianne
    Thomas, Dylan
    Colbourne, Penny
    Kinniburgh, David
    Bennett, Melissa
    Venner, Allison
    CLINICAL BIOCHEMISTRY, 2024, 133
  • [43] Evaluation of the new Sysmex OF-5000 fluorescence flow cytometry analyser for ruling out bacterial urinary tract infection and for prediction of Gram negative bacteria in urine cultures
    De Rosa, Rita
    Grosso, Shamanta
    Lorenzi, Giada
    Bruschetta, Graziano
    Camporese, Alessandro
    CLINICA CHIMICA ACTA, 2018, 484 : 171 - 178
  • [44] High performance liquid chromatographic measurement of iothalamate in human serum and urine for evaluation of glomerular filtration rate
    Bi, Daoqin
    Leary, Kevin J.
    Weitz, Julie A.
    Cherstniakova, Svetlana A.
    Reil, Michael A.
    Roy, Michael J.
    Cantilena, Louis R.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 856 (1-2): : 95 - 99
  • [45] Performance evaluation of the new fully automated urine particle analyser OF-5000 compared to the reference method of the Fuchs-Rosenthal chamber
    Previtali, Giulia
    Ravasio, Rudi
    Seghezzi, Michela
    Buoro, Sabrina
    Alessio, Maria Grazia
    CLINICA CHIMICA ACTA, 2017, 472 : 123 - 130
  • [46] Performance Evaluation of Urine Myoinositol Measurement Using Myoinositol Dehydrogenase - Applicability of the TBA200FR Device
    Yoshimoto, S.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A48 - A48
  • [47] Evaluation of artificial neural network algorithms for predicting the effect of the urine flow rate on the power performance of microbial fuel cells
    de Ramon-Fernandez, A.
    Salar-Garcia, M. J.
    Fernandez, D. Ruiz
    Greenman, J.
    Ieropoulos, I. A.
    ENERGY, 2020, 213
  • [48] Evaluation of artificial neural network algorithms for predicting the effect of the urine flow rate on the power performance of microbial fuel cells
    de Ramón-Fernández, A.
    Salar-García, M.J.
    Ruiz Fernández, D.
    Greenman, J.
    Ieropoulos, I.A.
    Energy, 2021, 213
  • [49] High performance liquid chromatographic measurement of iothalamate in human serum and urine for the estimation and long-term follow-up evaluation of glomerular filtration rate.
    Bi, D.
    Leary, K. J.
    Weitz, J. A.
    Cherstiniakova, S. A.
    Riel, M. A.
    Roy, M. J.
    Cantilena, L. R.
    CLINICAL PHARMACOLOGY & THERAPEUTICS, 2007, 81 : S55 - S55
  • [50] Evaluation of a particle enhanced turbidimetric assay for the measurement of neutrophil gelatinase-associated lipocalin in plasma and urine on Architect-8000: Analytical performance and establishment of reference values
    Makris, Konstantinos
    Stefani, Dimitra
    Makri, Eleni
    Panagou, Ioannis
    Lagiou, Maria
    Sarli, Antonia
    Lelekis, Moysis
    Kroupis, Christos
    CLINICAL BIOCHEMISTRY, 2015, 48 (18) : 1291 - 1297